Viscosity and binder composition effects on tyrosinase-based carbon paste electrode for detection of phenol and catechol

被引:35
作者
Rogers, KR
Becker, JY
Cembrano, J
Chough, SH
机构
[1] US EPA, Natl Exposure Res Lab, Las Vegas, NV 89119 USA
[2] Ben Gurion Univ Negev, Dept Chem, IL-84105 Beer Sheva, Israel
关键词
biosensor; tyrosinase; phenol; carbon paste electrode; viscosity;
D O I
10.1016/S0039-9140(01)00376-9
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The systematic study of the effect of binder viscosity on the sensitivity of a tyrosinase-based carbon paste electrode (CPE) biosensor for phenol and catechol is reported. Silicon oil binders with similar (polydimethylsiloxane) chemical composition were used to represent a wide range of viscosities (10-60 000 mPa s(-1) at 25 degreesC) while minimizing polarity effects. The highest response for both phenol and catechol was achieved using a silicon oil binder of intermediate viscosity (100 mPa s(-1)). The binder viscosity showed no appreciable effect on the direct oxidation of phenol and catechol using a plain CPE, suggesting the involvement of diffusion kinetics in the binder matrix for the enzyme-based CPE. The effect of the relative binder concentration in the carbon paste was measured over the range of 30-70%. Optimal results were obtained using 40% silicon oil. For comparison of the viscosity effects observed with the carbon paste electrode (CPE) containing silicon oil, other low and high viscosity mineral oils and paraffin waxes were also examined. (C) 2001 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:1059 / 1065
页数:7
相关论文
共 21 条
[1]   CARBON PASTE ELECTRODES [J].
ADAMS, RN .
ANALYTICAL CHEMISTRY, 1958, 30 (09) :1576-1576
[2]   Rate-limiting steps of tyrosinase-modified electrodes for the detection of catechol [J].
Burestedt, E ;
Narvaez, A ;
Ruzgas, T ;
Gorton, L ;
Emneus, J ;
Dominguez, E ;
MarkoVarga, G .
ANALYTICAL CHEMISTRY, 1996, 68 (09) :1605-1611
[3]  
*ENV HLTH CTR NAT, 2001, PHEN CHEM BACKGR
[4]  
GORDON L, 1995, ELECTROANAL, V7, P23
[5]  
*IRCH IND POLL PRE, 1996, DEV POLL PREV PROGR
[6]   SENSORS BASED ON CARBON-PASTE IN ELECTROCHEMICAL ANALYSIS - A REVIEW WITH PARTICULAR EMPHASIS ON THE PERIOD 1990-1993 [J].
KALCHER, K ;
KAUFFMANN, JM ;
WANG, J ;
SVANCARA, I ;
VYTRAS, K ;
NEUHOLD, C ;
YANG, Z .
ELECTROANALYSIS, 1995, 7 (01) :5-22
[7]  
Lindgren A, 1996, ANAL LETT, V29, P1055
[8]   EFFECTS OF DIFFERENT ADDITIVES ON A TYROSINASE BASED CARBON-PASTE ELECTRODE [J].
LUTZ, M ;
BURESTEDT, E ;
EMNEUS, J ;
LIDEN, H ;
GOBHADI, S ;
GORTON, L ;
MARKOVARGA, G .
ANALYTICA CHIMICA ACTA, 1995, 305 (1-3) :8-17
[9]   DEVELOPMENT OF ENZYME-BASED AMPEROMETRIC SENSORS FOR THE DETERMINATION OF PHENOLIC-COMPOUNDS [J].
MARKOVARGA, G ;
EMNEUS, J ;
GORTON, L ;
RUZGAS, T .
TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 1995, 14 (07) :319-328
[10]  
O'M Bockris J., 1973, Modern Electrochemistry: An Introduction to an Interdisciplinary Area