Comparison of two automated sphygmomanometers for use in the office setting

被引:37
作者
Myers, Martin G. [1 ,4 ]
Valdivieso, Miguel [1 ]
Kiss, Alexander [3 ]
Tobe, Sheldon W. [2 ,4 ]
机构
[1] Sunnybrook Hlth Sci Ctr, Div Cardiol, Schulich Heart Ctr, Toronto, ON M4N 3M5, Canada
[2] Sunnybrook Hlth Sci Ctr, Div Nephrol, Toronto, ON M4N 3M5, Canada
[3] Sunnybrook Hlth Sci Ctr, Inst Clin & Evaluat Sci, Toronto, ON M4N 3M5, Canada
[4] Univ Toronto, Dept Med, Toronto, ON, Canada
关键词
automated sphygmomanometers; blood pressure measurement; hypertension diagnosis; BLOOD-PRESSURE-MEASUREMENT; INTERNATIONAL-PROTOCOL; VALIDATION; HYPERTENSION; DEVICE;
D O I
10.1097/MBP.0b013e32831e314f
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
Objective To compare readings obtained using two automated blood pressure (BP) recording devices, BpTRU and Omron 907, in a clinical setting. Methods Two series of 50 patients attending a hypertension unit for 24-h ambulatory BP monitoring had BP recorded either every 1 or every 2 min using BpTRU and Omron 907 devices with the order of the measurements being randomized. Results No significant differences for systolic BP between the mean readings taken using the BpTRU or Omron 907 recorders at either 1 or 2 min were observed. Diastolic BP readings were similar using the 1-min interval setting but were 5.0 mmHg lower for the Omron 907 (P < 0.001) when readings were taken at 2-min intervals. Conclusion BP measurements can be made in the clinical setting using either the BpTRU or Omron 907 automated sphygmomanometers. Blood Press Monit 14:45-47 (C) 2009 Wolters Kluwer Health vertical bar Lippincott Williams & Wilkins.
引用
收藏
页码:45 / 47
页数:3
相关论文
共 16 条
[1]  
*ASS ADV MED INSTR, 1993, AM NAT STAND EL AUT, V3330
[2]   The BpTRU automatic blood pressure monitor compared to 24 hour ambulatory blood pressure monitoring in the assessment of blood pressure in patients with hypertension [J].
Beckett L. ;
Godwin M. .
BMC Cardiovascular Disorders, 5 (1)
[3]   Performance of the automated BpTRU™ measurement device in the assessment of white-coat hypertension and white-coat effect [J].
Culleton, BF ;
McKay, DW ;
Campbell, NR .
BLOOD PRESSURE MONITORING, 2006, 11 (01) :37-42
[4]   Validation of the Omron HEM-907 device for blood pressure measurement [J].
El Assaad, MA ;
Topouchian, JA ;
Darné, BM ;
Asmar, RG .
BLOOD PRESSURE MONITORING, 2002, 7 (04) :237-241
[5]  
Hemmelgarn Brenda R, 2005, Can J Cardiol, V21, P645
[6]  
Mancia G, 2007, EUR HEART J, V28, P1462, DOI [10.1093/eurheartj/ehm236, 10.1093/eurheartj/eht151]
[7]   Optimum frequency of office blood pressure measurement using an automated sphygmomanometer [J].
Myers, Martin G. ;
Valdivieso, Miguel ;
Kiss, Alexander .
BLOOD PRESSURE MONITORING, 2008, 13 (06) :333-338
[8]   Automated blood pressure measurement in routine clinical practice [J].
Myers, MG .
BLOOD PRESSURE MONITORING, 2006, 11 (02) :59-62
[9]   Ambulatory blood pressure monitoring for routine clinical practice [J].
Myers, MG .
HYPERTENSION, 2005, 45 (04) :483-484
[10]   Use of an automated blood pressure recording device, the BpTRU, to reduce the "White coat effect" in routine practice [J].
Myers, MG ;
Valdivieso, MA .
AMERICAN JOURNAL OF HYPERTENSION, 2003, 16 (06) :494-497