Integrated microfluidic system for the simultaneous determination of ammonia, creatinine, and urea

被引:25
作者
Suzuki, H [1 ]
Matsugi, Y [1 ]
机构
[1] Univ Tsukuba, Inst Mat Sci, Tsukuba, Ibaraki 3058573, Japan
关键词
ammonia; creatinine; urea; flow channel; air gap; differential mode;
D O I
10.1016/j.snb.2004.12.032
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
An integrated sensing system for ammonia, creatinine, and urea incorporated in a micro-flow channel was microfabricated. The transducers for these sensors were air-gap Severinghaus-type ammonia sensors. The air gap was formed between the electrolyte solution for the ammonia sensor and sample solution. Creatinine demiminase and urease were immobilized at the bottom of the flow channel facing the indicator electrodes for the ammonia sensors. The concentrations of the three analytes were determined separately by taking differential outputs in the presence of the other chemicals. The solution pH in the system was changed by fabricating an active flow channel. A three-electrode system was formed at the bottom of the flow channel. The pH change was caused by applying a negative potential to an actuator electrode. By switching on the flow channel, a neutral sample solution became basic, and distinct responses were observed for the three sensors. (c) 2004 Elsevier B.V. All rights reserved.
引用
收藏
页码:700 / 707
页数:8
相关论文
共 15 条
[1]   MULTIENZYME ELECTRODE USING HYDROGEN-ION-SENSITIVE FIELD-EFFECT TRANSISTORS [J].
HANAZATO, Y ;
NAKAKO, M ;
SHIONO, S .
IEEE TRANSACTIONS ON ELECTRON DEVICES, 1986, 33 (01) :47-51
[2]  
KARUBE I, 1986, ANAL CHIM ACTA, V185, P195
[3]   Creatinine biosensors: principles and designs [J].
Killard, AJ ;
Smyth, MR .
TRENDS IN BIOTECHNOLOGY, 2000, 18 (10) :433-437
[4]   Disposable creatinine sensor based on thick-film hydrogen peroxide electrode system [J].
Kim, EJ ;
Haruyama, T ;
Yanagida, Y ;
Kobatake, E ;
Aizawa, M .
ANALYTICA CHIMICA ACTA, 1999, 394 (2-3) :225-231
[5]   Microfabricated amperometric creatine and creatinine biosensors [J].
Madaras, MB ;
Popescu, IC ;
Ufer, S ;
Buck, RP .
ANALYTICA CHIMICA ACTA, 1996, 319 (03) :335-345
[6]   Photosensitive polyurethanes applied to the development of CHEMFET and ENFET devices for biomedical sensing [J].
Munoz, J ;
Jimenez, C ;
Bratov, A ;
Bartroli, J ;
Alegret, S ;
Dominguez, C .
BIOSENSORS & BIOELECTRONICS, 1997, 12 (07) :577-585
[7]   A LIFT-OFF METHOD FOR PATTERNING ENZYME-IMMOBILIZED MEMBRANES IN MULTI-BIOSENSORS [J].
NAKAMOTO, S ;
ITO, N ;
KURIYAMA, T ;
KIMURA, J .
SENSORS AND ACTUATORS, 1988, 13 (02) :165-172
[8]   Polyurethane-acrylate photocurable polymeric membrane for ion-sensitive field-effect transistor based urea biosensors [J].
Puig-Lleixà, C ;
Jiménez, C ;
Alonso, J ;
Bartrolí, J .
ANALYTICA CHIMICA ACTA, 1999, 389 (1-3) :179-188
[9]  
ROSS JW, 1973, PURE APPL CHEM, V36, P473
[10]   A miniaturized urea sensor based on the integration of both ammonium based urea enzyme field effect transistor and a reference field effect transistor in a single chip [J].
Senillou, A ;
Jaffrezic-Renault, N ;
Martelet, C ;
Cosnier, S .
TALANTA, 1999, 50 (01) :219-226