Outcomes of a multicenter trial of the Levitronix CentriMag ventricular assist system for short-term circulatory support

被引:152
作者
John, Ranjit [1 ]
Long, James W. [2 ]
Massey, H. Todd [3 ]
Griffith, Bartley P. [4 ]
Sun, Benjamin C. [5 ]
Tector, Alfred J. [6 ]
Frazier, O. Howard [7 ]
Joyce, Lyle D.
机构
[1] Univ Minnesota, Div Cardiothorac Surg, Med Ctr, Minneapolis, MN 55455 USA
[2] LDS Hosp, Salt Lake City, UT USA
[3] Univ Rochester, Strong Mem Hosp, Med Ctr, Rochester, NY 14642 USA
[4] Univ Maryland, Med Ctr, Baltimore, MD 21201 USA
[5] Ohio State Univ, Med Ctr, Columbus, OH 43210 USA
[6] St Lukes Hosp, Milwaukee, WI USA
[7] Texas Heart Inst, Houston, TX 77025 USA
关键词
EXTRACORPOREAL LIFE-SUPPORT; ACUTE CARDIOGENIC-SHOCK; BERLIN HEART EXCOR; CARDIAC-FAILURE; CLINICAL-EXPERIENCE; MECHANICAL ASSIST; DEVICE BRIDGE; TRANSPLANTATION; SURVIVAL; DECISION;
D O I
10.1016/j.jtcvs.2010.03.046
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objective: The Levitronix CentriMag (Levitronix LLC, Waltham, Mass) ventricular assist system is designed for temporary left, right, or biventricular support. Advantages include ease of use, excellent reliability, and low thrombosis risk,. which may allow wider application of short-term support and improved outcomes in patients with cardiogenic shock. This multi-institutional study evaluated safety, effectiveness, and outcomes of the CentriMag in patients with cardiogenic shock. Methods: Thirty-eight patients were supported at 7 centers. Patients included 12 after cardiotomy, 14 after myocardial infarction, and 12 with right ventricular failure after implantable left ventricular assist device placement. Devices were implanted in left (n = 8), right (n = 12), or biventricular (n = 18) configuration. Support was continued until recovery, transplantation, or implantation of long-term ventricular assist device. Results: Mean support duration for the entire cohort (n = 38) was 13 days (1-60 days), with 47% of patients (18/38) surviving 30 days after device removal. Mean CentriMag biventricular support (n = 18) duration was 15 days (1-60 days), with 44% (8/18) surviving at 30 days. Mean CentriMag right ventricular support with a commercially available left ventricular assist device (n = 12) duration was 14 days (1-29 days), with 58%(7/12) surviving at 30 days. Complications included bleeding (21%), infection (5%), respiratory failure (3%), hemolysis (5%), and neurologic dysfunction (11%). There were no CentriMag or pump failures. Conclusions: In this preliminary study, the CentriMag provided short-term support for patients with cardiogenic shock with a low incidence of device-related complications and no device failures. (J Thorac Cardiovasc Surg 2011;141:932-9)
引用
收藏
页码:932 / 939
页数:8
相关论文
共 22 条
[1]   Clinical performance with the levitronix centrimag short-term ventricular assist device [J].
De Robertis, F ;
Birks, EJ ;
Rogers, P ;
Dreyfus, G ;
Pepper, JR ;
Khaghani, A .
JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2006, 25 (02) :181-186
[2]   Bridge to decision using the levitronix CentriMag short-term ventricular assist device [J].
De Robertis, Fabio ;
Rogers, Paula ;
Amrani, Mohammed ;
Petrou, Mario ;
Pepper, John R. ;
Bahrami, Toufan ;
Dreyfus, Gilles D. ;
Khaghani, Asghar ;
Birks, Emma J. .
JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2008, 27 (05) :474-478
[3]   Medical progress - Implantable left ventricular assist devices [J].
Goldstein, DJ ;
Oz, MC ;
Rose, EA .
NEW ENGLAND JOURNAL OF MEDICINE, 1998, 339 (21) :1522-1533
[4]  
Gregoric ID, 2008, TEX HEART I J, V35, P125
[5]   Bridging patients after salvage from bridge to decision directly to transplant by means of prolonged support with the CentriMag short-term centrifugal pump [J].
Haj-Yahia, Saleem ;
Birks, Emma J. ;
Amrani, Mohammed ;
Petrou, Mario ;
Bahrami, Toufan ;
Dreyfus, Gilles ;
Khaghani, Asghar .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2009, 138 (01) :227-230
[6]   Outcome evaluation of the bridge to bridge concept in patients with cardiogenic shock [J].
Hoefer, Daniel ;
Ruttmann, Elfriede ;
Poelzl, Gerhard ;
Kilo, Juliane ;
Hoermann, Christoph ;
Margreiter, Raimund ;
Laufer, Guenther ;
Antretter, Herwig .
ANNALS OF THORACIC SURGERY, 2006, 82 (01) :28-34
[7]   Experience with the Levitronix CentriMag circulatory support system as a bridge to decision in patients with refractory acute cardiogenic shock and multisystem organ failure [J].
John, Ranjit ;
Liao, Kenneth ;
Lietz, Katherine ;
Kamdar, Forum ;
Colvin-Adams, Monica ;
Boyle, Andrew ;
Miller, Leslie ;
Joyce, Lyle .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2007, 134 (02) :351-358
[8]  
Kar B, 2006, TEX HEART I J, V33, P111
[9]   Implantation of a left ventricular assist device and the hub-and-spoke system in treating acute cardiogenic shock: Who survives? [J].
Kherani, AR ;
Cheema, FH ;
Oz, MC ;
Fal, JM ;
Morgan, JA ;
Topkara, VK ;
Wilson, DA ;
Vigilance, DW ;
Garrido, MJ ;
Naka, Y .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2003, 126 (05) :1634-1635
[10]   Levitronix CentriMag to Berlin Heart Excor: A "Bridge to Bridge" Solution in Refractory Cardiogenic Shock [J].
Loforte, Antonio ;
Potapov, Evgenij ;
Krabatsch, Thomas ;
Musci, Michele ;
Weng, Yuguo ;
Pasic, Miralem ;
Hetzer, Roland .
ASAIO JOURNAL, 2009, 55 (05) :465-468