A mechanical comparison of the locking compression plate (LCP) and the low contact-dynamic compression plate (DCP) in an osteoporotic bone model

被引:80
作者
Snow, Martyn [1 ]
Thompson, Graham [1 ]
Turner, Phillip G. [2 ]
机构
[1] Univ Manchester, Dept Mech Engn, Manchester, Lancs, England
[2] Stockport NHS Fdn Trust, Stepping Hill Hosp, Dept Orthopaed, Stockport, Cheshire, England
关键词
locking compression plates; dynamic compression plates; osteoporotic; mechanical testing; synthetic;
D O I
10.1097/BOT.0b013e318160c84c
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Objective: To determine if locking compression plates (LCP) are mechanically advantageous compared to low-contact dynamic compression plates (DCP) when used as a bridging plate in a synthetic model of osteoporotic bone. Methods: Five synthetic bars (Synbone Osteoporotic bone) were initially tested in compression and the Young's modulus determined. It was found to be comparable to that of tibial cancellous bone in an 80-year-old woman; thus, the synthetic bars were deemed usable to simulate some properties of osteoporotic bone. Six bars were then instrumented with an 8-hole narrow large fragment DCP with six 4.5-mm cortical screws (placed in holes 1-3 and holes 6-8). Six bars were instrumented with a narrow 8-hole LCP using four 5-mm locking screws (placed in the 1st, 3rd, 6th, and 8th holes). In a third group, 6 synbone bars were instrumented with a narrow 8-hole large fragment DCR Cortical screws were placed in holes 2-3 and holes 6-7. In holes 1 and 8, two 6.5 fully threaded cancellous screws were inserted. A 1-cm osteotomy was created in the Synbone at the center of each plate to represent a comminuted fracture. Initially, quasi-static testing was carried out on all specimens in compression to a maximum load of 450 N. Then 4-point bend tests were carried out in two planes (0 degrees and 90 degrees) with the maximum bending moment at 3.5 Nm. Finally, torsional testing was done to a maximum load of 3.5 Nm. The specimens were then cycled in axial compression 350 N at 5 Hz for 30,000 cycles. The static nondestructive tests were repeated. The slope of the load deformation curve indicated the relative stiffness of the construct. The slopes were determined pre- and postcycling and the loss of stiffness in each group compared. Statistical analysis was carried out using the paired t-test. The specimens were then loaded to failure in compression. Results: There was no statistical difference in the stiffness of the LCP or in the osteotomy gap postcycling. All specimens in the DCP groups failed initial static testing in axial compression. No fatigue testing could be undertaken in this group. Conclusions: In a synthetic model, the LCP was mechanically superior to the DCP when used as a bridging plate and tested in axial compression.
引用
收藏
页码:121 / 125
页数:5
相关论文
共 26 条
[1]   An in vitro biomechanical study of bone plate and interlocking nail in a canine diaphyseal femoral fracture model [J].
Bernarde, A ;
Diop, A ;
Maurel, N ;
Viguier, E .
VETERINARY SURGERY, 2001, 30 (05) :397-408
[2]  
*BRIT ORTH ASS, 2003, CAR FRAG FRACT
[3]   MECHANICAL PROPERTY DISTRIBUTIONS IN THE CANCELLOUS BONE OF THE HUMAN PROXIMAL FEMUR [J].
BROWN, TD ;
FERGUSON, AB .
ACTA ORTHOPAEDICA SCANDINAVICA, 1980, 51 (03) :429-437
[4]  
BURNSTEIN AH, 1976, J BONE JOINT SURG AM, V58, P20
[5]   Biomechanical considerations of fracture treatment and bone quality maintenance in elderly patients and patients with osteoporosis [J].
Chao, EYS ;
Inoue, N ;
Koo, TKK ;
Kim, YH .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2004, (425) :12-25
[6]   Age variations in the properties of human tibial trabecular bone [J].
Ding, M ;
Dalstra, M ;
Danielsen, CC ;
Kabel, J ;
Hvid, I ;
Linde, F .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1997, 79B (06) :995-1002
[7]  
DING M, 2000, ACTA ORTHOP SCAND, V292, P71
[8]   Fixation of diaphyseal fractures with a segmental defect: A biomechanical comparison of locked and conventional plating techniques [J].
Fulkerson, E ;
Egol, KA ;
Kubiak, EN ;
Liporace, F ;
Kummer, FJ ;
Koval, KJ .
JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE, 2006, 60 (04) :830-835
[9]   Fixation of periprosthetic femoral shaft fractures associated with cemented femoral stems - A biomechanical comparison of locked plating and conventional cable plates [J].
Fulkerson, E ;
Koval, K ;
Preston, CF ;
Iesaka, K ;
Kummer, FJ ;
Egol, KA .
JOURNAL OF ORTHOPAEDIC TRAUMA, 2006, 20 (02) :89-93
[10]  
Gardner Michael J, 2005, J Orthop Trauma, V19, P597, DOI 10.1097/01.bot.0000174033.30054.5f