Frequency of Conduction Disturbances After Transcatheter Implantation of an Edwards Sapien Aortic Valve Prosthesis

被引:67
作者
Godin, Matthieu [1 ]
Eltchaninoff, Helene [1 ]
Furuta, Akira [1 ]
Tron, Christophe [1 ]
Anselme, Frederic [1 ]
Bejar, Karim [1 ]
Sanchez-Giron, Carlos [1 ]
Bauer, Fabrice [1 ]
Litzler, Pierre-Yves [2 ]
Bessou, Jean-Paul [2 ]
Cribier, Alain [1 ]
机构
[1] Rouen Univ Hosp, Dept Cardiol, Rouen, France
[2] Rouen Univ Hosp, Dept Cardiothorac Surg, Rouen, France
关键词
ATRIOVENTRICULAR-BLOCK; HEART-VALVE; REPLACEMENT; STENOSIS;
D O I
10.1016/j.amjcard.2010.04.029
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
We evaluated the incidence of conduction abnormalities and requirement for permanent pacemaker in patients undergoing transcatheter aortic valve implantation (TAVI) with the Edwards Sapien prosthesis. In 2009, >8,000 patients were treated with TAVI using 1 of the 2 commercialized models of bioprosthesis (Edwards Sapien, Edwards Lifesciences, Irvine, California; and Core Valve, Medtronic, Irvine, California). Occurrence of conduction abnormalities including complete atrioventricular block requiring permanent pacemaker has been reported after TAVI with the 2 models of valve, more frequently with the Core Valve. We analyzed standard 12-lead electrocardiograms of 69 consecutive patients in whom an Edwards Sapien prosthesis was successfully implanted. Electrocardiograms were examined before treatment, at day 1, and at I-month follow-up. Heart rate, PR and QT intervals and QRS duration were measured and the presence of a first-, second-, or third-degree atrioventricular block was documented. There was a slight increase in heart rate and a discrete decrease in QT interval at day 1. These values had returned to baseline values at 1 month. There was no change in PR interval but a transitory increase in QRS duration was noted. Frequency of left bundle branch block increased from 14.5% at baseline to 27.5% at day 1 with a decreased incidence at day 30 (21.3%). Permanent pacemaker was required in only 3 patients (4.3%). In conclusion, in our experience, conductive disorders and requirement of a definitive pacemaker after implantation of an Edwards Sapien aortic bioprosthesis are infrequent. The physical properties of this prosthesis may explain this observation. (C) 2010 Published by Elsevier Inc. (Am J Cardiol 2010;106:707-712)
引用
收藏
页码:707 / 712
页数:6
相关论文
共 25 条
[21]   Minimally invasive transapical beating heart aortic valve implantation - proof of concept [J].
Walther, Thomas ;
Falk, Volkmar ;
Borger, Michael A. ;
Dewey, Todd ;
Wimmer-Greinecker, Gerhard ;
Schuler, Gerhard ;
Mack, Michael ;
Mohr, Friedrich W. .
EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 2007, 31 (01) :9-15
[22]   Percutaneous aortic valve implantation retrograde from the femoral artery [J].
Webb, JG ;
Chandavimol, M ;
Thompson, CR ;
Ricci, DR ;
Carere, RG ;
Munt, BI ;
Buller, CE ;
Pasupati, S ;
Lichtenstein, S .
CIRCULATION, 2006, 113 (06) :842-850
[23]   Percutaneous transarterial aortic valve replacement in selected high-risk patients with aortic stenosis [J].
Webb, John G. ;
Pasupati, Sanjeevan ;
Humphries, Karin ;
Thompson, Christopher ;
Altwegg, Lukas ;
Moss, Robert ;
Sinhal, Ajay ;
Carere, Ronald G. ;
Munt, Brad ;
Ricci, Donald ;
Ye, Jian ;
Cheung, Anson ;
Lichtenstein, Sam V. .
CIRCULATION, 2007, 116 (07) :755-763
[24]   Transcatheter Aortic Valve Implantation Impact on Clinical and Valve-Related Outcomes [J].
Webb, John G. ;
Altwegg, Lukas ;
Boone, Robert H. ;
Cheung, Anson ;
Ye, Jian ;
Lichtenstein, Samuel ;
Lee, May ;
Masson, Jean Bernard ;
Thompson, Christopher ;
Moss, Robert ;
Carere, Ron ;
Munt, Brad ;
Nietlispach, Fabian ;
Humphries, Karin .
CIRCULATION, 2009, 119 (23) :3009-3016
[25]   CRITERIA FOR INTRAVENTRICULAR-CONDUCTION DISTURBANCES AND PRE-EXCITATION [J].
WILLEMS, JL ;
DEMEDINA, EOR ;
BERNARD, R ;
COUMEL, P ;
FISCH, C ;
KRIKLER, D ;
MAZUR, NA ;
MEIJLER, FL ;
MOGENSEN, L ;
MORET, P ;
PISA, Z ;
RAUTAHARJU, PM ;
SURAWICZ, B ;
WATANABE, Y ;
WELLENS, HJJ .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1985, 5 (06) :1261-1275