Applying the right statistics: analyses of measurement studies

被引:1109
作者
Bland, JM
Altman, DG
机构
[1] Inst Hlth Sci, Ctr Stat Med, Canc Res UK Med Stat Grp, Oxford OX3 7LF, England
[2] Univ London St Georges Hosp, Sch Med, London SW17 0RE, England
关键词
agreement; correlation; error; measurement; regression;
D O I
10.1002/uog.122
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
The study of measurement error, observer variation and agreement between different methods of measurement are frequent topics in the imaging literature. We describe the problems of some applications of correlation and regression methods to these studies, using recent examples from this literature. We use a simulated example to show bow these problems and misinterpretations arise. We describe the 95% limits of agreement approach and a similar, appropriate, regression technique. We discuss the difference vs. mean plot, and the pitfalls of plotting difference against one variable only. We stress that these are questions of estimation, not significance tests, and show how confidence intervals can be found for these estimates. Copyright (C) 2003 ISUOG. Published by John Wiley Sons, Ltd.
引用
收藏
页码:85 / 93
页数:9
相关论文
共 20 条
  • [1] MEASUREMENT IN MEDICINE - THE ANALYSIS OF METHOD COMPARISON STUDIES
    ALTMAN, DG
    BLAND, JM
    [J]. JOURNAL OF THE ROYAL STATISTICAL SOCIETY SERIES D-THE STATISTICIAN, 1983, 32 (03) : 307 - 317
  • [2] Serial fetal lung volume measurement using three-dimensional ultrasound
    Bahmaie, A
    Hughes, SW
    Clark, T
    Milner, A
    Saunders, J
    Tilling, K
    Maxwell, DJ
    [J]. ULTRASOUND IN OBSTETRICS & GYNECOLOGY, 2000, 16 (02) : 154 - 158
  • [3] Renal volume measurements: Accuracy and repeatability of US compared with that of MR imaging
    Bakker, J
    Olree, M
    Kaatee, R
    de Lange, EE
    Moons, KGM
    Beutler, JJ
    Beek, FJA
    [J]. RADIOLOGY, 1999, 211 (03) : 623 - 628
  • [4] Pulmonary emphysema: Subjective visual grading versus objective quantification with macroscopic morphometry and thin-section CT densitometry
    Bankier, AA
    De Maertelaer, V
    Keyzer, C
    Gevenois, PA
    [J]. RADIOLOGY, 1999, 211 (03) : 851 - 858
  • [5] COMPARING METHODS OF MEASUREMENT - WHY PLOTTING DIFFERENCE AGAINST STANDARD METHOD IS MISLEADING
    BLAND, JM
    ALTMAN, DG
    [J]. LANCET, 1995, 346 (8982): : 1085 - 1087
  • [6] Measurement error and correlation coefficients
    Bland, JM
    Altman, DG
    [J]. BRITISH MEDICAL JOURNAL, 1996, 313 (7048) : 41 - 42
  • [7] SOME EXAMPLES OF REGRESSION TOWARDS THE MEAN .7.
    BLAND, JM
    ALTMAN, DG
    [J]. BRITISH MEDICAL JOURNAL, 1994, 309 (6957) : 780 - 780
  • [8] STATISTICS NOTES .2. REGRESSION TOWARDS THE MEAN
    BLAND, JM
    ALTMAN, DG
    [J]. BRITISH MEDICAL JOURNAL, 1994, 308 (6942) : 1499 - 1499
  • [9] Measuring agreement in method comparison studies
    Bland, JM
    Altman, DG
    [J]. STATISTICAL METHODS IN MEDICAL RESEARCH, 1999, 8 (02) : 135 - 160
  • [10] STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
    BLAND, JM
    ALTMAN, DG
    [J]. LANCET, 1986, 1 (8476) : 307 - 310