Minimization of acid effects at low consumption rates in an axially viewed inductively coupled plasma atomic emission spectrometer by using micronbulizer-based sample introduction systems

被引:53
作者
Todoli, JL [1 ]
Mermet, JM [1 ]
机构
[1] Univ Lyon 1, Sci Analyt Lab, UMR 5619, F-69622 Villeurbanne, France
关键词
low sample consumption rates; microconcentric nebulizer; direct injection nebulizer; aerosol desolvation; PTFE porous membrane; acid effects; inductively coupled plasma atomic emission spectrometry; axial viewing; ionic-to-atomic line intensity ratio;
D O I
10.1039/a801124i
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Two methods were applied in order to minimize acid effects at very low liquid uptake rates in an axially viewed ICP-AES system: aerosol solvent elimination and direct injection in the plasma base. The latter choice eliminated the use of a spray chamber. Four different sample introduction systems were evaluated based on the use of micronebulizers, The first system made use of a so-called microconcentric nebulizer (MCN) associated with a double-pass Scott spray chamber. Two other systems were based on the use of a desolvation unit that consisted of a first and a second stage in which the solvent was partially removed. The solvent elimination system was either two Liebig condensers connected in series or a porous PTFE membrane, In these cases two different MCNs were coupled. The last sample introduction system employed was a direct injection nebulizer (DIN) that eliminated the contribution of the spray chamber to the acid interference. The liquid flow rates ranged from 5 to 120 mu l min(-1), depending on the sample introduction system employed, Nitric, hydrochloric and sulfuric acid of 0.9 mol l(-1) were tested and 3.6 mol l(-1) nitric acid solution was also studied in order to evaluate the acid concentration effect. The Mg II to Mg I line intensity ratio was used to evaluate any possible changes in the plasma conditions. The results showed that, for the MCN coupled to the spray chamber, the lower the liquid flow rate, the greater was the acid interference. The extension of this effect was dependent on the MCN configuration. For the desolvation systems and DIN the behavior was the opposite, the acid interference being more important as the liquid flow rate increased. For the four acids employed, and at very low liquid how rates (below 30 mu l min(-1)), the acid effect was eliminated by using a conventional desolvation system with a heating temperature of 160 degrees C, the condensation temperatures for the two condensers being 10 and 0 degrees C, respectively. The use of a membrane was found to be advisable because, under some conditions, acids and water gave the same analytical signal. As regards the DIN, it was able to eliminate the interference for 0.9 mol l(-1) nitric and hydrochloric acid. Nevertheless, the two remaining solutions, i,e,, 0.9 mol l(-1) sulfuric and 3.6 mol l(-1) nitric acid, gave rise to a decrease in the signal of around 20%.
引用
收藏
页码:727 / 734
页数:8
相关论文
共 47 条
[1]   Determination of rare earth elements in wine by inductively coupled plasma mass spectrometry using a microconcentric nebulizer [J].
Augagneur, S ;
Medina, B ;
Szpunar, J ;
Lobinski, R .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1996, 11 (09) :713-721
[2]   A MEMBRANE INTERFACE FOR ORGANIC-SOLVENT SAMPLE INTRODUCTION INTO INDUCTIVELY COUPLED PLASMAS [J].
BACKSTROM, K ;
GUSTAVSSON, A ;
HIETALA, P .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1989, 44 (10) :1041-1048
[3]   REMOVAL OF A MATRIX EFFECT IN ICP-AES MULTIELEMENT ANALYSIS BY SIMPLEX OPTIMIZATION [J].
BELCHAMBER, RM ;
BETTERIDGE, D ;
WADE, AP ;
CRUICKSHANK, AJ ;
DAVISON, P .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1986, 41 (05) :503-505
[5]   Universal calibration for analysis of organic solutions by inductively coupled plasma atomic emission spectrometry [J].
Botto, RI ;
Zhu, JJ .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1996, 11 (09) :675-681
[6]   USE OF AN ULTRASONIC NEBULIZER WITH MEMBRANE DESOLVATION FOR ANALYSIS OF VOLATILE SOLVENTS BY INDUCTIVELY-COUPLED PLASMA-ATOMIC EMISSION-SPECTROMETRY [J].
BOTTO, RI ;
ZHU, JJ .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1994, 9 (09) :905-912
[7]   EFFECT OF NITRIC AND HYDROCHLORIC ACIDS ON RARE-EARTH ELEMENT (REE) INTENSITIES IN INDUCTIVELY-COUPLED PLASMA EMISSION-SPECTROMETRY [J].
BRENNER, IB ;
MERMET, JM ;
SEGAL, I ;
LONG, GL .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1995, 50 (4-7) :323-331
[8]   STUDY OF THE DEPRESSIVE EFFECTS OF NITRIC-ACID ON THE LINE-INTENSITIES OF RARE-EARTH ELEMENTS IN INDUCTIVELY-COUPLED PLASMA-ATOMIC EMISSION-SPECTROMETRY [J].
BRENNER, IB ;
SEGAL, I ;
MERMET, M ;
MERMET, JM .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1995, 50 (4-7) :333-340
[9]   MATRIX EFFECTS OF POTASSIUM-CHLORIDE AND PHOSPHORIC-ACID IN ARGON INDUCTIVELY-COUPLED PLASMA-ATOMIC EMISSION-SPECTROMETRY [J].
BUDIC, B ;
HUDNIK, V .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1994, 9 (01) :53-57
[10]   FUNDAMENTAL-STUDIES ON PNEUMATIC GENERATION AND AEROSOL TRANSPORT IN ATOMIC SPECTROMETRY - EFFECT OF MINERAL ACIDS ON EMISSION INTENSITY IN INDUCTIVELY-COUPLED PLASMA-ATOMIC EMISSION-SPECTROMETRY [J].
CANALS, A ;
HERNANDIS, V ;
TODOLI, JL ;
BROWNER, RF .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1995, 50 (4-7) :305-321