Quantitative analysis of chemical warfare agent degradation products in reaction masses using capillary electrophoresis

被引:43
作者
Nassar, AEF
Lucas, SV
Myler, CA
Jones, WR
Campisano, M
Hoffland, LD
机构
[1] Battelle Mem Inst, Bel Air, MD 21015 USA
[2] Thermo BioAnal Corp, Franklin, MA 02038 USA
[3] USA, Mat Command Treaty Lab, Aberdeen Proving Ground, MD 21010 USA
关键词
D O I
10.1021/ac9713870
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Quantitative methods have been developed for the analysis of chemical warfare agent degradation products in reaction masses using capillary electrophoresis (CE), This is the first report of a systematic validation of a CE-based method for the analysis of chemical warfare agent degradation products in agent neutralization matrixes (reaction masses). After neutralization with monoethanolamine/water, the nerve agent GB (isopropyl methylphosphonofluoridate, Sarin) gives isopropyl methylphosphonic acid (IMPA) and O-isopropyl O'-(2-amino)ethyl methylphosphonate (GB-MEA adduct). The nerve agent GD (pinacolyl methylphosphonofluoridate, Soman), [pinacolyl= 2-(3,3-dimethyl)butyl] produces pinacolyl methylphosphonic acid (PMPA) and O-pinacolyl O'-(2-amino)ethyl methylphosphonate (GD-MEA adduct), The samples were prepared by dilution of the reaction masses with deionized water before analysis by CE/indirect UV detection or CE/conductivity detection. Migration time precision was less than 4.0% RSD for IMPA and 5.0% RSD for PMPA on a day-to-day basis. The detection limit for both IMPA and PMPA is 100 mu g/L; the quantitation limit for both is 500 mu g/L. For calibration standards, IMPA and PMPA gave a linear response (R-2 = 0.9999) over the range 0.5-100 mu g/mL. The interday precision RSDs were 1.9, 1.0, and 0.7% for IMPA at 7.5, 37.5 and 75.0 mu g/mL, respectively, Corresponding values for PMPA (again, RSD) were 2.9, 1.1, and 1.0% at 7.5, 37.5 and 87.5 mu g/mL, respectively, as before. Analysis accuracy was assessed by spiking actual neutralization samples with IMPA or PMPA. For IMPA, the seven spike levels used ranged from 20 to 220% of the IMPA background level, and the incremental change in the found IMPA level ranged from 86 to 99% of the true spiking increment (R2 = 0.9987 for the linear regression). For PMPA, the five spike levels ranged from 10 to 150% of the matrix background level, and similarly, the accuracy obtained ranged from 95 to 97% of the true incremental value (R-2 = 0.9999 for the linear regression).
引用
收藏
页码:3598 / 3604
页数:7
相关论文
共 50 条
[31]   Separation of chemical warfare agent degradation products by the reversal of electroosmotic flow in capillary electrophoresis [J].
Nassar, AEF ;
Lucas, SV ;
Jones, WR ;
Hoffland, LD .
ANALYTICAL CHEMISTRY, 1998, 70 (06) :1085-1091
[32]  
NASSAR AEF, 1997, P AM CHEM SOC DIV EN, V37, P39
[33]  
NASSAR AEF, UNPUB
[34]   ANALYSIS OF NERVE AGENT DEGRADATION PRODUCTS USING CAPILLARY ION ELECTROPHORESIS [J].
OEHRLE, SA ;
BOSSLE, PC .
JOURNAL OF CHROMATOGRAPHY A, 1995, 692 (1-2) :247-252
[35]   OPTIMIZATION AND VALIDATION OF ANALYTICAL CONDITIONS FOR BOVINE SERUM-ALBUMIN USING CAPILLARY ELECTROPHORESIS [J].
PANDE, PG ;
NELLORE, RV ;
BHAGAT, HR .
ANALYTICAL BIOCHEMISTRY, 1992, 204 (01) :103-106
[36]   DETERMINATION OF ALKYLPHOSPHONIC ACIDS BY CAPILLARY ZONE ELECTROPHORESIS USING INDIRECT UV DETECTION [J].
PIANETTI, GA ;
TAVERNA, M ;
BAILLET, A ;
MAHUZIER, G ;
BAYLOCQFERRIER, D .
JOURNAL OF CHROMATOGRAPHY, 1993, 630 (1-2) :371-377
[37]   PREPARATION, STABILITY AND QUANTITATIVE-ANALYSIS BY GAS-CHROMATOGRAPHY AND GAS-CHROMATOGRAPHY ELECTRON-IMPACT MASS-SPECTROMETRY OF TERT-BUTYLDIMETHYLSILYL DERIVATIVES OF SOME ALKYLPHOSPHONIC AND ALKYL METHYLPHOSPHONIC ACIDS [J].
PURDON, JG ;
PAGOTTO, JG ;
MILLER, RK .
JOURNAL OF CHROMATOGRAPHY, 1989, 475 :261-272
[38]   FLUORESCENCE DETECTION OF ALKYLPHOSPHONIC ACIDS USING PARA-(9-ANTHROYLOXY)PHENACYL BROMIDE [J].
ROACH, MC ;
UNGAR, LW ;
ZARE, RN ;
REIMER, LM ;
POMPLIANO, DL ;
FROST, JW .
ANALYTICAL CHEMISTRY, 1987, 59 (07) :1056-1059
[39]   CAPILLARY ELECTROPHORETIC SEPARATION OF ORGANOPHOSPHONIC ACIDS USING BORATE ESTERIFICATION AND DIRECT UV DETECTION [J].
ROBINS, WH ;
WRIGHT, BW .
JOURNAL OF CHROMATOGRAPHY A, 1994, 680 (02) :667-673
[40]  
RUPPEL ML, 1976, BIOMED MASS SPECTROM, V3, P28