Effects of waveform and polarity on defibrillation thresholds in humans using a transvenous lead system

被引:36
作者
Shorofsky, SR
Gold, MR
机构
[1] Division of Cardiology, Department of Medicine, Univ. of Maryland Medical Center, Baltimore, MD
[2] Division of Cardiology, Univ. of Maryland Medical System, Baltimore, MD 21201
关键词
D O I
10.1016/S0002-9149(96)00284-6
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Minimizing defibrillation thresholds is important to allow for implantation of downsized pulse generators with reduced outputs while maintaining an adequate defibrillation safety margin, Recent studies have demonstrated a significant reduction in monophasic defibrillation thresholds with a transvenous lead when the polarity was reversed (proximal coil = cathode), However, conflicting data exist concerning the effect of polarity reversal on biphasic defibrillation thresholds, The present study was designed to evaluate prospectively the effect of waveform shape and polarity on defibrillation thresholds in humans, The group studied consisted of 26 patients undergoing cardioverter-defibrillator implantation for standard indications, All data were obtained with a transvenous lead alone configuration, Defibrillation thresholds were determined using a step-down protocol with the initial waveform and polarity randomized, Reversing polarity significantly decreased the delivered energy at defibrillation threshold with monophasic waveforms (14.8 +/- 7.1 vs 20.4 +/- 8.9 J; p <0.001), but had no effect on the overall efficacy of biphasic waveforms (11.1 +/- 5.5 vs 12.2 +/- 6.5 J), In the subgroup of patients with high biphasic defibrillation thresholds (greater than or equal to 15 J), reversing polarity decreased the defibrillation threshold from 18.2 +/- 5.1 to 13.3 +/- 5.8 J (p <0.001), Similarly, the improvement in defibrillation thresholds with reversing polarity of monophasic waveforms was confined to the subgroup of patients with higher defibrillation thresholds. Therefore, the lack of group effect of polarity on biphasic defibrillation thresholds may be simply due to the overall lowering of defibrillation thresholds by this waveform.
引用
收藏
页码:313 / 316
页数:4
相关论文
共 25 条
[11]   EFFECTS OF INITIAL POLARITY ON DEFIBRILLATION THRESHOLD WITH BIPHASIC PULSES [J].
NATALE, A ;
SRA, T ;
DHALA, A ;
JAZAYERI, M ;
DESHPANDE, S ;
AXTELL, K ;
AKHTAR, M .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1995, 18 (10) :1889-1893
[12]   THE AUTOMATIC IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR - EFFECT OF PATCH POLARITY ON DEFIBRILLATION THRESHOLD [J].
ONEILL, PG ;
BOAHENE, KA ;
LAWRIE, GM ;
HARVILL, LF ;
PACIFICO, A .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1991, 17 (03) :707-711
[13]   SERIAL DEFIBRILLATION THRESHOLD MEASURES IN MAN - A PROSPECTIVE CONTROLLED-STUDY [J].
POOLE, JE ;
BARDY, GH ;
DOLACK, GL ;
KUDENCHUK, PJ ;
ANDERSON, J ;
JOHNSON, G .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 1995, 6 (01) :19-25
[14]   INFLUENCE OF IMPLANTABLE CARDIOVERTER-DEFIBRILLATORS ON THE LONG-TERM PROGNOSIS OF SURVIVORS OF OUT-OF-HOSPITAL CARDIAC-ARREST [J].
POWELL, AC ;
FUCHS, T ;
FINKELSTEIN, DM ;
GARAN, H ;
CANNOM, DS ;
MCGOVERN, BA ;
KELLY, E ;
VLAHAKES, GJ ;
TORCHIANA, DF ;
RUSKIN, JN .
CIRCULATION, 1993, 88 (03) :1083-1092
[15]   DEFIBRILLATION WITH THE SEQUENTIAL PULSE TECHNIQUE - REPRODUCIBILITY WITH REPEATED SHOCKS [J].
RATTES, MF ;
JONES, DL ;
SOHLA, A ;
SHARMA, AD ;
JARVIS, E ;
KLEIN, GJ .
AMERICAN HEART JOURNAL, 1986, 111 (05) :874-878
[16]   DEFIBRILLATION THRESHOLD - A SIMPLE AND QUANTITATIVE ESTIMATE OF THE ABILITY TO DEFIBRILLATE [J].
RATTES, MF ;
JONES, DL ;
SHARMA, AD ;
KLEIN, GJ .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1987, 10 (01) :70-77
[17]   DEFIBRILLATION THRESHOLD - CLINICAL UTILITY AND THERAPEUTIC IMPLICATIONS [J].
SINGER, I ;
LANG, D .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1992, 15 (06) :932-949
[18]  
STRICKBERGER SA, 1995, J AM COLL CARDIOL, V25, P1605
[19]   EFFECT OF SHOCK POLARITY ON VENTRICULAR DEFIBRILLATION THRESHOLD USING A TRANSVENOUS LEAD SYSTEM [J].
STRICKBERGER, SA ;
HUMMEL, JD ;
HORWOOD, LE ;
JENTZER, J ;
DAOUD, E ;
NIEBAUER, M ;
BAKR, O ;
MAN, KC ;
WILLIAMSON, BD ;
KOU, W ;
MORADY, F .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1994, 24 (04) :1069-1072
[20]   OPTIMIZATION OF BIPHASIC WAVE-FORMS FOR HUMAN NONTHORACOTOMY DEFIBRILLATION [J].
SWARTZ, JF ;
FLETCHER, RD ;
KARASIK, PE .
CIRCULATION, 1993, 88 (06) :2646-2654