Coronary plaque imaging with 256-slice multidetector computed tomography: interobserver variability of volumetric lesion parameters with semiautomatic plaque analysis software

被引:28
作者
Klass, Oliver [1 ]
Kleinhans, Susanne [1 ]
Walker, Matthew J. [2 ]
Olszewski, Mark [2 ]
Feuerlein, Sebastian [1 ]
Juchems, Markus [1 ]
Hoffmann, Martin H. K. [1 ]
机构
[1] Univ Hosp Ulm, Dept Diagnost & Intervent Radiol, D-89075 Ulm, Germany
[2] CT Clin Sci, Philips Healthcare, Cleveland, OH USA
关键词
Atherosclerosis; Plaque analysis; Multidetector computed tomography; Noninvasive imaging; Interobserver variability; 256-slice MDCT; Gaussian mixture model; INTRAVASCULAR ULTRASOUND; ATHEROSCLEROTIC PLAQUE; ANGIOGRAPHY; CT; ACCURACY; SCORES; HEART; MODEL;
D O I
10.1007/s10554-010-9614-3
中图分类号
R5 [内科学];
学科分类号
100201 [内科学];
摘要
The purpose of this study was to evaluate the potential clinical value of coronary plaque imaging with a new generation CT scanner and the interobserver variability of coronary plaque assessment with a new semiautomatic plaque analysis application. Thirty-five isolated plaques of the left anterior descending coronary artery from 35 patients were evaluated with a new semiautomatic plaque analysis application. All patients were scanned with a 256-slice MDCT scanner (Brilliance iCT, Philips Healthcare, Cleveland OH, USA). Two independent observers evaluated lesion volume, maximum plaque burden, lesion CT number mean and standard deviation, and relative lesion composition. We found 10 noncalcified, 16 mixed, and 9 calcified lesions in our study cohort. Relative interobserver bias and variability for lesion volume were -37%, -13%, -49%, -44% and 28%, 16%, 37%, and 90% for all, noncalcified, mixed, and calcified lesions, respectively. Absolute interobserver bias and variability for relative lesion composition were 1.2%, 0.5%, 1.5%, 1.3% and 3.3%, 4.5%, 7.0%, and 4.4% for all, noncalcified, mixed, and calcified lesions, respectively. While mixed and calcified lesions demonstrated a high degree of lesion volume interobserver variability, noncalcified lesions had a lower degree of lesion volume interobserver variability. In addition, relative noncalcified lesion composition had a very low interobserver variability. Therefore, there may a role for MDCT in serial noncalcified plaque assessment with semiautomatic analysis software.
引用
收藏
页码:711 / 720
页数:10
相关论文
共 21 条
[1]
Detection of calcified and noncalcified coronary atherosclerotic plaque by contrast-enhanced, submillimeter multidetector spiral computed tomography - A segment-based comparison with intravascular ultrasound [J].
Achenbach, S ;
Moselewski, F ;
Ropers, D ;
Ferencik, M ;
Hoffmann, U ;
MacNeill, B ;
Pohle, K ;
Baum, U ;
Anders, K ;
Jang, I ;
Daniel, WG ;
Brady, TJ .
CIRCULATION, 2004, 109 (01) :14-17
[2]
[Anonymous], 2004, NRPBPE12004
[3]
Characterization of coronary atherosclerosis by dual-source computed tomography and HU-based color mapping: a pilot study [J].
Brodoefel, H. ;
Reimann, A. ;
Heuschmid, M. ;
Tsiflikas, I. ;
Kopp, A. F. ;
Schroeder, S. ;
Claussen, C. D. ;
Clouse, M. E. ;
Burgstahler, C. .
EUROPEAN RADIOLOGY, 2008, 18 (11) :2466-2474
[4]
Coronary Calcium Predicts Events Better With Absolute Calcium Scores Than Age-Sex-Race/Ethnicity Percentiles MESA (Multi-Ethnic Study of Atherosclerosis) [J].
Budoff, Matthew J. ;
Nasir, Khurram ;
McClelland, Robyn L. ;
Detrano, Robert ;
Wong, Nathan ;
Blumenthal, Roger S. ;
Kondos, George ;
Kronmal, Richard A. .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2009, 53 (04) :345-352
[5]
Computer-aided detection and evaluation of lipid-rich plaque on noncontrast cardiac CT [J].
Dey, Damini ;
Callister, Tracy ;
Slomka, Piotr ;
Aboul-Enein, Fatma ;
Nishina, Hidetaka ;
Kang, Xingping ;
Gransar, Heidi ;
Wong, Nathan D. ;
Miranda-Peats, Romalisa ;
Hayes, Sean ;
Friedman, John D. ;
Berman, Daniel S. .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2006, 186 (06) :S407-S413
[6]
Duda R.O., 2000, PATTERN CLASSIFICATI, V2nd
[7]
Automatic model-based segmentation of the heart in CT images [J].
Ecabert, Olivier ;
Peters, Jochen ;
Schramm, Hauke ;
Lorenz, Cristian ;
von Berg, Jens ;
Walker, Matthew J. ;
Vembar, Mani ;
Olszewski, Mark E. ;
Subramanyan, Krishna ;
Lavi, Guy ;
Weese, Juergen .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2008, 27 (09) :1189-1201
[8]
Is there a relation between non-calcifying coronary plaques and acute coronary syndromes? A retrospective study using multislice computed tomography [J].
Feuchtner, Gudrun ;
Postel, Thomas ;
Weidinger, Franz ;
Frick, Matthias ;
Alber, Hannes ;
Dichtl, Wolfgang ;
Jodocy, Daniel ;
Mallouhi, Ammar ;
Pachinger, Otmar ;
Nedden, Dieter Zur ;
Friedrich, Guy J. .
CARDIOLOGY, 2008, 110 (04) :241-248
[9]
A method for calcium quantification by means of CT coronary angiography using 64-multidetector CT: very high correlation with agatston and volume scores [J].
Glodny, Bernhard ;
Helmel, Bettina ;
Trieb, Thomas ;
Schenk, Claudia ;
Taferner, Bernadette ;
Unterholzner, Verena ;
Strasak, Alexander ;
Petersen, Johannes .
EUROPEAN RADIOLOGY, 2009, 19 (07) :1661-1668
[10]
Reproducibility of volumetric intravascular ultrasound radiofrequency-based analysis of coronary plaque composition in vivo [J].
Hartmann, Marc ;
Mattern, Eline S. K. ;
Huisman, Jennifer ;
van Houwelingen, Gert K. ;
de Man, Frits H. A. F. ;
Stoel, Martin G. ;
Danse, Peter W. ;
Louwerenburg, Hans W. ;
von Birgelen, Clemens .