Ambulatory measurement of knee motion and physical activity: preliminary evaluation of a smart activity monitor

被引:38
作者
Huddleston, James
Alaiti, Amer
Goldvasser, Dov
Scarborough, Donna
Freiberg, Andrew
Rubash, Harry
Malchau, Henrik
Harris, William
Krebs, David
机构
[1] Harvard Univ, Sch Med, Harris Orthopaed Biomech & Biomat Lab, Massachusetts Gen Hosp, Boston, MA 02114 USA
[2] Harvard Univ, Massachusetts Gen Hosp, Sch Med, Biomot Lab,MGH Inst Hlth, Boston, MA 02129 USA
[3] Stanford Univ, Sch Med, Dept Orthopaed Surg, Stanford, CA 94305 USA
关键词
D O I
10.1186/1743-0003-3-21
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Background: There is currently a paucity of devices available for continuous, long-term monitoring of human joint motion. Non-invasive, inexpensive devices capable of recording human activity and joint motion have many applications for medical research. Such a device could be used to quantify range of motion outside the gait laboratory. The purpose of this study was to test the accuracy of the modified Intelligent Device for Energy Expenditure and Activity (IDEEA) in measuring knee flexion angles, to detect different physical activities, and to quantify how often healthy subjects use deep knee flexion in the ambulatory setting. Methods: We compared Biomotion Laboratory (BML) "gold standard" data to simultaneous IDEEA measures of knee motion and gait, step up/down, and stair descent in 5 healthy subjects. In addition, we used a series of choreographed physical activities outside the BML to confirm the IDEEA's ability to accurately measure 7 commonly-performed physical activities. Subjects then continued data collection during ordinary activities outside the gait laboratory. Results: Pooled correlations between the BML and IDEEA knee flexion angles were .97 +/- .03 for step up/down, .98 +/- .02 for stair descent, and .98 +/- .01 for gait. In the BML protocol, the IDEEA accurately identified gait, but was less accurate in identifying step up/down and stair descent. During sampling outside the BML, the IDEEA accurately detected walking, running, stair ascent, stair descent, standing, lying, and sitting. On average, subjects flexed their knees > 120 degrees for 0.17% of their data collection periods outside the BML. Conclusion: The modified IDEEA system is a useful clinical tool for evaluating knee motion and multiple physical activities in the ambulatory setting. These five healthy subjects rarely flexed their knees > 120 degrees.
引用
收藏
页数:10
相关论文
共 25 条
[1]   VALIDATION OF A SIMPLE MECHANICAL ACCELEROMETER (PEDOMETER) FOR THE ESTIMATION OF WALKING ACTIVITY [J].
BASSEY, EJ ;
DALLOSSO, HM ;
FENTEM, PH ;
IRVING, JM ;
PATRICK, JM .
EUROPEAN JOURNAL OF APPLIED PHYSIOLOGY AND OCCUPATIONAL PHYSIOLOGY, 1987, 56 (03) :323-330
[2]  
BOUTEN CV, 1994, MED SCI SPORT EXER, V26, P1516
[3]   Vestibular rehabilitation outcomes: velocity trajectory analysis of repeated bench stepping [J].
Goldvasser, D ;
McGibbon, CA ;
Krebs, DE .
CLINICAL NEUROPHYSIOLOGY, 2000, 111 (10) :1838-1842
[4]   The early results of high-flex total knee arthroplasty - A minimum of 2 years of follow-up [J].
Huang, HT ;
Su, AY ;
Wang, GJ .
JOURNAL OF ARTHROPLASTY, 2005, 20 (05) :674-679
[5]  
JANZ KF, 1994, MED SCI SPORT EXER, V26, P369
[6]   INVESTIGATION INTO THE MASS-DISTRIBUTION PROPERTIES OF THE HUMAN-BODY AND ITS SEGMENTS [J].
KALEPS, I ;
CLAUSER, CE ;
YOUNG, JW ;
CHANDLER, RF ;
ZEHNER, GF ;
MCCONVILLE, JT .
ERGONOMICS, 1984, 27 (12) :1225-1237
[7]   Validation of the Tracmor triaxial accelerometer system for walking [J].
Levine, JA ;
Baukol, PA ;
Westerterp, KR .
MEDICINE AND SCIENCE IN SPORTS AND EXERCISE, 2001, 33 (09) :1593-1597
[8]   Knee kinematics with a high-flexion posterior stabilized total knee prosthesis: An in vitro robotic experimental investigation [J].
Li, G ;
Most, E ;
Sultan, PG ;
Schule, S ;
Zayontz, S ;
Park, SE ;
Rubash, HE .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2004, 86A (08) :1721-1729
[9]   METHODS TO ASSESS PHYSICAL-ACTIVITY WITH SPECIAL REFERENCE TO MOTION SENSORS AND ACCELEROMETERS [J].
MEIJER, GAL ;
WESTERTERP, KR ;
VERHOEVEN, FMH ;
KOPER, HBM ;
TENHOOR, F .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1991, 38 (03) :221-229
[10]   Physical activity assessment: A review of methods [J].
Melanson, EL ;
Freedson, PS .
CRITICAL REVIEWS IN FOOD SCIENCE AND NUTRITION, 1996, 36 (05) :385-396