Performance study of a 3D small animal PET scanner based on BaF2 crystals and a photo sensitive wire chamber

被引:36
作者
Bruyndonckx, P
Liu, XA
Tavernier, S
Zhang, SP
机构
[1] Vrije Universiteit Brussel, IIHE, B-1050 Brussel
关键词
D O I
10.1016/S0168-9002(97)00281-7
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A 3D small animal PET scanner, designed and built at the University of Brussel, has been operational since the beginning of 1996. The scanner has a transaxial field of view (FOV) of 110 mm and an axial FOV of 52 mm. The absolute sensitivity is around 35,000 coincidences s(-1) MBq(-1) for a point source at the center of the scanner and the time resolution is 29 ns FWHM. To measure the achievable spatial resolution, a thin Na-22 source was scanned at various distances from the scanner axis. The resolution in a reconstructed image for a source close to the center is 3.0 mm FWHM. This figure can be improved at the expense of the sensitivity by lowering the voltage on the anodes in the wire chamber. Finally, to assess the overall image quality, scans were made of a cylindrical phantom with holes of varying diameter.
引用
收藏
页码:407 / 413
页数:7
相关论文
共 10 条
  • [1] COUPLING OF A BAF2 SCINTILLATOR TO A TMAE PHOTO-CATHODE AND A LOW-PRESSURE WIRE CHAMBER
    ANDERSON, DF
    BOUCLIER, R
    CHARPAK, G
    MAJEWSKI, S
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH, 1983, 217 (1-2): : 217 - 223
  • [2] THE DESIGN AND PHYSICAL CHARACTERISTICS OF A SMALL ANIMAL POSITRON EMISSION TOMOGRAPH
    BLOOMFIELD, PM
    RAJESWARAN, S
    SPINKS, TJ
    HUME, SP
    MYERS, R
    ASHWORTH, S
    CLIFFORD, KM
    JONES, WF
    BYARS, LG
    YOUNG, J
    ANDREACO, M
    WILLIAMS, CW
    LAMMERTSMA, AA
    JONES, T
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1995, 40 (06) : 1105 - 1126
  • [3] ON THE LOW-PRESSURE OPERATION OF MULTISTEP AVALANCHE CHAMBERS
    BRESKIN, A
    CHARPAK, G
    MAJEWSKI, S
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1984, 220 (2-3) : 349 - 355
  • [4] BAF2 SCINTILLATORS WITH WIRE CHAMBER READOUT FOR POSITRON EMISSION TOMOGRAPHY
    BRUYNDONCKX, P
    DEBRUYNE, J
    ETIENNE, L
    GRUWE, M
    GUERARD, B
    TAVERNIER, S
    SHUPING, Z
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1991, 310 (1-2) : 107 - 115
  • [5] STUDY OF THE SPATIAL-RESOLUTION IN GAMMA-DETECTORS USING PHOTOSENSITIVE WIRE CHAMBERS IN VIEW OF THEIR USE AS HIGH-RESOLUTION PET SCANNERS
    BRUYNDONCKX, P
    SHUPING, Z
    TAVERNIER, S
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1992, 323 (1-2) : 54 - 59
  • [6] CHERRY SR, 1995 IEEE NSS MIC C, P1766
  • [7] DAUBEWITHERSPOON ME, 1987, J NUCL MED, V28, P1717
  • [8] A FACTORIZATION METHOD FOR THE 3D X-RAY TRANSFORM
    DEFRISE, M
    [J]. INVERSE PROBLEMS, 1995, 11 (05) : 983 - 994
  • [9] LECOMTE R, 1995 IEEE NSS MIC C, P1353
  • [10] LIU XA, 1996, NUCL INSTRUM METH A, V382, P589