Non-uniform versus uniform attenuation correction in brain perfusion SPET of healthy volunteers

被引:25
作者
Van Laere, K
Koole, M
Versijpt, J
Dierckx, R
机构
[1] State Univ Ghent Hosp, Div Nucl Med, B-9000 Ghent, Belgium
[2] Fac Sci Appl, Dept Elect & Informat Syst, B-9000 Ghent, Belgium
关键词
transmission scanning; attenuation correction; cerebral perfusion; single-photon emission tomography; semi-quantification;
D O I
10.1007/s002590000407
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Although non-uniform attenuation correction (NUAC) can supply more accurate absolute quantification, it is not entirely clear whether NUAC provides clearcut benefits in the routine clinical practice of brain SPET imaging. The aim of this study was to compare the effect of NUAC versus uniform attenuation correction (UAC) on volume of interest (VOI)-based semi-quantification of a large age- and gender-stratified brain perfusion normal database. Eighty-nine healthy volunteers (46 females and 43 males, aged 20-81 years) underwent standardised high-resolution single-photon emission tomography (SPET) with 925 MBq Tc-99m-ethyl cysteinate dimer (ECD) on a Toshiba GCA-9300A camera with Gd-153 or (TC)-T-99m transmission CT scanning. Emission images were reconstructed by filtered back-projection and scatter corrected using the triple-energy window correction method. Both non-uniform Chang attenuation correction (one iteration) and uniform Sorenson correction (attenuation coefficient 0.09 cm(-1)) were applied, Images were automatically reoriented to a stereotactic template on which 35 predefined VOIs were defined for semi-quantification (normalisation on total VOI counts). Small but significant differences between relative VOI uptake values for NUAC versus UAC in the infratentorial region were found. VOI standard deviations were significantly smaller for UAC, 4.5% (range 2.6-7.5), than for NUAC, 5.0% (2.3-9.0) (P<0.01). Higher filter roll-off values of the transmission reconstruction filter increased noise in the emission images and altered estimated cortical attenuation coefficients as well as uptake values. In conclusion, semi-quantification based upon reconstruction with UAC results in very similar Tc-99m-ECD uptake values in healthy volunteers to those obtained with NUAC, although values for the infratentorial region are slightly lower. NUAC produces a slight increase in inter-subject variability. Further study is necessary in various patient populations to establish the full clinical impact of NUAC in brain perfusion SPET.
引用
收藏
页码:90 / 98
页数:9
相关论文
共 42 条
[1]   Dosimetry of transmission measurements in nuclear medicine: a study using anthropomorphic phantoms and thermoluminescent dosimeters [J].
Almeida, P ;
Bendriem, B ;
de Dreuille, O ;
Peltier, A ;
Perrot, C ;
Brulon, V .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE, 1998, 25 (10) :1435-1441
[2]   Transmission scanning in emission tomography [J].
Bailey, DL .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE, 1998, 25 (07) :774-787
[3]   Regional cerebral blood flow pattern in normal young and aged volunteers: A Tc-99m-HMPAO SPET Study [J].
Catafau, AM ;
Lomena, FJ ;
Pavia, J ;
Parellada, E ;
Bernardo, M ;
Setoain, J ;
Tolosa, E .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE, 1996, 23 (10) :1329-1337
[5]   PROBLEMS IN SIGNAL-TO-NOISE RATIO FOR ATTENUATION CORRECTION IN HIGH-RESOLUTION PET [J].
DAHLBOM, M ;
HOFFMAN, EJ .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1987, 34 (01) :288-293
[6]  
Dey D, 1999, J NUCL MED, V40, P448
[7]   Comparison of ordered subsets expectation maximization and Chang's attenuation correction method in quantitative cardiac SPET: A phantom study [J].
Dey, D ;
Slomka, PJ ;
Hahn, LJ ;
Kloiber, R .
NUCLEAR MEDICINE COMMUNICATIONS, 1998, 19 (12) :1149-1157
[8]  
FICARO EP, 1995, J NUCL MED, V36, P921
[9]  
Freedman NMT, 1996, J NUCL MED, V37, P690
[10]   TC-99M ATTENUATION COEFFICIENTS IN WATER-FILLED PHANTOMS DETERMINED WITH GAMMA CAMERAS [J].
HARRIS, CC ;
GREER, KL ;
JASZCZAK, RJ ;
FLOYD, CE ;
FEARNOW, EC ;
COLEMAN, RE .
MEDICAL PHYSICS, 1984, 11 (05) :681-685