Reproducibility of nerve fiber layer thickness measurements using optical coherence tomography

被引:540
作者
Schuman, JS
PedutKloizman, T
Hertzmark, E
Hee, MR
Wilkins, JR
Coker, JG
Puliafito, CA
Fujimoto, JG
Swanson, EA
机构
[1] TUFTS UNIV,SCH MED,NEW ENGLAND EYE CTR,BOSTON,MA 02111
[2] MIT,CAMBRIDGE,MA 02139
[3] MIT,LINCOLN LAB,LEXINGTON,MA 02173
关键词
D O I
10.1016/S0161-6420(96)30410-7
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: Optical coherence tomography (OCT) is a new technology that uses near-infrared light in an interferometer to produce approximately 10-mu m resolution cross-sectional images of the tissue of interest. The authors performed repeated quantitative assessment of nerve fiber layer thickness in individuals with normal and glaucomatous eyes, and they evaluated the reproducibility of these measurements. Methods: The authors studied 21 eyes of 21 subjects by OCT. Each subject underwent five repetitions of a series of scans on five separate occasions within a 1-month period. Each series consisted of three circular scans around the optic nerve head (diameters, 2.9, 3.4, and 4.5 mm). Each series was performed separately using internal (fixation with same eye being studied) and external (fixation with contralateral eye) fixation techniques, The eye studied and the sequence of testing were assigned randomly. Results: Internal fixation (IF), in general, provides a slightly higher degree of reproducibility than external fixation (EF). Reproducibility was better in a given eye on a given visit than from visit to visit. Reproducibility as measured by intraclass correlation coefficients were as follows: circle diameter (CD), 2.9 mm, 0.51/0.57 (normal/glaucoma) (IF), 0.43/0.54 (EF); CD, 3.4 mm, 0.56/0.52 (IF), 0.43/0.61 (EF); CD, 4.5 mm, 0.53/0.43 (IF), 0.42/0.49 (EF). Conclusions: Nerve fiber layer thickness can be reproducibly measured using OCT. Internal is superior to external fixation; each circle diameter tested provides adequate reproducibility.
引用
收藏
页码:1889 / 1898
页数:10
相关论文
共 14 条
  • [1] REPRODUCIBILITY OF TOPOGRAPHIC MEASUREMENTS OF THE NORMAL AND GLAUCOMATOUS OPTIC-NERVE HEAD WITH THE LASER TOMOGRAPHIC SCANNER
    DREHER, AW
    TSO, PC
    WEINREB, RN
    [J]. AMERICAN JOURNAL OF OPHTHALMOLOGY, 1991, 111 (02) : 221 - 229
  • [2] OPTICAL COHERENCE TOMOGRAPHY OF MACULAR HOLES
    HEE, MR
    PULIAFITO, CA
    WONG, C
    DUKER, JS
    REICHEL, E
    SCHUMAN, JS
    SWANSON, EA
    FUJIMOTO, JG
    [J]. OPHTHALMOLOGY, 1995, 102 (05) : 748 - 756
  • [3] OPTICAL COHERENCE TOMOGRAPHY OF THE HUMAN RETINA
    HEE, MR
    IZATT, JA
    SWANSON, EA
    HUANG, D
    SCHUMAN, JS
    LIN, CP
    PULIAFITO, CA
    FUJIMOTO, JG
    [J]. ARCHIVES OF OPHTHALMOLOGY, 1995, 113 (03) : 325 - 332
  • [4] Hoskins H D, 1994, J Glaucoma, V3, P17
  • [5] OPTICAL COHERENCE TOMOGRAPHY
    HUANG, D
    SWANSON, EA
    LIN, CP
    SCHUMAN, JS
    STINSON, WG
    CHANG, W
    HEE, MR
    FLOTTE, T
    GREGORY, K
    PULIAFITO, CA
    FUJIMOTO, JG
    [J]. SCIENCE, 1991, 254 (5035) : 1178 - 1181
  • [6] IZATT JA, 1993, P SOC PHOTO-OPT INS, V1877, P136, DOI 10.1117/12.147520
  • [7] MICROMETER-SCALE RESOLUTION IMAGING OF THE ANTERIOR EYE IN-VIVO WITH OPTICAL COHERENCE TOMOGRAPHY
    IZATT, JA
    HEE, MR
    SWANSON, EA
    LIN, CP
    HUANG, D
    SCHUMAN, JS
    PULIAFITO, CA
    FUJIMOTO, JG
    [J]. ARCHIVES OF OPHTHALMOLOGY, 1994, 112 (12) : 1584 - 1589
  • [8] KRUSE FE, 1989, OPHTHALMOLOGY, V96, P1320
  • [9] PULIAFITO CA, 1995, OPHTHALMOLOGY, V102, P217
  • [10] QUANTIFICATION OF NERVE-FIBER LAYER THICKNESS IN NORMAL AND GLAUCOMATOUS EYES USING OPTICAL COHERENCE TOMOGRAPHY - A PILOT-STUDY
    SCHUMAN, JS
    HEE, MR
    PULIAFITO, CA
    WONG, C
    PEDUTKLOIZMAN, T
    LIN, CP
    HERTZMARK, E
    IZATT, JA
    SWANSON, EA
    FUJIMOTO, JG
    [J]. ARCHIVES OF OPHTHALMOLOGY, 1995, 113 (05) : 586 - 596