Evaluation of anterior segment parameters in keratoconic eyes measured with the Pentacam system

被引:127
作者
Emre, Sinan [1 ]
Doganay, Selim [1 ]
Yologlu, Saim [1 ]
机构
[1] Inonu Univ, Sch Med, Turgut Ozal Med Ctr, Dept Ophthalmol, Malatya, Turkey
关键词
D O I
10.1016/j.jcrs.2007.06.020
中图分类号
R77 [眼科学];
学科分类号
100212 [眼科学];
摘要
PURPOSE: To evaluate the alteration in anterior chamber parameters with the progression of keratoconus using a rotating Scheimpflug camera. SETTING: Inonu University, Turgut Ozal Medical Center, Ophthalmology Department, Malatya, Turkey. METHODS: Two hundred sixteen eyes of 123 patients diagnosed with keratoconus and 224 eyes of 112 normal subjects were evaluated with a Pentacam Scheimpflug camera (Oculus Inc.). Keratoconic eyes were divided into 3 groups according to mean keratometer (K) readings: mild (K = less than 47.0 diopters [D]), moderate (K = 47.0 to 52.0 D), and severe (K = 52.0 D or higher). The following parameters were obtained: thinnest corneal thickness (TCT), anterior chamber depth (ACD), corneal volume (CV), anterior chamber angle (ACA), and anterior chamber volume (ACV). RESULTS: One hundred twenty-nine eyes had mild keratoconus, 59 had moderate keratoconus, and 35 had severe keratoconus. There were no statistically significant differences between the keratoconus group and control group in age, sex, or eye distribution (P>.05). With the progression of the disease, TCT and ACID were statistically different between all groups (P<.05). There were statistically significant differences in ACA and CV measurements between the mild keratoconus and severe keratoconus groups (P<.05). Also, CV readings were significantly different between the control group and all keratoconus groups (P<.05), although there was no significant difference in ACA measurements (P>.05). Despite a progressive increase in the ACV with disease progression, the differences between groups were not statistically significant (P>.05) except between the severe keratoconus group and the control group (P<.05). CONCLUSION: Anterior segment parameters of the eye showed significant alterations with the progression of keratoconus.
引用
收藏
页码:1708 / 1712
页数:5
相关论文
共 23 条
[1]
Intracorneal ring segments for keratoconus correction:: Long-term follow-up [J].
Alio, Jorge L. ;
Shabayek, Mohamed H. ;
Artola, Alberto .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2006, 32 (06) :978-985
[2]
Corneal-thickness spatial profile and corneal-volume distribution: Tomographic indices to detect keratoconus [J].
Ambrosio, Renato, Jr. ;
Alonso, Ruiz Simonato ;
Luz, Allan ;
Coca Velarde, Luis Guillermo .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2006, 32 (11) :1851-1859
[3]
Central corneal thickness measurement with the Pentacam Scheimpflug system, optical low-coherence reflectometry pachymeter, and ultrasound pachymetry [J].
Barkana, Y ;
Gerber, Y ;
Elbaz, U ;
Schwartz, S ;
Ken-Dror, G ;
Avni, I ;
Zadok, D .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2005, 31 (09) :1729-1735
[4]
Budo C, 2005, J REFRACT SURG, V21, P218
[5]
Comparison of three methods of measuring corneal thickness and anterior chamber depth [J].
Buehl, W ;
Stojanac, D ;
Sacu, S ;
Drexler, W ;
Findl, O .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 2006, 141 (01) :7-12
[6]
Defining subclinical keratoconus using corneal first-surface higher-order aberrations [J].
Buehren, Jens ;
Kuehne, Christoph ;
Kohnen, Thomas .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 2007, 143 (03) :381-389
[7]
Chopra Ira, 2005, Clin Exp Optom, V88, P146
[8]
European clinical evaluation: Use of Intacs for the treatment of keratoconus [J].
Colin, J .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2006, 32 (05) :747-755
[9]
Corneal pachymetry in normal and keratoconic eyes: Orbscan II versus ultrasound [J].
Gherghel, D ;
Hosking, SL ;
Mantry, S ;
Banerjee, S ;
Naroo, SA ;
Shah, S .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2004, 30 (06) :1272-1277
[10]
Hellstedt T, 2005, J REFRACT SURG, V21, P236