Long-term results of porcine bioprostheses in the tricuspid position

被引:10
作者
Kuwaki, K [1 ]
Komatsu, K [1 ]
Morishita, K [1 ]
Tsukamoto, M [1 ]
Abe, T [1 ]
机构
[1] Sapporo Med Univ, Sch Med, Dept Cardiovasc & Thorac Surg, Chuo Ku, Sapporo, Hokkaido 060, Japan
来源
SURGERY TODAY-THE JAPANESE JOURNAL OF SURGERY | 1998年 / 28卷 / 06期
关键词
tricuspid valve replacement; porcine bioprosthesis;
D O I
10.1007/s005950050191
中图分类号
R61 [外科手术学];
学科分类号
摘要
From January 1976 to September 1994, 26 patients underwent a tricuspid valve replacement (TVR) with porcine bioprostheses (Hancock II in 19 patients, Carpentier-Edwards in 3, Hancock standard in 2, Medtronic-Intact in 1, and Liotta in 1), The mean age of these patients was 38.4 Sears (range, 9 to 70 years). The mean follow-up period was 6.7 years (range, 1.8 to 16.5 years). There was a complete follow-up in 100%. There were two hospital deaths and three late deaths. The actuarial survival rate at 16 years was 78.7% +/- 8.6%. Five patients developed structural deterioration. Four of these 5 patients underwent re-TVR, The actuarial freedom rate from structural deterioration at 10 years was 47.1% +/- 19.1%. There were no instances of thromboembolism and prosthetic valve endocarditis. Our experience indicated the major problem of the porcine bioprostheses in the tricuspid position thus to be valve durability: however, the long-term actuarial survival rate was satisfactory, including no operative morality with re-TVR due to tricuspid prosthetic structural deterioration. We therefore conclude that the choice of a porcine bioprosthesis in the tricuspid position should be considered not only for patients with isolated tricuspid valve disease, but also for the patients who have not previously received mechanical valves in the mitral or the aortic valve position.
引用
收藏
页码:599 / 603
页数:5
相关论文
共 19 条
[1]  
BOSKOVIC D, 1986, J THORAC CARDIOV SUR, V91, P1
[2]  
BOURDILLON PDV, 1976, BRIT HEART J, V38, P1149
[3]   COMPARISON OF LATE (62 TO 140 MONTHS) DEGENERATIVE CHANGES IN SIMULTANEOUSLY IMPLANTED AND EXPLANTED PORCINE (HANCOCK) BIOPROSTHESES IN THE TRICUSPID AND MITRAL-VALVE POSITIONS IN 6 PATIENTS [J].
COHEN, SR ;
SILVER, MA ;
MCINTOSH, CL ;
ROBERTS, WC .
AMERICAN JOURNAL OF CARDIOLOGY, 1984, 53 (11) :1599-1602
[4]   LONG-TERM RESULTS OF IONESCU-SHILEY VALVE IN THE TRICUSPID POSITION [J].
ENG, J ;
RAVICHANDRAN, PS ;
KAY, PH ;
MURDAY, AJ .
ANNALS OF THORACIC SURGERY, 1991, 51 (02) :200-203
[5]   DETERMINANTS OF REOPERATION AFTER 960 VALVE REPLACEMENTS WITH CARPENTIER-EDWARDS PROSTHESES [J].
GLOWER, DD ;
WHITE, WD ;
HATTON, AC ;
SMITH, LR ;
YOUNG, WG ;
WOLFE, WG ;
LOWE, JE .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 1994, 107 (02) :381-393
[6]   IN-HOSPITAL AND LONG-TERM OUTCOME AFTER PORCINE TRICUSPID-VALVE REPLACEMENT [J].
GLOWER, DD ;
WHITE, WD ;
SMITH, LR ;
YOUNG, WG ;
OLDHAM, HN ;
WOLFE, WG ;
LOWE, JE .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 1995, 109 (05) :877-884
[7]  
GUERRA F, 1990, J THORAC CARDIOV SUR, V99, P838
[8]  
HOROWITZ MS, 1978, AM J CARDIOL, V42, P691, DOI 10.1016/0002-9149(78)90647-1
[9]   TRICUSPID-VALVE REPLACEMENT - FACTORS INFLUENCING EARLY AND LATE MORTALITY [J].
KAUL, TK ;
MERCER, JL .
THORACIC AND CARDIOVASCULAR SURGEON, 1990, 38 (04) :229-235
[10]   EXCELLENT DURABILITY OF THE HANCOCK PORCINE BIOPROSTHESIS IN THE TRICUSPID POSITION [J].
KAWACHI, Y ;
TOMINAGA, R ;
HISAHARA, M ;
NAKASHIMA, A ;
YASUI, H ;
TOKUNAGA, K .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 1992, 104 (06) :1561-1566