On-line coupling of sol-gel-generated immunoaffinity columns with high-performance liquid chromatography

被引:39
作者
Cichna, M
Markl, P
Knopp, D
Niessner, R
机构
[1] Univ Vienna, Inst Analyt Chem, A-1090 Vienna, Austria
[2] Tech Univ Munich, Inst Hydrochem, D-81377 Munich, Germany
关键词
sol-gel; immunosorbents; water analysis; sample preparation; HPLC; on-line coupling; pyrene;
D O I
10.1016/S0021-9673(01)00798-1
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The paper demonstrates the possibility to use sol-gel-generated immunoaffinity columns as selective sample preparation step in on-line combination with HPLC. In the past sol-gel-generated immunoaffinity columns have only been included in off-line sample preparation schemes. Compared with conventional RP-materials on-line coupling of sol-gel-generated silica matrices with a pore structure designed to retain antibodies poses additional problems caused by their lower pressure tolerance and by the necessity to match the mobile phases not only to take into account the chromatographic properties but also the conformational stability of the antibodies. These problems have been overcome by an on-line system which can be regarded as a prototype for similar systems which exploit the selectivity of sol-gel immunoaffinity columns. The system consists of a sol-gel-generated immunoaffinity column coupled to an RP enrichment column and an analytical column. The practicality of such systems is demonstrated using the example of anti-pyrene immunoaffinity columns applied for the determination of pyrene in aqueous solutions. (C) 2001 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:51 / 58
页数:8
相关论文
共 28 条
[1]   Class-selective immunosorbent for trace-level determination of polycyclic aromatic hydrocarbons in complex sample matrices, used in off-line procedure or on-line coupled with liquid chromatography fluorescence and diode array detections in series [J].
Bouzige, M ;
Pichon, V ;
Hennion, MC .
ENVIRONMENTAL SCIENCE & TECHNOLOGY, 1999, 33 (11) :1916-1925
[2]   Selective on-line immunoextraction coupled to liquid chromatography for the trace determination of benzidine, congeners and related azo dyes in surface water and industrial effluents [J].
Bouzige, M ;
Legeay, P ;
Pichon, V ;
Hennion, MC .
JOURNAL OF CHROMATOGRAPHY A, 1999, 846 (1-2) :317-329
[3]   Sol-gel matrixes doped with atrazine antibodies: Atrazine binding properties [J].
Bronshtein, A ;
Aharonson, N ;
Avnir, D ;
Turniansky, A ;
Altstein, M .
CHEMISTRY OF MATERIALS, 1997, 9 (11) :2632-2639
[4]   Immunoaffinity chromatography of polycyclic aromatic hydrocarbons in columns prepared by the sol-gel method [J].
Cichna, M ;
Knopp, D ;
Niessner, R .
ANALYTICA CHIMICA ACTA, 1997, 339 (03) :241-250
[5]   Optimization of the selectivity of pyrene immunoaffinity columns prepared by the sol-gel method [J].
Cichna, M ;
Markl, P ;
Knopp, D ;
Niessner, R .
CHEMISTRY OF MATERIALS, 1997, 9 (11) :2640-2646
[6]   SOL-GEL ENCAPSULATION METHODS FOR BIOSENSORS [J].
DAVE, BC ;
DUNN, B ;
VALENTINE, JS ;
ZINK, JI .
ANALYTICAL CHEMISTRY, 1994, 66 (22) :A1120-A1127
[7]   ENCAPSULATION OF PROTEINS IN TRANSPARENT POROUS SILICATE-GLASSES PREPARED BY THE SOL-GEL METHOD [J].
ELLERBY, LM ;
NISHIDA, CR ;
NISHIDA, F ;
YAMANAKA, SA ;
DUNN, B ;
VALENTINE, JS ;
ZINK, JI .
SCIENCE, 1992, 255 (5048) :1113-1115
[8]   IMMUNOAFFINITY PRE-COLUMN FOR SELECTIVE ONLINE SAMPLE PRE-TREATMENT IN HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY DETERMINATION OF 19-NORTESTOSTERONE [J].
FARJAM, A ;
DEJONG, GJ ;
FREI, RW ;
BRINKMAN, UAT ;
HAASNOOT, W ;
HAMERS, ARM ;
SCHILT, R ;
HUF, FA .
JOURNAL OF CHROMATOGRAPHY, 1988, 452 :419-433
[9]   ONLINE IMMUNOAFFINITY SAMPLE PRETREATMENT FOR COLUMN LIQUID-CHROMATOGRAPHY - EVALUATION OF DESORPTION TECHNIQUES AND OPERATING-CONDITIONS USING AN ANTIESTROGEN IMMUNO-PRECOLUMN AS A MODEL SYSTEM [J].
FARJAM, A ;
BRUGMAN, AE ;
LINGEMAN, H ;
THBRINKMAN, UA .
ANALYST, 1991, 116 (09) :891-896
[10]   IMMUNOAFFINITY PRECOLUMN FOR SELECTIVE SAMPLE PRETREATMENT IN COLUMN LIQUID-CHROMATOGRAPHY - IMMUNOSELECTIVE DESORPTION [J].
FARJAM, A ;
BRUGMAN, AE ;
SOLDAAT, A ;
TIMMERMAN, P ;
LINGEMAN, H ;
DEJONG, GJ ;
FREI, RW ;
BRINKMAN, UAT .
CHROMATOGRAPHIA, 1991, 31 (9-10) :469-477