A high-performance liquid chromatographic system equipped with on-line reducer, clean-up and concentrator columns for determination of trace levels of nitropolycyclic aromatic hydrocarbons in airborne particulates
被引:33
作者:
Hayakawa, K
Noji, K
Tang, N
Toriba, A
Kizu, R
Sakai, S
Matsumoto, Y
机构:
[1] Kanazawa Univ, Fac Pharmaceut Sci, Kanazawa, Ishikawa 9200934, Japan
[2] Hokkaido Res Inst Environm Sci, Kita Ku, Sapporo, Hokkaido 0600819, Japan
关键词:
nitropolycyclic aromatic hydrocarbon;
platinum/rhodium;
HPLC;
fluorescence;
D O I:
10.1016/S0003-2670(01)01279-X
中图分类号:
O65 [分析化学];
学科分类号:
070302 ;
081704 ;
摘要:
A HPLC system with on-line reducer, clean-up and concentrator columns has been developed for determining nitropoly-cyclic aromatic hydrocarbons (NPAHs). Maximum efficiency of the reducer column (Pt/Rh-coated alumina) was achieved after the column was activated with a flow of air/hydrogen at 350 degreesC for 3 h and when ethanol-acetate buffer was used as the sample carrier solution. After a benzene-ethanol extract from airborne particulates was washed with sodium hydroxide, sulfuric acid and water, an aliquot was injected into the system. NPAHs were separated from interfering substances by the clean-up column and reduced to their amino derivatives by the reducer column. The derivatives were concentrated on the concentrator column and eluted into the separator column. NPAHs such as 1,3-, 1,6-, 1,8-dinitropyrenes and 1-nitropyrene were determined chemilumigenically with linear calibration graphs from I x 10(-11) to 1 x 10(-8) and I x 10(-10) to I X 10(-7) M, respectively, at an injection volume of 100 mul. These NPAHs in airborne particulates were determined by the proposed method. (C) 2001 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:205 / 212
页数:8
相关论文