Sensing and tachyarrhythmia detection problems in implantable cardioverter defibrillators

被引:49
作者
Reiter, MJ
Mann, DE
机构
[1] Department of Medicine, Univ. of Colorado Hlth. Sci. Center, Denver, CO
[2] Division of Cardiology, B-130, Univ. of Colorado Hlth. Sci. Center, Denver, CO 80262
关键词
implantable cardioverter defibrillator; overdetection; sensing; underdetection; ventricular tachycardia; ventricular fibrillation;
D O I
10.1111/j.1540-8167.1996.tb00562.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
ICD Sensing-Detection Errors. Sensing of cardiac activity and detection of tachyarrhythmias in implantable cardioverter defibrillators (ICDs) are complex functions and errors occur. Sources of sensing-detection errors include the variable nature of intracardiac electrograms, the occasional inability of automatically adjusting signal amplifiers to cope with this variability, problems with sensing leads, inappropriate programming, and limitations of tachyarrhythmia detection algorithms, which are optimized to avoid underdetection of ventricular tachyarrhythmias. Current ICDs vary considerably in details of sensing and detection function, programmability, and diagnostic data, so that a thorough knowledge of each device is necessary to diagnose and correct these problems. Stored intracardiac electrograms and/or marker channels available in most of these de,ices have contributed much to our understanding of sensing-detection errors, Undersensing of individual signals, most frequently due to signal variability and/or inability of the amplifier to adjust adequately, can lead to delay or failure of tachyarrhythmia detection. Delay or failure of tachyarrhythmia detection can also occur if algorithms to enhance specificity, such as sudden onset or rate stability, are utilized. Oversensing of T waves or noise can lead to false detection; however, the most common cause of false detection is the inability of current detection algorithms to distinguish supraventricular from ventricular tachyarrhythmias. New algorithms that incorporate atrial sensing, electrogram morphology analysis, or hemodynamic monitoring may result in improved detection accuracy of ICDs in the future.
引用
收藏
页码:542 / 558
页数:17
相关论文
共 42 条
[1]   INAPPROPRIATE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR SHOCKS SECONDARY TO SENSING LEAD FAILURE - UTILITY OF STORED ELECTROGRAMS [J].
ALMEIDA, HF ;
BUCKINGHAM, TA .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1993, 16 (03) :407-411
[2]  
BARDY GH, 1993, PACE, V16, P858
[3]   COMPARISON OF INITIAL DETECTION AND REDETECTION OF VENTRICULAR-FIBRILLATION IN A TRANSVENOUS DEFIBRILLATOR SYSTEM WITH AUTOMATIC GAIN-CONTROL [J].
BERUL, CI ;
CALLANS, DJ ;
SCHWARTZMAN, DS ;
PREMINGER, MW ;
GOTTLIEB, CD ;
MARCHLINSKI, FE .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1995, 25 (02) :431-436
[4]   HOLTER DOCUMENTED SUDDEN-DEATH IN A PATIENT WITH AN IMPLANTED DEFIBRILLATOR [J].
BIRGERSDOTTERGREEN, U ;
ROSENQVIST, M ;
LINDEMANS, FW ;
RYDEN, L ;
RADEGRAN, K .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1992, 15 (07) :1008-1014
[5]   CLINICAL INTERACTIONS BETWEEN PACEMAKERS AND AUTOMATIC IMPLANTABLE CARDIOVERTER-DEFIBRILLATORS [J].
CALKINS, H ;
BRINKER, J ;
VELTRI, EP ;
GUARNIERI, T ;
LEVINE, JH .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1990, 16 (03) :666-673
[6]   UNIQUE SENSING ERRORS IN 3RD-GENERATION IMPLANTABLE CARDIOVERTER-DEFIBRILLATORS [J].
CALLANS, DJ ;
HOOK, BG ;
KLEIMAN, RB ;
MITRA, RL ;
FLORES, BT ;
MARCHLINSKI, FE .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1993, 22 (04) :1135-1140
[7]   POSTSHOCK SENSING PERFORMANCE IN TRANSVENOUS DEFIBRILLATION LEAD SYSTEMS - ANALYSIS OF DETECTION AND REDETECTION OF VENTRICULAR-FIBRILLATION [J].
CALLANS, DJ ;
SWARNA, US ;
SCHWARTZMAN, D ;
GOTTLIEB, CD ;
MARCHLINSKI, FE .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 1995, 6 (08) :604-612
[8]   EFFECT OF RATE AND COUPLING INTERVAL ON ENDOCARDIAL-R WAVE AMPLITUDE VARIABILITY IN PERMANENT VENTRICULAR SENSING LEAD SYSTEMS [J].
CALLANS, DJ ;
HOOK, BG ;
MARCHLINSKI, FE .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1993, 22 (03) :746-750
[9]   THE USE AND INTERACTION OF PERMANENT PACEMAKERS AND THE AUTOMATIC IMPLANTABLE CARDIOVERTER DEFIBRILLATOR [J].
COHEN, AI ;
WISH, MH ;
FLETCHER, RD ;
MILLER, FC ;
MCCORMICK, D ;
SHUCK, J ;
SHAPIRA, N ;
DELNEGRO, AA .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1988, 11 (06) :704-711
[10]   INCIDENCE, PRESENTATION, DIAGNOSIS, AND MANAGEMENT OF MALFUNCTIONING IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR RATE-SENSING LEADS [J].
DAOUD, EG ;
KIRSH, MM ;
BOLLING, SF ;
STRICKBERGER, SA ;
MORADY, F ;
KOU, WH .
AMERICAN HEART JOURNAL, 1994, 128 (05) :892-895