Cardioverter-defibrillator oversensing due to double counting of ventricular tachycardia electrograms

被引:11
作者
Boriani, G
Biffi, M
Frabetti, L
Parlapiano, M
Galli, R
Branzi, A
Magnani, B
机构
[1] Univ Bologna, Policlin S Orsola, Inst Cardiol, I-40138 Bologna, Italy
[2] Univ Bologna, Policlin S Orsola, Dept Cardiac Surg, I-40138 Bologna, Italy
关键词
cardioverter-defibrillator; ventricular tachycardia; troubleshooting;
D O I
10.1016/S0167-5273(98)00189-2
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
We report a case of a patient with idiopathic dilated cardiomyopathy and recurrent ventricular tachycardias refractory to antiarrhythmic treatment with amiodarone, A cardioverter defibrillator implantation was performed by the transvenous technique, but ventricular tachycardia detection resulted to be inappropriate because of constant double sensing of ventricular tachycardia electrograms (QRS width=250 ms). Device programmability didn't allow a satisfactory solution to this problem, therefore a more appropriate sensing system was considered. Through an anterior thoracotomy two epicardial wires were positioned acid sensing by these wires, placed closer to ventricular tachycardia origin, resulted appropriate. An electrophysiologic study and subsequent follow up confirmed appropriate ICD detection of ventricular tachycardias. This case emphasizes how in some cases sensing by epicardial wires may be a solution for QRS double counting occurring with endocardial leads during ventricular tachycardia. (C) 1998 Elsevier Science Ireland Ltd. All rights reserved.
引用
收藏
页码:91 / 95
页数:5
相关论文
共 9 条
[1]   CLINICAL-EXPERIENCE WITH A TIERED-THERAPY, MULTIPROGRAMMABLE ANTIARRHYTHMIA DEVICE [J].
BARDY, GH ;
TROUTMAN, C ;
POOLE, JE ;
KUDENCHUK, PJ ;
DOLACK, GL ;
JOHNSON, G ;
HOFER, B .
CIRCULATION, 1992, 85 (05) :1689-1698
[2]  
HOOK BG, 1991, J AM COLL CARDIOL, V17, P985
[3]   OVERSENSING DURING VENTRICULAR PACING IN PATIENTS WITH A 3RD-GENERATION IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR [J].
KELLY, PA ;
MANN, DE ;
DAMLE, RS ;
REITER, MJ .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1994, 23 (07) :1531-1534
[4]   ICD DATA-STORAGE - VALUE IN ARRHYTHMIA MANAGEMENT [J].
MARCHLINSKI, FE ;
GOTTLIEB, CD ;
SARTER, B ;
FINKLE, J ;
HOOK, B ;
CALLANS, D ;
SCHWARTZMAN, D .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1993, 16 (03) :527-534
[5]   USE OF TELEMETRY FUNCTIONS IN THE ASSESSMENT OF IMPLANTED ANTITACHYCARDIA DEVICE EFFICACY [J].
NEWMAN, D ;
DORIAN, P ;
DOWNAR, E ;
HARRIS, L ;
CAMERON, D ;
WAXMAN, M ;
HAMILTON, R ;
GOW, R ;
HARDY, J .
AMERICAN JOURNAL OF CARDIOLOGY, 1992, 70 (06) :616-621
[6]  
ONUNAIN SO, 1995, CIRCULATION, V91, P2204
[7]   Sensing and tachyarrhythmia detection problems in implantable cardioverter defibrillators [J].
Reiter, MJ ;
Mann, DE .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 1996, 7 (06) :542-558
[8]   POSTOPERATIVE LEAD-RELATED COMPLICATIONS IN PATIENTS WITH NONTHORACOTOMY DEFIBRILLATION LEAD SYSTEMS [J].
SCHWARTZMAN, D ;
NALLAMOTHU, N ;
CALLANS, DJ ;
PREMINGER, MW ;
GOTTLIEB, CD ;
MARCHLINSKI, FE .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1995, 26 (03) :776-786
[9]   SENSING PACING LEAD COMPLICATIONS WITH A NEWER GENERATION IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR - WORLDWIDE EXPERIENCE FROM THE GUARDIAN-ATP-4210 CLINICAL-TRIAL [J].
STAMBLER, BS ;
WOOD, MA ;
DAMIANO, RJ ;
GREENWAY, PS ;
SMUTKA, ML ;
ELLENBOGEN, KA ;
BATSFORD, W ;
BAUERNFEIND, R ;
BECKMAN, K ;
BROWNSTEIN, S ;
CASTLE, L ;
DEERING, T ;
ELLENBOGEN, K ;
ESTES, NAM ;
FRANKLIN, J ;
FREIHLING, T ;
HAISTY, K ;
KEHOE, R ;
LOVE, C ;
LUCERI, R ;
NACCARELLI, G ;
PACIFICO, A ;
PARSONNET, V ;
PETROPOULOS, T ;
SAKSENA, S ;
SHARMA, A ;
SINGER, I ;
VENDITTI, F ;
WALTER, P ;
ZIPES, D ;
DAVIS, MJ ;
HELLESTRAND, K ;
KUCHAR, D ;
POWELL, A ;
STAFFORD, W ;
UTHER, J ;
VOHRA, J ;
DORIAN, P ;
GULAMHUSEIN, S ;
KERR, CR ;
ROSENGARTEN, M ;
ROY, D ;
TANG, ASL ;
WILLIAMS, W ;
YEE, R ;
MOLLER, M ;
THOMSEN, PEB ;
ALIOT, E ;
LECLERQ, JF ;
LEMAREC, H .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1994, 23 (01) :123-132