EOS Micro-dose Protocol: First Full-spine Radiation Dose Measurements in Anthropomorphic Phantoms and Comparisons with EOS Standard-dose and Conventional Digital Radiology

被引:28
作者
Pedersen, Peter Heide [1 ]
Petersen, Asger Greval [2 ]
Ostgaard, Svend Erik [1 ]
Tvedebrink, Torben [3 ]
Eiskjaer, Soren P. [1 ]
机构
[1] Aalborg Univ Hosp, Dept Orthoped Surg, Hobrovej 18-22, DK-9000 Aalborg, Denmark
[2] Reg Nordjylland, Rontgenfysik, Denmark
[3] Aalborg Univ, Dept Math Sci, Aalborg, Denmark
关键词
anthropomorphic phantoms; cancer risk; cumulative dose; effective dose; EOS stereo-radiography; full-spine radiography; low-dose; micro-dose; organ dose; phantom dosimetry; radiation dose; scoliosis; slot-scanning system; X-RAY EXAMINATIONS; SCOLIOSIS; MORTALITY; CANCER;
D O I
10.1097/BRS.0000000000002696
中图分类号
R74 [神经病学与精神病学];
学科分类号
100204 [神经病学];
摘要
Study Design. A comparative study of radiation dose measured in anthropomorphic phantoms. Objectives. The aim of this study was to first report the first organ dose and effective dose measurements in anthropomorphic phantoms using the new EOS imaging micro-dose protocol in full-spine examinations, and to compare these measurements of radiation dose to measurements in the EOS standard-dose protocol and conventional digital radiology (CR). Summary of Background Data. Few studies evaluating organ dose and effective dose for the EOS low-dose scanner exist, and mainly for the standard-dose protocol. To the best of our knowledge, no studies of effective dose based on anthropomorphic phantom measurements exist for the new micro-dose protocol. Methods. Two anthropomorphic phantoms, representing a 5year-old (pediatric) and a 15-year-old (adolescent). The phantoms were exposed to EOS micro-dose and standard-dose protocols during full-spine imaging. Additionally, CR in scoliosis settings was performed. For all modalities, organ doses were measured and effective doses were calculated using thermoluminescent dosimeters. Results. We found a 17-fold reduction (94%) of effective dose in micro-dose protocol compared with our CR system in the adolescent phantom. Micro-dose versus standard-dose protocol, showed a 6-fold reduction (83%), and for standard-dose versus our CR system a 2.8-fold reduction (64%) reduction of effective dose was observed. For the pediatric phantom, a 5-fold reduction (81%) of effective dose in micro-dose protocol compared to our CR system was observed. Micro-dose versus standard-dose protocol, showed a seven-fold (86%) reduction. However, we observed an increase in absorbed dose of 38% when comparing the EOS standard-dose protocol with our CR system. Conclusion. The EOS imaging micro-dose option exposes patients to lower radiation doses than any currently available modality for full-spine examination. Expected reduction of dose was established for the adolescent phantom when comparing CR and standard-dose protocol. However, no reduction of effective dose with EOS standard-dose protocol compared to our reference CR system was observed in the pediatric phantom.
引用
收藏
页码:E1313 / E1321
页数:9
相关论文
共 26 条
[1]
[Anonymous], 2015, EUR SPINE J, V24, pS740, DOI DOI 10.1007/S00586-015-4130-8
[2]
Effective dose reduction in spine radiographic imaging by choosing the less radiation-sensitive side of the body [J].
Ben-Shlomo, Avi ;
Bartal, Gabriel ;
Mosseri, Morris ;
Avraham, Boaz ;
Leitner, Yosef ;
Shabat, Shay .
SPINE JOURNAL, 2016, 16 (04) :558-563
[3]
Body W, 2013, ATOM DOSIMETRY PHANT
[4]
Reduction of radiation risks in patients undergoing some X-ray examinations by using optimal projections: A Monte Carlo program-based mathematical calculation [J].
Chaparian, A. ;
Kanani, A. ;
Baghbanian, M. .
JOURNAL OF MEDICAL PHYSICS, 2014, 39 (01) :32-39
[5]
Organ radiation exposure with EOS: GATE simulations versus TLD measurements [J].
Clavel, A. H. ;
Thevenard-Berger, P. ;
Verdun, F. R. ;
Letang, J. M. ;
Darbon, A. .
MEDICAL IMAGING 2016: PHYSICS OF MEDICAL IMAGING, 2016, 9783
[6]
Occupational and patient exposure as well as image quality for full spine examinations with the EOS imaging system [J].
Damet, J. ;
Fournier, P. ;
Monnin, P. ;
Sans-Merce, M. ;
Ceroni, D. ;
Zand, T. ;
Verdun, F. R. ;
Baechler, S. .
MEDICAL PHYSICS, 2014, 41 (06)
[7]
Diagnostic Imaging of Spinal Deformities Reducing Patients Radiation Dose With a New Slot-Scanning X-ray Imager [J].
Deschenes, Sylvain ;
Charron, Guy ;
Beaudoin, Gilles ;
Labelle, Hubert ;
Dubois, Josee ;
Miron, Marie-Claude ;
Parent, Stefan .
SPINE, 2010, 35 (09) :989-994
[8]
Doody MM, 2000, SPINE, V25, P2052
[9]
EVALUATION OF ORGAN DOSES IN CT EXAMINATIONS WITH AN INFANT ANTHROPOMORPHIC PHANTOM [J].
Fujii, K. ;
Akahane, K. ;
Miyazaki, O. ;
Horiuchi, T. ;
Shimada, A. ;
Nagmatsu, H. ;
Yamauchi, M. ;
Yamauchi-Kawaura, C. ;
Kawasaki, T. .
RADIATION PROTECTION DOSIMETRY, 2011, 147 (1-2) :151-155
[10]
Comparison of dose from radiological examination for scoliosis in children among two pediatric hospitals by monte carlo simulation [J].
Gialousis, G. ;
Yiakoumakis, E. N. ;
Makri, T. K. ;
Papadoupoulou, D. ;
Karlatira, M. ;
Karaiskos, P. ;
Papaodysseas, S. ;
Evlogias, N. ;
Dimitriou, P. A. ;
Georgiou, E. K. .
HEALTH PHYSICS, 2008, 94 (05) :471-478