Polymer-based lanthanide luminescent sensor for detection of the hydrolysis product of the nerve agent Soman in water

被引:231
作者
Jenkins, AL
Uy, OM
Murray, GM
机构
[1] Johns Hopkins Univ, Appl Phys Lab, Laurel, MD 20723 USA
[2] Univ Maryland Baltimore Cty, Dept Chem & Biochem, Baltimore, MD 21250 USA
关键词
D O I
10.1021/ac980985r
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The techniques of molecular imprinting and sensitized lanthanide luminescence have been combined to create the basis for a sensor that can selectively measure the hydrolysis product of the nerve agent Soman in water. The sensor functions by selectively and reversibly binding the phosphonate hydrolysis product of this agent to a functionality-imprinted copolymer possessing a coordinatively bound luminescent lanthanide ion, Eu3+. Instrumental support for this device is designed to monitor the appearance of a narrow luminescence band in the 610-mn region of the Eu3+ spectrum that results when the analyte is coordinated to the copolymer. The ligand held shifted luminescence was excited using 1 mW of the 465.8-nm line of an argon ion laser and monitored via an optical fiber using a miniature spectrometer. For this configuration, the limit of detection for the hydrolysis product is 7 parts per trillion (ppt) in solution with a linear range from 10 ppt to 10 ppm. Chemical and spectroscopic selectivities have been combined to reduce the likelihood of false positive analyses. Chemically analogous organophosphorus pesticides tested against the sensor have been shown to not interfere with determination.
引用
收藏
页码:373 / 378
页数:6
相关论文
共 18 条