Linear calibration function for optical oxygen sensors based on quenching of ruthenium fluorescence

被引:28
作者
Chuang, H [1 ]
Arnold, MA [1 ]
机构
[1] Univ Iowa, Dept Chem, Iowa City, IA 52242 USA
关键词
oxygen sensors; optical sensors; ruthenium fluorescence quenching; radioluminescent oxygen sensors;
D O I
10.1016/S0003-2670(98)00191-3
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A mathematical model is derived from the Stern-Volmer equation as a practical calibration function for various optical oxygen sensors based on quenching of ruthenium fluorescence. The model is simple and flexible, and is ideal for computing oxygen partial pressure from 0 to 760 Torr in gaseous samples, and 0 to 200 Torr in aqueous samples. Feasibility of this model for dissolved oxygen calibration is demonstrated by applying the model to data acquired from a set of three radioluminescence-based optical oxygen sensors. The model is applicable to various sensor designs and adequately accounts for stray radiation and other sources of non-quenchable light. (C) 1998 Elsevier Science B.V.
引用
收藏
页码:83 / 89
页数:7
相关论文
共 15 条
[1]   DETERMINATION OF OXYGEN CONCENTRATIONS BY LUMINESCENCE QUENCHING OF A POLYMER-IMMOBILIZED TRANSITION-METAL COMPLEX [J].
BACON, JR ;
DEMAS, JN .
ANALYTICAL CHEMISTRY, 1987, 59 (23) :2780-2785
[2]   PHOTOPHYSICS AND PHOTOCHEMISTRY OF OXYGEN SENSORS BASED ON LUMINESCENT TRANSITION-METAL COMPLEXES [J].
CARRAWAY, ER ;
DEMAS, JN ;
DEGRAFF, BA ;
BACON, JR .
ANALYTICAL CHEMISTRY, 1991, 63 (04) :337-342
[3]   Radioluminescent light source for optical oxygen sensors [J].
Chuang, H ;
Arnold, MA .
ANALYTICAL CHEMISTRY, 1997, 69 (10) :1899-1903
[4]  
CHUANG H, 1997, THESIS U IOWA
[5]   MODELING OF LUMINESCENCE QUENCHING-BASED SENSORS - COMPARISON OF MULTISITE AND NONLINEAR GAS SOLUBILITY MODELS [J].
DEMAS, JN ;
DEGRAFF, BA ;
XU, WY .
ANALYTICAL CHEMISTRY, 1995, 67 (08) :1377-1380
[6]   LUMINESCENCE QUENCHING BEHAVIOR OF AN OXYGEN SENSOR-BASED ON A RU(II) COMPLEX DISSOLVED IN POLYSTYRENE [J].
HARTMANN, P ;
LEINER, MJP ;
LIPPITSCH, ME .
ANALYTICAL CHEMISTRY, 1995, 67 (01) :88-93
[7]   OXYGEN-SENSITIVE LUMINESCENT MATERIALS BASED ON SILICONE-SOLUBLE RUTHENIUM DIIMINE COMPLEXES [J].
KLIMANT, I ;
WOLFBEIS, OS .
ANALYTICAL CHEMISTRY, 1995, 67 (18) :3160-3166
[8]   DUAL-ANALYTE FIBEROPTIC SENSOR FOR THE SIMULTANEOUS AND CONTINUOUS MEASUREMENT OF GLUCOSE AND OXYGEN [J].
LI, L ;
WALT, DR .
ANALYTICAL CHEMISTRY, 1995, 67 (20) :3746-3752
[9]   FIBRE-OPTIC OXYGEN SENSOR WITH THE FLUORESCENCE DECAY TIME AS THE INFORMATION CARRIER [J].
LIPPITSCH, ME ;
PUSTERHOFER, J ;
LEINER, MJP ;
WOLFBEIS, OS .
ANALYTICA CHIMICA ACTA, 1988, 205 (1-2) :1-6
[10]   OXYGEN OPTRODE FOR USE IN A FIBEROPTIC GLUCOSE BIOSENSOR [J].
MORENOBONDI, MC ;
WOLFBEIS, OS ;
LEINER, MJP ;
SCHAFFAR, BPH .
ANALYTICAL CHEMISTRY, 1990, 62 (21) :2377-2380