Arterial-phase three-dimensional gadolinium magnetic resonance angiography of the renal arteries - Strategies for timing and contrast media injection: Original investigation

被引:42
作者
Schoenberg, SO
Knopp, MV
Prince, MR
Londy, F
Knopp, MA
机构
[1] German Canc Res Ctr, Dept Radiol, D-69120 Heidelberg, Germany
[2] Univ Michigan, Dept Radiol, Ann Arbor, MI 48109 USA
关键词
magnetic resonance angiography; gadolinium chelates; bolus injection; bolus timing;
D O I
10.1097/00004424-199809000-00005
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
RATIONALE AND OBJECTIVES. The authors review different imaging and contrast-media infusion strategies for arterial-phase three-dimensional (3D) gadolinium-enhanced magnetic resonance angiography (Gd-MRA). METHODS. The influence of physicochemical factors on the infusion of contrast media, including viscosity, how rate, inline pressure, and cannula size, is assessed, The combination of manual or automated contrast-media administration with timing-dependent or -independent 3D Gd-MRA techniques is reviewed regarding the aspects of effectiveness, robustness, image quality, and costs. RESULTS. For effective bolus delivery with high flow rates, the type and temperature of the contrast media, the size of the cannula, and an immediate saline hush must he considered. Timing-dependent techniques based on a test bolus and using automated contrast-media infusion as well as timing independent techniques such as MR SmartPrep or multiphase 3D Gd-MRA by using a manual injection with a SmartSet tubing set, are all effective procedures for arterial phase 3D Gd-MRA. CONCLUSIONS. Manual contrast-media injection with a tubing set can be used for timing-independent MRA techniques, The multiphase 3D Gd-MRA approach seems to be favorable for different MR systems, robustness, and speed.
引用
收藏
页码:506 / 514
页数:9
相关论文
共 18 条
[1]   PHASE ERRORS IN MULTISHOT ECHO-PLANAR IMAGING [J].
FEINBERG, DA ;
OSHIO, K .
MAGNETIC RESONANCE IN MEDICINE, 1994, 32 (04) :535-539
[2]  
GOLDFARB JW, 1997, P ISMRM 5 ANN M VANC, P200
[3]   Evaluation of the aortoiliac and renal arteries: Comparison of breath-hold, contrast-enhanced, three-dimensional MR angiography with conventional catheter angiography [J].
Hany, TF ;
Debatin, JF ;
Leung, DA ;
Pfammatter, T .
RADIOLOGY, 1997, 204 (02) :357-362
[4]   Optimization of contrast timing for breath-hold three-dimensional MR angiography [J].
Hany, TF ;
McKinnon, GC ;
Leung, DA ;
Pfammatter, T ;
Debatin, JF .
JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1997, 7 (03) :551-556
[5]   Test bolus examination in the carotid artery at dynamic gadolinium-enhanced MR angiography [J].
Kim, JK ;
Farb, RI ;
Wright, GA .
RADIOLOGY, 1998, 206 (01) :283-289
[6]   Time-resolved contrast-enhanced 3D MR angiography [J].
Korosec, FR ;
Frayne, R ;
Grist, TM ;
Mistretta, CA .
MAGNETIC RESONANCE IN MEDICINE, 1996, 36 (03) :345-351
[7]  
LAUB G, 1996, P 4 M INT SOC MAGN R, P239
[8]  
LEUNG DA, 1996, RADIOLOGY, V201, P569
[9]   The effects of time varying intravascular signal intensity and k-space acquisition order on three-dimensional MR angiography image quality [J].
Maki, JH ;
Prince, MR ;
Londy, FJ ;
Chenevert, TL .
JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1996, 6 (04) :642-651
[10]  
MARGOSIAN PM, 1996, ENCY NUCL MAGNETIC R, P3462