Reproducibility of nerve fiber layer thickness measurements by use of optical coherence tomography

被引:266
作者
Blumenthal, EZ
Williams, JM
Weinreb, RN
Girkin, CA
Berry, CC
Zangwill, LM
机构
[1] Univ Calif San Diego, Claucoma Ctr 0946, La Jolla, CA 92093 USA
[2] Univ Calif San Diego, Dept Ophthalmol, Diagnost Imaging Lab, La Jolla, CA 92093 USA
[3] Univ Calif San Diego, Dept Family & Prevent Med, La Jolla, CA 92093 USA
关键词
D O I
10.1016/S0161-6420(00)00341-9
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Objective: To evaluate the reproducibility of optical coherence tomograph (OCT) retinal nerve fiber layer (RNFL) measurements in normal and glaucomatous eyes by means of the commercially available OCT 2000 instrument (Humphrey Systems, Dublin, CA). Design: Prospective instrument validation study. Participants: One eye each from 10 normal subjects and 10 glaucoma patients. Methods: Twenty subjects underwent a total of eight scanning sessions during two independent visits. in each session, five circular scans centered on the optic nerve head were performed. The first two sessions were performed by two experienced technicians. Followed by a 30-minute break, a third and a fourth session was completed by the same technicians. This sequence was duplicated on a second visit. Intrasession, intersession, intervisit, and interoperator reproducibility of quadrant and global RNFL measurements were calculated by use of a components of variance model. Main Outcome Measures: RNFL thickness. Results: The coefficient of variation for the mean RNFL thickness was significantly smaller (P = 0.02) in normal eyes (6.9%) than in glaucoma eyes (11.8%). The estimated root mean squared error based on the statistical model using three scans per patient was 5.8 and 8.0 mum for normal and glaucoma eyes, respectively. A components of variance model showed most of the variance (79%) to be due to differences between patients. Only a modest contribution to variability was found for session (1%), visit (5%), and operator (2%). Conclusion: With the commercially available OCT, our results indicate that the RNFL measurements are reproducible for both normal and glaucomatous eyes. (C) 2000 by the American Academy of Ophthalmology.
引用
收藏
页码:2278 / 2282
页数:5
相关论文
共 18 条
[1]  
Baumann M, 1998, OPHTHALMIC SURG LAS, V29, P280
[2]  
Bowd C, 2000, ARCH OPHTHALMOL-CHIC, V118, P22
[3]   New technology for high-speed and high-resolution optical coherence tomography [J].
Fujimoto, JG ;
Bouma, B ;
Tearney, GJ ;
Boppart, SA ;
Pitris, C ;
Southern, JF ;
Brezinski, ME .
ADVANCES IN OPTICAL BIOPSY AND OPTICAL MAMMOGRAPHY, 1998, 838 :95-107
[4]  
Gurses-Ozden R, 1999, J GLAUCOMA, V8, P238
[5]   Topography of diabetic macular edema with optical coherence tomography [J].
Hee, MR ;
Puliafito, CA ;
Duker, JS ;
Reichel, E ;
Coker, JG ;
Wilkins, JR ;
Schuman, JS ;
Swanson, EA ;
Fujimoto, JG .
OPHTHALMOLOGY, 1998, 105 (02) :360-370
[6]   OPTICAL COHERENCE TOMOGRAPHY OF MACULAR HOLES [J].
HEE, MR ;
PULIAFITO, CA ;
WONG, C ;
DUKER, JS ;
REICHEL, E ;
SCHUMAN, JS ;
SWANSON, EA ;
FUJIMOTO, JG .
OPHTHALMOLOGY, 1995, 102 (05) :748-756
[7]   OPTICAL COHERENCE TOMOGRAPHY OF THE HUMAN RETINA [J].
HEE, MR ;
IZATT, JA ;
SWANSON, EA ;
HUANG, D ;
SCHUMAN, JS ;
LIN, CP ;
PULIAFITO, CA ;
FUJIMOTO, JG .
ARCHIVES OF OPHTHALMOLOGY, 1995, 113 (03) :325-332
[8]   RETINAL NERVE-FIBER LAYER PHOTOGRAPHY IN GLAUCOMATOUS PATIENTS [J].
HITCHINGS, RA ;
POINOOSAWMY, D ;
POPLAR, N ;
SHETH, GP .
EYE-TRANSACTIONS OF THE OPHTHALMOLOGICAL SOCIETIES OF THE UNITED KINGDOM, 1987, 1 :621-625
[9]   OPTICAL COHERENCE TOMOGRAPHY [J].
HUANG, D ;
SWANSON, EA ;
LIN, CP ;
SCHUMAN, JS ;
STINSON, WG ;
CHANG, W ;
HEE, MR ;
FLOTTE, T ;
GREGORY, K ;
PULIAFITO, CA ;
FUJIMOTO, JG .
SCIENCE, 1991, 254 (5035) :1178-1181
[10]   Heidelberg retina tomography and optical coherence tomography in normal, ocular-hypertensive, and glaucomatous eyes [J].
Mistlberger, A ;
Liebmann, JM ;
Greenfield, DS ;
Pons, ME ;
Hoh, ST ;
Ishikawa, H ;
Ritch, R .
OPHTHALMOLOGY, 1999, 106 (10) :2027-2032