Sensing and tachyarrhythmia detection problems in implantable cardioverter defibrillators

被引:49
作者
Reiter, MJ
Mann, DE
机构
[1] Department of Medicine, Univ. of Colorado Hlth. Sci. Center, Denver, CO
[2] Division of Cardiology, B-130, Univ. of Colorado Hlth. Sci. Center, Denver, CO 80262
关键词
implantable cardioverter defibrillator; overdetection; sensing; underdetection; ventricular tachycardia; ventricular fibrillation;
D O I
10.1111/j.1540-8167.1996.tb00562.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
ICD Sensing-Detection Errors. Sensing of cardiac activity and detection of tachyarrhythmias in implantable cardioverter defibrillators (ICDs) are complex functions and errors occur. Sources of sensing-detection errors include the variable nature of intracardiac electrograms, the occasional inability of automatically adjusting signal amplifiers to cope with this variability, problems with sensing leads, inappropriate programming, and limitations of tachyarrhythmia detection algorithms, which are optimized to avoid underdetection of ventricular tachyarrhythmias. Current ICDs vary considerably in details of sensing and detection function, programmability, and diagnostic data, so that a thorough knowledge of each device is necessary to diagnose and correct these problems. Stored intracardiac electrograms and/or marker channels available in most of these de,ices have contributed much to our understanding of sensing-detection errors, Undersensing of individual signals, most frequently due to signal variability and/or inability of the amplifier to adjust adequately, can lead to delay or failure of tachyarrhythmia detection. Delay or failure of tachyarrhythmia detection can also occur if algorithms to enhance specificity, such as sudden onset or rate stability, are utilized. Oversensing of T waves or noise can lead to false detection; however, the most common cause of false detection is the inability of current detection algorithms to distinguish supraventricular from ventricular tachyarrhythmias. New algorithms that incorporate atrial sensing, electrogram morphology analysis, or hemodynamic monitoring may result in improved detection accuracy of ICDs in the future.
引用
收藏
页码:542 / 558
页数:17
相关论文
共 42 条
[11]  
Ellenbogen K A, 1992, J Electrocardiol, V25 Suppl, P143, DOI 10.1016/0022-0736(92)90081-A
[12]   APPROPRIATE SENSING OF VENTRICULAR-FIBRILLATION AFTER FAILED SHOCKS IN A TRANSVENOUS CARDIOVERTER-DEFIBRILLATOR SYSTEM [J].
ELLIS, JR ;
MARTIN, DT ;
VENDITTI, FJ .
CIRCULATION, 1994, 90 (04) :1820-1825
[13]   FAILURE OF ONE CONDUCTOR IN A NONTHORACOTOMY IMPLANTABLE DEFIBRILLATOR LEAD CAUSING INAPPROPRIATE SENSING AND POTENTIALLY INEFFECTIVE SHOCK DELIVERY [J].
EPSTEIN, AE ;
SHEPARD, RB .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1993, 16 (04) :796-800
[14]   EFFICACY OF AUTOMATIC MULTIMODAL DEVICE THERAPY FOR VENTRICULAR TACHYARRHYTHMIAS AS DELIVERED BY A NEW IMPLANTABLE PACING CARDIOVERTER-DEFIBRILLATOR - RESULTS OF A EUROPEAN MULTICENTER STUDY OF 102 IMPLANTS [J].
FROMER, M ;
BRACHMANN, J ;
BLOCK, M ;
SIEBELS, J ;
HOFFMANN, E ;
ALMENDRAL, J ;
OHM, OJ ;
DENDULK, K ;
COUMEL, P ;
CAMM, AJ ;
TOUBOUL, P .
CIRCULATION, 1992, 86 (02) :363-374
[15]  
HERRE JM, 1993, CIRCULATION, V88, P154
[16]  
HOOK BG, 1991, J AM COLL CARDIOL, V17, P985
[17]   IMPORTANCE OF ABORTIVE SHOCK CAPABILITY WITH ELECTROGRAM STORAGE IN CARDIOVERTER-DEFIBRILLATOR DEVICES [J].
HURWITZ, JL ;
HOOK, BG ;
FLORES, BT ;
MARCHLINSKI, FE .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1993, 21 (04) :895-900
[18]   ARRHYTHMIAS INDUCED BY DEVICE ANTITACHYCARDIA THERAPY DUE TO DIAGNOSTIC NONSPECIFICITY [J].
JOHNSON, NJ ;
MARCHLINSKI, FE .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1991, 18 (05) :1418-1425
[19]   FAILURE OF AN IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR TO REDETECT VENTRICULAR-FIBRILLATION IN PATIENTS WITH A NONTHORACOTOMY LEAD SYSTEM [J].
JUNG, W ;
MANZ, M ;
MOOSDORF, R ;
LUDERITZ, B .
CIRCULATION, 1992, 86 (04) :1217-1222
[20]   THE AUTOMATIC IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR - EFFICACY, COMPLICATIONS AND SURVIVAL IN PATIENTS WITH MALIGNANT VENTRICULAR ARRHYTHMIAS [J].
KELLY, PA ;
CANNOM, DS ;
GARAN, H ;
MIRABAL, GS ;
HARTHORNE, JW ;
HURVITZ, RJ ;
VLAHAKES, GJ ;
JACOBS, ML ;
ILVENTO, JP ;
BUCKLEY, MJ ;
RUSKIN, JN .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1988, 11 (06) :1278-1286