摘要
Purpose: To improve the quantification of myocardial perfusion by registering the time series of magnetic resonance (MR) images with injection of gadolinium. Materials and Methods: Eight patients underwent MR scans to perform myocardial perfusion exam. Two short axis views of the left ventricle (LV) were acquired in free breathing. Two masks for performing the spatial registration of the images were evaluated. The registration was based on pixel intensity in a multi-resolution scheme. The efficiency of this correction was evaluated by calculating geometric residual displacement of the LV and by fitting the data to a compartment model fit with two parameters: K1, the blood-to-myocardium transfer coefficient, and Vd, the distribution volume of the contrast media. Results: The registration stage allowed a decrease in the observed motion of the LV from more than 1.98 +/- 0.68 mm to less than 0.56 +/- 0.18 rorn (mean +/- SD). Variability obtained in the perfusion analysis decreased from 46 103% to 5 +/- 4% for KI parameter and from 18 +/- 21% to 5 +/- 5% for Vd parameter. Conclusion: As with manual correction, this automatic motion correction leads to accurate perfusion parameters in dynamic cardiac MR imaging after contrast agent injection. This automatic stage requires placing only one mask over one frame of the perfusion study instead of manually shifting each image to fit a reference image of the perfusion study.
引用
收藏
页码:160 / 168
页数:9
相关论文
共 21 条
[1]   Automated registration of dynamic MR images for the quantification of myocardial perfusion [J].
Bidaut, LM ;
Vallée, JP .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2001, 13 (04) :648-655
[2]   ULTRASOUND QUANTITATION OF RESPIRATORY ORGAN MOTION IN THE UPPER ABDOMEN [J].
DAVIES, SC ;
HILL, AL ;
HOLMES, RB ;
HALLIWELL, M ;
JACKSON, PC .
BRITISH JOURNAL OF RADIOLOGY, 1994, 67 (803) :1096-1102
[3]   SIMULTANEOUS TEMPORAL RESOLUTION OF CARDIAC AND RESPIRATORY MOTION IN MR-IMAGING [J].
FREDRICKSON, JO ;
WEGMULLER, H ;
HERFKENS, RJ ;
PELC, NJ .
RADIOLOGY, 1995, 195 (01) :169-175
[4]   SEMIAUTOMATED ROI ANALYSIS IN DYNAMIC MR STUDIES .1. IMAGE-ANALYSIS TOOLS FOR AUTOMATIC CORRECTION OF ORGAN DISPLACEMENTS [J].
GERIG, G ;
KIKINIS, R ;
KUONI, W ;
VONSCHULTHESS, GK ;
KUBLER, O .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1991, 15 (05) :725-732
[5]   Global optimization of mutual information: application to three-dimensional retrospective registration of magnetic resonance images [J].
He, RJ ;
Narayana, PA .
COMPUTERIZED MEDICAL IMAGING AND GRAPHICS, 2002, 26 (04) :277-292
[6]   Myocardial perfusion modeling using MRI [J].
Larsson, HBW ;
FritzHansen, T ;
Rostrup, E ;
Sondergaard, L ;
Ring, P ;
Henriksen, O .
MAGNETIC RESONANCE IN MEDICINE, 1996, 35 (05) :716-726
[7]   Multimodality image registration by maximization of mutual information [J].
Maes, F ;
Collignon, A ;
Vandermeulen, D ;
Marchal, G ;
Suetens, P .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1997, 16 (02) :187-198
[8]  
Maintz J B, 1998, Med Image Anal, V2, P1, DOI 10.1016/S1361-8415(01)80026-8
[9]   SIMULTANEOUS USAGE OF HOMOLOGOUS POINTS, LINES, AND PLANES FOR OPTIMAL, 3-D, LINEAR REGISTRATION OF MULTIMODALITY IMAGING DATA [J].
MEYER, CR ;
LEICHTMAN, GS ;
BRUNBERG, JA ;
WAHL, RL ;
QUINT, LE .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1995, 14 (01) :1-11
[10]  
NETSCH T, 2002, P 19 ANN M ES MRMB C, P105