Accuracy and noise in optical Doppler tomography studied by Monte Carlo simulation

被引:20
作者
Lindmo, T [1 ]
Smithies, DJ
Chen, ZP
Nelson, JS
Milner, TE
机构
[1] Norwegian Univ Sci & Technol, Dept Phys, N-7034 Trondheim, Norway
[2] Univ Calif Irvine, Beckman Laser Inst, Irvine, CA 92612 USA
[3] Univ Calif Irvine, Med Clin, Irvine, CA 92612 USA
关键词
D O I
10.1088/0031-9155/43/10/025
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A Monte Carlo model has been developed for optical Doppler tomography (ODT) within the framework of a model for optical coherence tomography (OCT). A phantom situation represented by blood Bowing in a horizontal 100 mu m diameter vessel placed at 250 mu m axial depth in 2% intralipid solution was implemented for the Monte Carlo simulation, and a similar configuration used for experimental ODT measurements in the laboratory. Simulated depth profiles through the centre of the vessel of average Doppler frequency demonstrated an accuracy of 3-4% deviation in frequency values and position localization of flow borders, compared with true values. Stochastic Doppler frequency noise was experimentally observed as a shadowing in regions underneath the vessel and also seen in simulated Doppler frequency depth profiles. By Monte Carlo simulation, this Doppler noise was shown to represent a nearly constant level over an investigated 100 mu m interval of depth underneath the vessel. The noise level was essentially independent of the numerical aperture of the detector and angle between the flow velocity and the direction of observation, as long as this angle was larger than 60 degrees. Since this angle determines the magnitude of the Doppler frequency for backscattering from the flow region, this means that the signal-to-noise ratio between Doppler signal from the flow region to Doppler noise from regions underneath the flow is improved by decreasing the angle between the flow direction and direction of observation. Doppler noise values from Monte Carlo simulations were compared with values from statistical analysis.
引用
收藏
页码:3045 / 3064
页数:20
相关论文
共 21 条
[1]   Noninvasive imaging of in vivo blood flow velocity using optical Doppler tomography [J].
Chen, ZP ;
Milner, TE ;
Srinivas, S ;
Wang, XJ ;
Malekafzali, A ;
vanGemert, MJC ;
Nelson, JS .
OPTICS LETTERS, 1997, 22 (14) :1119-1121
[2]   Optical Doppler tomographic imaging of fluid flow velocity in highly scattering media [J].
Chen, ZP ;
Milner, TE ;
Dave, D ;
Nelson, JS .
OPTICS LETTERS, 1997, 22 (01) :64-66
[3]   Optical Doppler tomography for noninvasive imaging of in vivo blood flow [J].
Chen, ZP ;
Milner, TE ;
Srinivas, S ;
Lindmo, T ;
Dave, D ;
Nelson, JS .
COHERENCE DOMAIN OPTICAL METHODS IN BIOMEDICAL SCIENCE AND CLINICAL APPLICATIONS, PROCEEDINGS OF, 1997, 2981 :112-118
[4]   LASER-DOPPLER VELOCIMETRY AND MONTE-CARLO SIMULATIONS ON MODELS FOR BLOOD PERFUSION IN TISSUE [J].
DEMUL, FFM ;
KOELINK, MH ;
KOK, ML ;
HARMSMA, PJ ;
GREVE, J ;
GRAAFF, R ;
AARNOUDSE, JG .
APPLIED OPTICS, 1995, 34 (28) :6595-6611
[5]   MONTE-CARLO SIMULATIONS OF LASER DOPPLER BLOOD-FLOW MEASUREMENTS IN TISSUE - REPLY TO COMMENT [J].
DEMUL, FFM .
APPLIED OPTICS, 1992, 31 (04) :440-440
[6]   OPTICAL-PROPERTIES OF INTRALIPID - A PHANTOM MEDIUM FOR LIGHT-PROPAGATION STUDIES [J].
FLOCK, ST ;
JACQUES, SL ;
WILSON, BC ;
STAR, WM ;
VANGEMERT, MJC .
LASERS IN SURGERY AND MEDICINE, 1992, 12 (05) :510-519
[7]  
HAWKEY CM, 1989, COLOR ATLAS COMP VET, P17
[8]  
JACQUES SL, 1996, SC34 SPIE SHORT COUR, P26
[9]   MONTE-CARLO SIMULATIONS OF LASER DOPPLER BLOOD-FLOW MEASUREMENTS IN TISSUE [J].
JENTINK, HW ;
DEMUL, FFM ;
HERMSEN, RGAM ;
GRAAFF, R ;
GREVE, J .
APPLIED OPTICS, 1990, 29 (16) :2371-2381
[10]   LASER-DOPPLER BLOOD FLOWMETRY USING 2 WAVELENGTHS - MONTE-CARLO SIMULATIONS AND MEASUREMENTS [J].
KOELINK, MH ;
DEMUL, FFM ;
GREVE, J ;
GRAAFF, R ;
DASSEL, ACM ;
AARNOUDSE, JG .
APPLIED OPTICS, 1994, 33 (16) :3549-3558