A controlled trial of smart infusion pumps to improve medication safety in critically ill patients

被引:181
作者
Rothschild, JM
Keohane, CA
Cook, EF
Orav, EJ
Burdick, E
Thompson, S
Hayes, J
Bates, DW
机构
[1] Harvard Univ, Sch Med, Brigham & Womens Hosp, Div Gen Med, Boston, MA 02120 USA
[2] Harvard Univ, Sch Med, Brigham & Womens Hosp, Dept Med, Boston, MA 02120 USA
[3] Brigham & Womens Hosp, Dept Nursing, Boston, MA 02115 USA
[4] Harvard Univ, Sch Publ Hlth, Boston, MA 02115 USA
关键词
medication safety; adverse drug events; intravenous infusion pump; bedside decision support; human factors; critical care nursing; intensive care units; ADVERSE DRUG EVENTS; INTENSIVE-CARE-UNIT; HOSPITALIZED-PATIENTS; INTRAVENOUS DRUG; ERRORS; TECHNOLOGY; VALIDATION; PREVENTION; MORTALITY; DEVICES;
D O I
10.1097/01.ccm.0000155912.73313.cd
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Objective: Intravenous medications are vital during inpatient management. Errors associated with the administration of medications through intravenous infusion pumps to critically ill patients can result in adverse drug events. We sought to assess the impact of smart pumps with integrated decision support software on the incidence and nature of medication errors and adverse drug events. Design: We performed a prospective, randomized time-series trial and compared the serious medication error rate between intervention (decision support on) and control (decision support off) periods. Serious medication errors included both near-misses and preventable adverse drug events. Pump software produced log reports to help identify potential events. Events were presented to physicians for rating of event type, preventability, and severity. Setting: Cardiac surgical intensive care and step-down units between February and December 2002. Patients: Pump data were available for 744 cardiac surgery admissions. Interventions: Decision support during medication administration provided feedback including alerts, reminders, and unit-specific drug rate limits. Measurements and Main Results: We found a total of 180 serious medication errors, including 14 and 11 preventable adverse drug events and 73 and 82 nonintercepted potential adverse drug events in the control and intervention periods, respectively. The serious medication error rates in the control and intervention periods were 2.03 and 2.41 per 100 patient-pump-days, respectively (p = .124). We also found numerous opportunities for safety improvement. Violations of infusion practice during the intervention periods included 571 (25%) bypasses of the drug library. Medications were also frequently administered without documentation of physician orders in both periods (n = 823; 7.7%). Conclusion: Intravenous medication errors and adverse drug events were frequent and could be detected using smart pumps. We found no measurable impact on the serious medication error rate, likely in part due to poor compliance. Although smart pumps have great promise, technological and nursing behavioral factors must be addressed if these pumps are to achieve their potential for improving medication safety.
引用
收藏
页码:533 / 540
页数:8
相关论文
共 38 条
[21]   Identifying adverse drug events: Development of a computer-based monitor and comparison with chart review and stimulated voluntary report [J].
Jha, AK ;
Kuperman, GJ ;
Teich, JM ;
Leape, L ;
Shea, B ;
Rittenberg, E ;
Burdick, E ;
Seger, DL ;
Vander Vliet, M ;
Bates, DW .
JOURNAL OF THE AMERICAN MEDICAL INFORMATICS ASSOCIATION, 1998, 5 (03) :305-314
[22]  
Kohn LT, 2000, To Err Is Human: Building a Safer Health System
[23]  
Kovner C, 2002, HEALTH SERV RES, V37, P611
[24]   SYSTEMS-ANALYSIS OF ADVERSE DRUG EVENTS [J].
LEAPE, LL ;
BATES, DW ;
CULLEN, DJ ;
COOPER, J ;
DEMONACO, HJ ;
GALLIVAN, T ;
HALLISEY, R ;
IVES, J ;
LAIRD, N ;
LAFFEL, G ;
NEMESKAL, R ;
PETERSEN, LA ;
PORTER, K ;
SERVI, D ;
SHEA, BF ;
SMALL, SD ;
SWEITZER, BJ ;
THOMPSON, BT ;
VANDERVLIET, M .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1995, 274 (01) :35-43
[25]   THE NATURE OF ADVERSE EVENTS IN HOSPITALIZED-PATIENTS - RESULTS OF THE HARVARD MEDICAL-PRACTICE STUDY-II [J].
LEAPE, LL ;
BRENNAN, TA ;
LAIRD, N ;
LAWTHERS, AG ;
LOCALIO, AR ;
BARNES, BA ;
HEBERT, L ;
NEWHOUSE, JP ;
WEILER, PC ;
HIATT, H .
NEW ENGLAND JOURNAL OF MEDICINE, 1991, 324 (06) :377-384
[26]   Patient safety, potential adverse drug events, and medical device design: A human factors engineering approach [J].
Lin, L ;
Vicente, KJ ;
Doyle, DJ .
JOURNAL OF BIOMEDICAL INFORMATICS, 2001, 34 (04) :274-284
[27]  
NCC MERP, 1998, Taxonomy of medication errors
[28]   Nurse-staffing levels and the quality of care in hospitals [J].
Needleman, J ;
Buerhaus, P ;
Mattke, S ;
Stewart, M ;
Zelevinsky, K .
NEW ENGLAND JOURNAL OF MEDICINE, 2002, 346 (22) :1715-1722
[29]   Modelling border-line tolerated conditions of use (BTCU) and associated risks [J].
Polet, P ;
Vanderhaegen, F ;
Amalberti, R .
SAFETY SCIENCE, 2003, 41 (2-3) :111-136
[30]  
Schneider J., 1999, HOSP PHARM, V34, P1044