The influence of inserting a Fuji pressure sensitive film between the tibiofemoral joint of knee prosthesis on actual contact characteristics

被引:42
作者
Liau, JJ
Hu, CC
Cheng, CK
Huang, CH
Lo, WH
机构
[1] Natl Yang Ming Univ, Inst Biomed Engn, Orthopaed Biomech Lab, Taipei, Taiwan
[2] Mackay Mem Hosp, Dept Orthopaed Surg, Taipei, Taiwan
[3] Vet Gen Hosp, Dept Orthopaed & Traumatol, Taipei, Taiwan
关键词
tibiofemoral joint; contact characteristics; finite element model; Fuji pressure sensitive film; knee prosthesis;
D O I
10.1016/S0268-0033(00)00040-1
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Objective. To investigate the influence of inserting a Fuji pressure sensitive film between the tibiofemoral joint of knee prosthesis on actual contact characteristics. Design. A finite element analysis was used to investigate the alteration of contact characteristics of the tibiofemoral joint due to inserting a pressure sensitive film between the contacting surfaces. Background. The discrepancy between actual contact behaviors of tibiofemoral joint of knee prosthesis and the measurement using Fuji pressure sensitive film was not discussed extensively. The change of direct contact circumstance of the tibiofemoral joint due to inserting a pressure sensitive film was not well reported. Method. A two-dimensional finite element model of the tibiofemoral joint of knee prosthesis in the sagittal plane was constructed, Four different radii of the femoral component were used to investigate the conformity effect. Two-layers of plane strain element were used to model the pressure sensitive film, The influence of inserting a pressure sensitive film on the actual contact characteristics was analyzed by comparing the results of the change in contact radius. Results. Inserting a pressure sensitive film between contacting surfaces would disturb the original contact behaviors, especially in the lowest conformity design. The maximum difference of contact radius between the model simulating actual contact circumstance and the model with inserting a pressure sensitive film between contacting surface was 77% in the lowest conformity condition at the smallest load. Conclusions. This study proposes a quantitative analysis of contact characteristics in the tibiofemoral joint of knee prosthesis between the models with and without inserting a pressure sensitive film into the contact surface. The measurement of contact area in artificial tibiofemoral joints by using Fuji pressure sensitive film is always overestimated its true contact area by 14-77%.
引用
收藏
页码:160 / 166
页数:7
相关论文
共 25 条
[1]   A STEREOPHOTOGRAMMETRIC METHOD FOR DETERMINING IN-SITU CONTACT AREAS IN DIARTHROIDAL JOINTS, AND A COMPARISON WITH OTHER METHODS [J].
ATESHIAN, GA ;
KWAK, SD ;
SOSLOWSKY, LJ ;
MOW, VC .
JOURNAL OF BIOMECHANICS, 1994, 27 (01) :111-124
[2]   THE INFLUENCE OF CONTACT STRESS ON THE WEAR OF UHMWPE FOR TOTAL REPLACEMENT HIP PROSTHESES [J].
BARBOUR, PSM ;
BARTON, DC ;
FISHER, J .
WEAR, 1995, 181 :250-257
[3]   THE EFFECT OF CONFORMITY, THICKNESS, AND MATERIAL ON STRESSES IN ULTRAHIGH MOLECULAR-WEIGHT COMPONENTS FOR TOTAL JOINT REPLACEMENT [J].
BARTEL, DL ;
BICKNELL, VL ;
WRIGHT, TM .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1986, 68A (07) :1041-1051
[4]  
BARTEL DL, 1995, CLIN ORTHOP RELAT R, V317, P76
[5]  
BLACK JD, 1981, CLIN ORTHOP RELAT R, P265
[6]   Non-uniformity of contact stress on polyethylene inserts in total knee arthroplasty [J].
Bristol, RE ;
Fitzpatrick, DC ;
Brown, TD ;
Callaghan, JJ .
CLINICAL BIOMECHANICS, 1996, 11 (02) :75-80
[7]  
COLLIER JP, 1991, CLIN ORTHOP RELAT R, V273, P232
[8]   Residual stresses in ultra-high molecular weight polyethylene loaded cyclically by a rigid moving indenter in nonconforming geometries [J].
Estupiñán, JA ;
Bartel, DL ;
Wright, TM .
JOURNAL OF ORTHOPAEDIC RESEARCH, 1998, 16 (01) :80-88
[9]   CONTACT STRESS GRADIENT DETECTION LIMITS OF PRESSENSOR FILM [J].
HALE, JE ;
BROWN, TD .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 1992, 114 (03) :352-357
[10]   An improved method for measuring tibiofemoral contact areas in total knee arthroplasty: a comparison of K-scan sensor and Fuji film [J].
Harris, ML ;
Morberg, P ;
Bruce, WJM ;
Walsh, WR .
JOURNAL OF BIOMECHANICS, 1999, 32 (09) :951-958