Corneal pachymetry in normal and keratoconic eyes: Orbscan II versus ultrasound

被引:76
作者
Gherghel, D
Hosking, SL [1 ]
Mantry, S
Banerjee, S
Naroo, SA
Shah, S
机构
[1] Aston Univ, Sch Life & Hlth Sci, Neurosci Res Inst, Birmingham B4 7ET, W Midlands, England
[2] Heartlands & Solihull NHS Trust, Dept Ophthalmol, Birmingham, W Midlands, England
关键词
D O I
10.1016/j.jcrs.2003.11.049
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To compare corneal thickness measurements using Orbscan II (OII) and ultrasonic (US) pachymetry in normal and in keratoconic eyes. Setting: Eye Department, Heartlands and Solihull NHS Trust, Birmingham, United Kingdom. Methods: Central corneal thickness (CCT) was measured by means of Oil and US pachymetry in 1 eye of 72 normal subjects and 36 keratoconus patients. The apical corneal thickness (ACT) in keratoconus patients was also evaluated using each method. The mean of the difference, standard deviation (SD), and 95% limits of agreement (LoA = mean +/- 2 SD), with and without applying the default linear correction factor (LCF), were determined for each sample. The Student t test was used to identify significant differences between methods, and the correlation between methods was determined using the Pearson bivariate correlation. Bland-Altman analysis was performed to confirm that the results of the 2 instruments were clinically comparable. Results: In normal eyes, the mean difference (+/- 95% LoA) in CCT was 1.04 mum +/- 68.52 (SD) (P>.05; r = 0.71) when the LCF was used and 46.73 +/- 75.40 mum (P = .0001; r = 0.71) without the LCF In keratoconus patients, the mean difference ( 95% LoA) in CCT between methods was 42.46 +/- 66.56 mum (P<.0001: r = 0.85) with the LCF, and 2.51 +/- 73.00 mu m (P>.05: r = 0.85) without the LCF. The mean difference (+/- 95% LoA) in ACT for this group was 49.24 +/- 60.88 mum (P<.0001: r = 0.89) with the LCF and 12.71 +/- 68.14 mu m (P = .0077; r = 0.89) when the LCF was not used. Conclusions: This study suggests that Oil and US pachymetry provide similar readings for CCT in normal subjects when an LCF is used. In keratoconus patients, Oil provides a valid clinical tool for the noninvasive assessment of CCT when the LCF is not applied.
引用
收藏
页码:1272 / 1277
页数:6
相关论文
共 21 条
[1]   Keratoconus evaluation using the Orbscan Topography System [J].
Auffarth, GU ;
Wang, L ;
Völcker, HE .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2000, 26 (02) :222-228
[2]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[3]   Accuracy of Orbscan optical pachymetry in corneas with haze [J].
Boscia, F ;
La Tegola, MG ;
Alessio, G ;
Sborgia, C .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2002, 28 (02) :253-258
[4]   Comparison of corneal thickness measurements using ultrasound and Orbscan slit-scanning topography in normal and post-LASIK eyes [J].
Chakrabarti, HS ;
Craig, JP ;
Brahma, A ;
Malik, TY ;
McGhee, CNJ .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2001, 27 (11) :1823-1828
[5]   Comparison of corneal pachymetry using ultrasound and Orbscan II [J].
Fakhry, MA ;
Artola, A ;
Belda, JL ;
Ayala, MJ ;
Alió, JL .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2002, 28 (02) :248-252
[6]  
Franco S, 2000, J REFRACT SURG, V16, pS661
[7]   Central and peripheral corneal thickness measurement with Orbscan II and topographical ultrasound pachymetry [J].
González-Méijome, JM ;
Cerviño, A ;
Yebra-Pimentel, E ;
Parafita, MA .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2003, 29 (01) :125-132
[8]   INTERFEROMETRIC MEASUREMENT OF CORNEAL THICKNESS WITH MICROMETER PRECISION [J].
HITZENBERGER, CK ;
BAUMGARTNER, A ;
DREXLER, W ;
FERCHER, AF .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 1994, 118 (04) :468-476
[9]   A histochemical and x-ray diffraction study of keratoconus epikeratoplasty - Report of two cases [J].
Huang, YF ;
Tuft, SJ ;
Meek, KM .
CORNEA, 1996, 15 (03) :320-328
[10]   Accuracy of Orbscan pachymetry measurements and DHG ultrasound pachymetry in primary laser in situ keratomileusis and LASIK enhancement procedures [J].
Iskander, NG ;
Penno, EA ;
Peters, NT ;
Gimbel, HV ;
Ferensowicz, M .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2001, 27 (05) :681-685