Radiation exposure in full-field digital mammography with a flat-panel x-ray detector based on amorphous silicon in comparison with conventional screen-film mammography.

被引:18
作者
Hermann, KP [1 ]
Obenauer, S [1 ]
Grabbe, E [1 ]
机构
[1] Univ Gottingen Klinikum, Zentrum Radiol, Abt Rontgendiagnost 1, D-37075 Gottingen, Germany
来源
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN | 2000年 / 172卷 / 11期
关键词
mammography; digital mammography; flat panel detector; radiation exposure; image quality;
D O I
10.1055/s-2000-8370
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Comparison of radiation exposure between a digital amorphous silicon and a screen-film based mammography system. Evaluation of a possible potential of full-field digital mammography in order to decrease the radiation dose. Methods: The average glandular dose for phantom thicknesses from 30 to 60 mm was calculated from experimentally determined entrance surface air kerma for a digital and a conventional mammography system. The effect of reducing the detector dose and of changing the radiation quality on radiation exposure and on image quality were investigated. Results: By using the delivered settings of the automatic exposure control (AEC) devices, both mammographic systems needed nearly the same doses. Regulations and guidelines on radiation doses were complied. With the digital system, a reduction of radiation exposure of up to 40% by using a higher radiation quality and decreasing slightly the detector dose without loss of diagnostic image quality, might be possible. Conclusion: The potential of fullfield digital mammography for radiation dose reduction, as shown in the present phantom study, needs however, a careful examination under clinical conditions.
引用
收藏
页码:940 / 945
页数:6
相关论文
共 21 条
  • [1] [Anonymous], 1999, MAMMOGRAPHY QUALITY
  • [2] CHANG CH, 1992, EUR RADIOL, V2, P483
  • [3] *EUR COMM, 1999, EUR PROT QUAL CONTR
  • [4] European Commission, 1996, European protocol on dosimetry in mammography
  • [5] EWEN K, 1998, Z MED PHYS, V8, P105
  • [6] Image quality and radiation exposure in digital storage plate mammography with magnification technique.
    Fiedler, E
    Aichinger, U
    Böhner, C
    Säbel, M
    Schulz-Wendtland, R
    Bautz, W
    [J]. ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 1999, 171 (01): : 60 - 64
  • [7] FRIEDRICH M, 1995, STRAHLENEXPOSITION M, V30, P113
  • [8] A magnification mammography technique using storage phosphor plates: Experimental investigations concerning spatial resolution and detection of microcalcifications.
    Funke, M
    Hermann, KP
    Breiter, N
    HundertMark, C
    Sachs, J
    Gruhl, T
    Sperner, W
    Grabbe, E
    [J]. ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 1997, 167 (02): : 174 - 179
  • [9] FUNKE M, 1995, ROFO FORTSCHR RONTG, V163, P20
  • [10] Jung H, 1998, ROFO-FORTSCHR RONTG, V169, P336