A miniaturized urea sensor based on the integration of both ammonium based urea enzyme field effect transistor and a reference field effect transistor in a single chip

被引:83
作者
Senillou, A
Jaffrezic-Renault, N
Martelet, C
Cosnier, S
机构
[1] Ecole Cent Lyon, UMR CNRS 5621, Lab Ingenierie & Fonctionnalisation Surfaces, F-69131 Ecully, France
[2] Univ Grenoble 1, UMR CNRS 5630, Lab Electrochim Organ & Photochim Redox, F-38041 Grenoble 9, France
关键词
biosensor; enzyme field effect transistor; urease; urea;
D O I
10.1016/S0039-9140(99)00122-8
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A urea biosensor prepared by covalent binding of urease directly to the surface of an ammonium-sensitive field effect transistor (FET) is described. Nonactin incorporated in carboxylated polyvinyl chloride was used to obtain the sensitive membrane of the ammonium-sensitive FET. The grafting of urease on the polyvinylchloride-COOH membrane surface was performed through carbodiimide coupling. The activity of the immobilized enzyme was spectrometrically controlled through the time-dependent disappearance of the absorbance of NADH at 340 nm. An apparent activity of 50% was found, compared with free enzyme. The sensitivity of the urea enzyme FET is 50 mV/pUrea working in a differential mode of 2 mu M to 1 mM, this sensitivity being constant during 15 days. Finally, in order to test the potentialities of the urea biosensor for the environmental applications, the detection of heavy metal ions such as Cu(II) and Hg(II) in solution was performed by measuring the remaining activity of the inhibited enzyme. (C) 1999 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:219 / 226
页数:8
相关论文
共 23 条
[1]  
ALEGRET S, 1993, SENSOR ACTUAT B-CHEM, V15, P453
[2]   HOW ELECTRICAL AND CHEMICAL-REQUIREMENTS FOR REFETS MAY COINCIDE [J].
BERGVELD, P ;
VANDENBERG, A ;
VANDERWAL, PD ;
SKOWRONSKAPTASINSKA, M ;
SUDHOLTER, EJR ;
REINHOUDT, DN .
SENSORS AND ACTUATORS, 1989, 18 (3-4) :309-327
[3]  
BOUBRYAK OA, 1995, SENSOR ACTUAT B-CHEM, V26, P429
[4]   DEVELOPMENT OF AMPEROMETRIC BIOSENSORS BASED ON THE IMMOBILIZATION OF ENZYMES IN POLYMER-FILMS ELECTROGENERATED FROM A SERIES OF AMPHIPHILIC PYRROLE DERIVATIVES [J].
COCHEGUERENTE, L ;
COSNIER, S ;
INNOCENT, C ;
MAILLEY, P .
ANALYTICA CHIMICA ACTA, 1995, 311 (01) :23-30
[5]   FABRICATION AND CHARACTERIZATION OF ENFET DEVICES FOR BIOMEDICAL APPLICATIONS AND ENVIRONMENTAL MONITORING [J].
COLAPICCHIONI, C ;
BARBARO, A ;
PORCELLI, F .
SENSORS AND ACTUATORS B-CHEMICAL, 1992, 6 (1-3) :202-207
[6]   A NEW STRATEGY FOR THE CONSTRUCTION OF A TYROSINASE-BASED AMPEROMETRIC PHENOL AND O-DIPHENOL SENSOR [J].
COSNIER, S ;
INNOCENT, C .
BIOELECTROCHEMISTRY AND BIOENERGETICS, 1993, 31 (02) :147-160
[7]   Application of charged polymeric materials as additional permselective membranes for improvement of the performance characteristics of urea-sensitive enzymatic field effect transistors .1. Determination of urea in model solutions [J].
Gorchkov, DV ;
Soldatkin, AP ;
Poyard, S ;
JaffrezicRenault, N ;
Martelet, C .
MATERIALS SCIENCE & ENGINEERING C-BIOMIMETIC MATERIALS SENSORS AND SYSTEMS, 1997, 5 (01) :23-28
[8]   Application of the charged polymeric materials as additional permselective membranes for improvement of the performance characteristics of urea-sensitive ENFETs .2. Urea determination in blood serum [J].
Gorchkov, DV ;
Poyard, S ;
Soldatkin, AP ;
JaffrezicRenault, N ;
Martelet, C .
MATERIALS SCIENCE & ENGINEERING C-BIOMIMETIC MATERIALS SENSORS AND SYSTEMS, 1997, 5 (01) :29-34
[9]   Penicillin enzyme biosensors based on pH membrane electrodes [J].
Koncki, R ;
Leszczynska, E .
ANALYTICA CHIMICA ACTA, 1996, 321 (01) :27-34
[10]   UREA SENSOR BASED ON AN AMMONIUM-ION-SENSITIVE FIELD-EFFECT TRANSISTOR [J].
MIYAHARA, Y ;
TSUKADA, K ;
MIYAGI, H ;
SIMON, W .
SENSORS AND ACTUATORS B-CHEMICAL, 1991, 3 (04) :287-293