A compact planar positron imaging system

被引:55
作者
Uchida, H [1 ]
Okamoto, T [1 ]
Ohmura, T [1 ]
Shimizu, K [1 ]
Satoh, N [1 ]
Koike, T [1 ]
Yamashita, T [1 ]
机构
[1] Hamamatsu Photon KK, Res Grp 5, Shizuoka 4348601, Japan
关键词
positron imaging device; planar imaging; PS-PMT;
D O I
10.1016/j.nima.2003.08.165
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A planar imaging system for positron-emitting radiotracers was developed and its physical performance was evaluated. The new device consists of two opposing planar detectors, each having 4 (columns) x 6 (rows) detect or units, and each unit composed of 10 x 10 arrays of 2 x 2 x 20 mm(3) pillars of Bi(4)Gey(3)O(12) (BGO) scintillators and a metal-packaged position-sensitive photomultiplier tube. The system is very compact and has a simple structure that allows versatile positioning of detectors in the horizontal or vertical mode. Focal plane images are constructed from coincidence data collected by opposing planar detectors. The system provides good quality images even in a short period of time or with low tracer activity. Spatial resolutions of 1.6-2.1 mm FWHM were achieved for the entire field of view. The sensitivity was 107 cps/kBq/ml measured in a uniform flat phantom. In this study, the capability of the new system was demonstrated by the imaging of radiotracer distribution in rats and plants. (C) 2003 Elsevier B.V. All rights reserved.
引用
收藏
页码:564 / 574
页数:11
相关论文
共 31 条
[1]   Three-dimensional performance of a small-diameter positron emission tomograph [J].
Bloomfield, PM ;
Myers, R ;
Hume, SP ;
Spinks, TJ ;
Lammertsma, AA ;
Jones, T .
PHYSICS IN MEDICINE AND BIOLOGY, 1997, 42 (02) :389-400
[2]  
BRASSE D, 1997, IEEE MED IM C REC 19
[3]  
BROWNELL GL, 1977, RECONSTRUCTION TOMOG, P293
[4]   Performance study of a 3D small animal PET scanner based on BaF2 crystals and a photo sensitive wire chamber [J].
Bruyndonckx, P ;
Liu, XA ;
Tavernier, S ;
Zhang, SP .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1997, 392 (1-3) :407-413
[5]   EMISSION COMPUTER-ASSISTED TOMOGRAPHY WITH SINGLE-PHOTON AND POSITRON-ANNIHILATION PHOTON EMITTERS [J].
BUDINGER, TF ;
DERENZO, SE ;
GULLBERG, GT ;
GREENBERG, WL ;
HUESMAN, RH .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1977, 1 (01) :131-145
[6]   Real-time [11C]methionine translocation in barley in relation to mugineic acid phytosiderophore biosynthesis [J].
Bughio, N ;
Nakanishi, H ;
Kiyomiya, S ;
Matsuhashi, S ;
Ishioka, NS ;
Watanabe, S ;
Uchida, H ;
Tsuji, A ;
Osa, A ;
Kume, T ;
Hashimoto, S ;
Sekine, T ;
Mori, S .
PLANTA, 2001, 213 (05) :708-715
[7]  
Chatziioannou AF, 1999, J NUCL MED, V40, P1164
[8]   MicroPET: A high resolution PET scanner for imaging small animals [J].
Cherry, SR ;
Shao, Y ;
Silverman, RW ;
Meadors, K ;
Siegel, S ;
Chatziioannou, A ;
Young, JW ;
Jones, WF ;
Moyers, JC ;
Newport, D ;
Boutefnouchet, A ;
Farquhar, TH ;
Andreaco, M ;
Paulus, MJ ;
Binkley, DM ;
Nutt, R ;
Phelps, ME .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1997, 44 (03) :1161-1166
[9]   A small animal PET prototype based on LSO crystals read out by avalanche photodiodes [J].
Fries, O ;
Bradbury, SM ;
Gebauer, J ;
Holl, I ;
Lorenz, E ;
Renker, D ;
Ziegler, SI .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1997, 387 (1-2) :220-224
[10]  
GUERRA AD, 1998, IEEE T NUCL SCI, V45, P3105