Laser ablation inductively coupled plasma isotope dilution mass spectrometry (LA-ICP-IDMS) with a special laser ablation system for bulk analyses (LINA-Spark (TM)-Atomiser) was applied for direct determinations of chlorine, bromine, and iodine in rock and sediment samples. Special attention was focused on possible inter-halogen fractionations and analyte/spike isotope fractionations by using LA-ICP-MS and LA-ICP-IDMS, respectively. A variation of Br/Cl and I/Cl element intensity ratios by a factor of 1.3-3 was observed when changing the nebulizer and the laser power density in the range of 2-10 GW cm(-2) respectively. When using an internal gas flow rate in the range of 0.84-1.0 L min(-1) standard for halogen quantification in LA-ICP-MS, this inter-element fractionation can cause systematic errors, which can be avoided by applying the isotope dilution technique. However, at high laser power densities (> 5.7 GW cm(-2) for iodine and > 4.0 GW cm(-2) for bromine and chlorine) the corresponding measured isotope ratio of the isotope-diluted sample deviates significantly from the target value. Under optimised conditions concentrations in the range of 30 mu g g(-1) 16 x 10(3) mu g g(-1) for chlorine, < 2-140 mu g g(-1) for bromine, and < 0.1-31 mu g g(-1) for iodine were determined by LA-ICP-IDMS in two sediment reference materials (SRM 1646, SRM 2704) and three rock reference samples (GS-N, Granite; BX-N. Bauxite; DT-N. Disthene), which have not been certified for these halogens. The sediment results agree well within the given uncertainties with indicative values by different methods and the results of the rock samples with those obtained by negative thermal ionisation isotope dilution mass spectrometry. The detection limits of LA-ICP-IDMS are 8 mu g g(-1) for chlorine, 1.7 mu g g(-1) for bromine, and 0.1 mu g g for iodine. (c) 2004 Elsevier B.V. All rights reserved.