Organ radiation exposure with EOS: GATE simulations versus TLD measurements

被引:3
作者
Clavel, A. H. [1 ,2 ,3 ]
Thevenard-Berger, P. [3 ,4 ]
Verdun, F. R. [1 ]
Letang, J. M. [2 ]
Darbon, A. [3 ]
机构
[1] CHU Vaudois, Inst Radiat Phys, Rue Grand Pre 1, CH-1007 Lausanne, Switzerland
[2] Univ Lyon, CNRS, UMR5220, CREATIS,Inserm,U1044, Lyon, France
[3] EOS Imaging, 10 Rue Mercoeur, F-75011 Paris, France
[4] Ecole Cent Elect, 37 Quai Grenelle, F-75015 Paris, France
来源
MEDICAL IMAGING 2016: PHYSICS OF MEDICAL IMAGING | 2016年 / 9783卷
关键词
Monte-Carlo; GAPE; radiation exposure; X-ray; organ dosimetry; EOS; phantom; GAS MICROSTRIP DETECTOR;
D O I
10.1117/12.2217097
中图分类号
O43 [光学];
学科分类号
070207 [光学];
摘要
EOS (R) is an innovative X-ray imaging system allowing the acquisition of two simultaneous images of a patient in the standing position, during the vertical scan of two orthogonal fan beams. This study aimed to compute organs radiation exposure to a patient, in the particular geometry of this system. Two different positions of the patient in the machine were studied, corresponding to postero-anterior plus left lateral projections (PA-LLAT) and antero-posterior plus right lateral projections (AP-RLAT). To achieve this goal, a Monte-Carlo simulation was developed based on a GAPE environment. To model the physical properties of the patient, a computational phantom was produced based on computed tomography scan data of an anthropomorphic phantom. The simulations provided several organs doses, which were compared to previously published dose results measured with Thermo Luminescent Detectors (TLD) in the same conditions and with the same phantom. The simulation results showed a good agreement with measured doses at the TLD locations, for both AP-RLAT and PA-LLAT projections. This study also showed that the organ dose assessed only from a sample of locations, rather than considering the whole organ, introduced significant bias, depending on organs and projections.
引用
收藏
页数:7
相关论文
共 13 条
[1]
Dynamic Partitioning of GATE Monte-Carlo Simulations on EGEE [J].
Camarasu-Pop, Sorina ;
Glatard, Tristan ;
Moscicki, Jakub T. ;
Benoit-Cattin, Hugues ;
Sarrut, David .
JOURNAL OF GRID COMPUTING, 2010, 8 (02) :241-259
[2]
Characterization of scatter in cone-beam CT breast imaging: Comparison of experimental measurements and Monte Carlo simulation [J].
Chen, Yu ;
Liu, Bob ;
O'Connor, J. Michael ;
Didier, Clay S. ;
Glick, Stephen J. .
MEDICAL PHYSICS, 2009, 36 (03) :857-869
[3]
Clavel A. H., 2015, RADIAT PROT DOSIM
[4]
Occupational and patient exposure as well as image quality for full spine examinations with the EOS imaging system [J].
Damet, J. ;
Fournier, P. ;
Monnin, P. ;
Sans-Merce, M. ;
Ceroni, D. ;
Zand, T. ;
Verdun, F. R. ;
Baechler, S. .
MEDICAL PHYSICS, 2014, 41 (06)
[5]
Physical characteristics of a low-dose gas microstrip detector for orthopedic x-ray imaging [J].
Després, P ;
Beaudoin, G ;
Gravel, P ;
de Guise, JA .
MEDICAL PHYSICS, 2005, 32 (04) :1193-1204
[6]
Evaluation of a full-scale gas microstrip detector for low-dose X-ray imaging [J].
Després, P ;
Beaudoin, G ;
Gravel, P ;
de Guise, JA .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2005, 536 (1-2) :52-60
[7]
A Virtual Imaging Platform for Multi-Modality Medical Image Simulation [J].
Glatard, Tristan ;
Lartizien, Carole ;
Gibaud, Bernard ;
da Silva, Rafael Ferreira ;
Forestier, Germain ;
Cervenansky, Frederic ;
Alessandrini, Martino ;
Benoit-Cattin, Hugues ;
Bernard, Olivier ;
Camarasu-Pop, Sorina ;
Cerezo, Nadia ;
Clarysse, Patrick ;
Gaignard, Alban ;
Hugonnard, Patrick ;
Liebgott, Herve ;
Marache, Simon ;
Marion, Adrien ;
Montagnat, Johan ;
Tabary, Joachim ;
Friboulet, Denis .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2013, 32 (01) :110-118
[8]
Golikov V., 2015, P OXMI 2015 C GUTH S
[9]
SpekCalc: a program to calculate photon spectra from tungsten anode x-ray tubes [J].
Poludniowski, G. ;
Landry, G. ;
DeBlois, F. ;
Evans, P. M. ;
Verhaegen, F. .
PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (19) :N433-N438
[10]
A review of the use and potential of the GATE Monte Carlo simulation code for radiation therapy and dosimetry applications [J].
Sarrut, David ;
Bardies, Manuel ;
Boussion, Nicolas ;
Freud, Nicolas ;
Jan, Sebastien ;
Letang, Jean-Michel ;
Loudos, George ;
Maigne, Lydia ;
Marcatili, Sara ;
Mauxion, Thibault ;
Papadimitroulas, Panagiotis ;
Perrot, Yann ;
Pietrzyk, Uwe ;
Robert, Charlotte ;
Schaart, Dennis R. ;
Visvikis, Dimitris ;
Buvat, Irene .
MEDICAL PHYSICS, 2014, 41 (06)