Evaluation of an automated sphygmomanometer for use in the office setting

被引:30
作者
Myers, Martin G. [1 ,2 ]
Valdivieso, Miguel [1 ,2 ]
机构
[1] Sunnybrook Hlth Sci Ctr, Div Cardiol, Schulich Heart Program, Toronto, ON M4N 3M5, Canada
[2] Univ Toronto, Dept Med, Toronto, ON, Canada
关键词
automated sphygmomanometers; blood pressure measurement; hypertension diagnosis; BLOOD-PRESSURE-MEASUREMENT; VALIDATION; DIAGNOSIS; DEVICE; HOME;
D O I
10.1097/MBP.0b013e3283540785
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
Objective To determine whether the WatchBP Office sphygmomanometer can be used to obtain automated office blood pressure (AOBP) readings that are similar to the awake ambulatory BP. Methods One hundred patients referred for 24 h ambulatory BP monitoring had BP recorded three times using the WatchBP Office fully automated sphygmomanometer in accordance with standard AOBP measurement guidelines. The mean AOBP was compared with the mean awake ambulatory BP. Results The mean (+/- SD) AOBP (138.6 +/- 13.7/79.7 +/- 9.0 mmHg) was similar to the mean awake ambulatory BP (136.8 +/- 12.4/79.0 +/- 10.8 mmHg). The small difference in systolic BP (1.8 mmHg) was statistically significant (P = 0.03), but was within the accepted range (5 mmHg) recommended by guidelines for equivalence between BP readings. There was a strong correlation (P < 0.001) between the systolic/diastolic AOBP and awake ambulatory BP readings (r = 0.819/0.801). Conclusion The WatchBP Office produces BP readings that closely approximate the awake ambulatory BP, confirming that this automated sphygmomanometer is suitable for recording AOBP in clinical practice. Blood Press Monit 17:116-119 (C) 2012 Wolters Kluwer Health vertical bar Lippincott Williams & Wilkins.
引用
收藏
页码:116 / 119
页数:4
相关论文
共 20 条
[1]   Automated Office Blood Pressure and 24-h Ambulatory Measurements are Equally Associated With Left Ventricular Mass Index [J].
Andreadis, Emmanuel A. ;
Agaliotis, Gerasimos D. ;
Angelopoulos, Epameinondas T. ;
Tsakanikas, Athanasios P. ;
Chaveles, Loannis A. ;
Mousoulis, George P. .
AMERICAN JOURNAL OF HYPERTENSION, 2011, 24 (06) :661-666
[2]  
[Anonymous], 2009, 8106022009 ANSIAAMII
[3]   The BpTRU automatic blood pressure monitor compared to 24 hour ambulatory blood pressure monitoring in the assessment of blood pressure in patients with hypertension [J].
Beckett L. ;
Godwin M. .
BMC Cardiovascular Disorders, 5 (1)
[4]  
BeNational Institute for Health and Clinical Excellence, 2011, HYP NICE CLIN GUID
[5]   Correlations between different measures of clinic, home, and ambulatory blood pressure in hypertensive patients [J].
Eguchi, Kazuo ;
Kuruvilla, Sujith ;
Ishikawa, Joji ;
Ogedegbe, Gbenga ;
Gerin, William ;
Schwartz, Joseph E. ;
Pickering, Thomas G. .
BLOOD PRESSURE MONITORING, 2011, 16 (03) :142-148
[6]   Relative effectiveness of clinic and home blood pressure monitoring compared with ambulatory blood pressure monitoring in diagnosis of hypertension: systematic review [J].
Hodgkinson, J. ;
Mant, J. ;
Martin, U. ;
Guo, B. ;
Hobbs, F. D. R. ;
Deeks, J. J. ;
Heneghan, C. ;
Roberts, N. ;
McManus, R. J. .
BMJ-BRITISH MEDICAL JOURNAL, 2011, 342
[7]   Can sphygmomanometers designed for self-measurement of blood pressure in the home be used in office practice? [J].
Myers, Martin G. ;
Valdivieso, Miguel ;
Chessman, Mary ;
Kiss, Alexander .
BLOOD PRESSURE MONITORING, 2010, 15 (06) :300-304
[8]   A proposed algorithm for diagnosing hypertension using automated office blood pressure measurement [J].
Myers, Martin G. .
JOURNAL OF HYPERTENSION, 2010, 28 (04) :703-708
[9]   Measurement of Blood Pressure in the Office Recognizing the Problem and Proposing the Solution [J].
Myers, Martin G. ;
Godwin, Marshall ;
Dawes, Martin ;
Kiss, Alexander ;
Tobe, Sheldon W. ;
Kaczorowski, Janusz .
HYPERTENSION, 2010, 55 (02) :195-200
[10]   Comparison of two automated sphygmomanometers for use in the office setting [J].
Myers, Martin G. ;
Valdivieso, Miguel ;
Kiss, Alexander ;
Tobe, Sheldon W. .
BLOOD PRESSURE MONITORING, 2009, 14 (01) :45-47