Determination of nano-molar levels of formaldehyde in drinking water using flow-injection system with immobilized formaldehyde dehydrogenase after off-line solid-phase extraction

被引:42
作者
Kiba, N [1 ]
Sun, L [1 ]
Yokose, S [1 ]
Kazue, MT [1 ]
Suzuki, TT [1 ]
机构
[1] Yamanashi Univ, Fac Engn, Dept Appl Chem & Biotechnol, Kofu, Yamanashi 4008511, Japan
关键词
flow injection; formaldehyde; immobilization; solid-phase extraction;
D O I
10.1016/S0003-2670(98)00636-9
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Low levels of formaldehyde were determined by preconcentration with poly(allylamine) beads and analysis by flow-injection system with immobilized formaldehyde dehydrogenase. Formaldehyde dehydrogenase was immobilized on tresylate-poly(vinyl alcohol) beads and packed into a stainless-steel column (5 cm x 4 mm). The column was incorporated in a flow-injection system with fluorimetic detection. The calibration graph was linear from 0.5 to 10 mu M (15-300 mu g l(-1)). The poly(allylamine) beads (0.5 g) were used to adsorb formaldehyde present at 20-400 nM (0.6-12 mu g l(-1)) from drinking water (11). Formaldehyde was eluted with 1 M HCl (10 ml). The solution (50 mu l) adjusted to pH 9 was injected into the flow-injection system. Concentration factor was 25-fold. Recovery of formaldehyde spiked into purified water was >96% with a relative standard deviation of <3.0%. (C) 1999 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:169 / 175
页数:7
相关论文
共 28 条
[1]   Determination of formaldehyde in frozen fish with formaldehyde dehydrogenase using a flow injection system with an incorporated gel-filtration chromatography column [J].
Bechmann, IE .
ANALYTICA CHIMICA ACTA, 1996, 320 (2-3) :155-164
[2]   OCCURRENCE OF ORGANOHALIDES IN CHLORINATED DRINKING WATERS [J].
BELLAR, TA ;
LICHTENBERG, JJ ;
KRONER, RC .
JOURNAL AMERICAN WATER WORKS ASSOCIATION, 1974, 66 (12) :703-706
[3]   EFFECT OF PH ON THE REACTION OF 2,4-DINITROPHENYLHYDRAZINE WITH FORMALDEHYDE AND ACETALDEHYDE [J].
BICKING, MKL ;
COOKE, WM ;
KAWAHARA, FK ;
LONGBOTTOM, JE .
JOURNAL OF CHROMATOGRAPHY, 1988, 455 :310-315
[4]   LOW-LEVEL DETERMINATION OF FORMALDEHYDE IN WATER BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY [J].
COTSARIS, E ;
NICHOLSON, BC .
ANALYST, 1993, 118 (03) :265-268
[5]   DETERMINATION OF FORMALDEHYDE BY HPLC AS THE DNPH DERIVATIVE FOLLOWING HIGH-VOLUME AIR SAMPLING ONTO BISULFITE-COATED CELLULOSE FILTERS [J].
DEANDRADE, JB ;
TANNER, RL .
ATMOSPHERIC ENVIRONMENT PART A-GENERAL TOPICS, 1992, 26 (05) :819-825
[6]   A MODEL BASED ON MOLECULAR-STRUCTURE DESCRIPTORS FOR PREDICTING MUTAGENICITY OF ORGANIC-COMPOUNDS [J].
DEFLORA, S ;
KOCH, R ;
STROBEL, K ;
NAGEL, M .
TOXICOLOGICAL AND ENVIRONMENTAL CHEMISTRY, 1985, 10 (02) :157-170
[7]   SELECTIVE FLOW-INJECTION DETERMINATION OF METHANOL USING IMMOBILIZED ENZYME REACTORS [J].
DEMARIA, CG ;
MANZANO, T ;
DUARTE, R ;
ALONSO, A .
ANALYTICA CHIMICA ACTA, 1995, 309 (1-3) :241-250
[8]   FAST FLUOROMETRIC FLOW-INJECTION ANALYSIS OF FORMALDEHYDE IN ATMOSPHERIC WATER [J].
DONG, S ;
DASGUPTA, PK .
ENVIRONMENTAL SCIENCE & TECHNOLOGY, 1987, 21 (06) :581-588
[9]   OZONATION BY-PRODUCTS .2. IMPROVEMENT OF AN AQUEOUS-PHASE DERIVATIZATION METHOD FOR THE DETECTION OF FORMALDEHYDE AND OTHER CARBONYL-COMPOUNDS FORMED BY THE OZONATION OF DRINKING-WATER [J].
GLAZE, WH ;
KOGA, M ;
CANCILLA, D .
ENVIRONMENTAL SCIENCE & TECHNOLOGY, 1989, 23 (07) :838-847