Multiple-beam X-ray interferometry for phase-contrast microtomography

被引:13
作者
Bonse, U [1 ]
Beckmann, F
机构
[1] Univ Dortmund, Inst Phys, D-44221 Dortmund, Germany
[2] DESY, Hamburger Synchrontronstschlungslab HASYLAB, D-22603 Hamburg, Germany
关键词
microtomography; multiple-beam X-ray interferometry; phase-contrast microtomography;
D O I
10.1107/S0909049500015843
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The first successful operation of an X-ray interferometer under conditions of non-planar three-beam diffraction for phase-contrast X-ray microtomography is reported. Intrinsic phase differences of the reflections used cancel from the three-dimensional phase image of the specimen. With simultaneous hkl and hk (l) over bar reflections of a synchrotron radiation beam in a side-by-side geometry, the size of the usable field of view is doubled and the investigated specimen volume is increased by a factor of four. As an example, the reconstructed slice of a mouse kidney is shown in phase contrast at 71 keV. Optimized choices of three-beam reflections and matching interferometer geometries useful for applications are presented.
引用
收藏
页码:1 / 5
页数:5
相关论文
共 12 条
[1]   TEST MEASUREMENTS WITH A PERFECT CRYSTAL NEUTRON INTERFEROMETER [J].
BAUSPIESS, W ;
BONSE, U ;
RAUCH, H ;
TREIMER, W .
ZEITSCHRIFT FUR PHYSIK, 1974, 271 (02) :177-182
[2]   X-ray microtomography (mu CT) using phase contrast for the investigation of organic matter [J].
Beckmann, F ;
Bonse, U ;
Busch, F ;
Gunnewig, O .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1997, 21 (04) :539-553
[3]  
BECKMANN F, 1995, HASYLAB ANN REP, V2, P691
[4]   AN X-RAY INTERFEROMETER WITH LONG SEPARATED INTERFERING BEAM PATHS (E) [J].
BONSE, U ;
HART, M .
APPLIED PHYSICS LETTERS, 1965, 7 (04) :99-&
[5]   X-ray computed microtomography (mu CT) using synchrotron radiation (SR) [J].
Bonse, U ;
Busch, F .
PROGRESS IN BIOPHYSICS & MOLECULAR BIOLOGY, 1996, 65 (1-2) :133-169
[6]   AN X-RAY INTERFEROMETER [J].
BONSE, U ;
HART, M .
APPLIED PHYSICS LETTERS, 1965, 6 (08) :155-&
[7]   3-BEAM CASE X-RAY INTERFEROMETER [J].
GRAEFF, W ;
BONSE, U .
ZEITSCHRIFT FUR PHYSIK B-CONDENSED MATTER, 1977, 27 (01) :19-32
[8]  
GRAEFF W, 1991, HDB SYNCHROTRON RAD, V4, P361
[9]  
GRAEFF W, 1996, P INT SCH PHYS ENR F, P155
[10]  
Kak AC, 1987, PRINCIPLES COMPUTERI