Temperature quantification using the proton frequency shift technique: In vitro and in vivo validation in an open 0.5 tesla interventional MR scanner during RF ablation

被引:27
作者
Botnar, RM [1 ]
Steiner, P [1 ]
Dubno, B [1 ]
Erhart, P [1 ]
von Schulthess, GK [1 ]
Debatin, JF [1 ]
机构
[1] Univ Zurich Hosp, Dept Radiol, CH-8091 Zurich, Switzerland
关键词
temperature mapping; proton frequency shift; RF thermal ablations; interventional MR; medical instrumentation;
D O I
10.1002/jmri.1063.abs
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Open magnetic resonance (MR) scanners allow MR-guided targeting of tumors, as well as temperature monitoring of radio frequency (RF) ablation. The proton frequency shift (PFS) technique, an accurate and fast imaging method for temperature quantification, was used to synthesize thermal maps after RF ablation in an open 0.5 T MR system under ex vivo and in vivo conditions. Calibration experiments with 1.5% agarose gel yielded a chen-Acal shift factor of 0.011 -k 0.001 ppm/degreesC (r(2) = 0.96). Three gradient echo (GRE) pulse sequences were tested for thermal mapping by comparison with fiberoptic thermometer (Luxtron Model 760) readings. Temperature uncertainty decreased from high to low bandwidths (BW): +/-5.9 degreesC at BW = 15.6 kHz, +/-1.4 degreesC at BW = 3.9 kHz, and +/-0.8 degreesC at BW = 2.5 kHz. In vitro experiments (N = 9) in the paraspinal muscle yielded a chemical shift factor of 0.008 +/- 0.001 ppm/degreesC. Temperature uncertainty was determined as +/-2.7 degreesC (BW = 3.9 kHz, TE = 19.3 msec). The same experiments carried out in the paraspinal muscle (N = 9) of a fully anesthetized pig resulted In a temperature uncertainty of +/-4.3 degreesC (BW = 3.9 kHz, TE = 19.3 msec), which is higher than it is in vitro conditions (P < 0.15). Quantitative temperature monitoring of RF ablation Is feasible in a 0.5 T open-configured MR scanner under ex vivo and in vivo conditions using the PFS technique. (C) 2001 Wiley-Liss, Inc.
引用
收藏
页码:437 / 444
页数:8
相关论文
共 25 条
[1]   REAL-TIME MAGNETIC-RESONANCE-IMAGING OF LASER HEAT DEPOSITION IN TISSUE [J].
BLEIER, AR ;
JOLESZ, FA ;
COHEN, MS ;
WEISSKOFF, RM ;
DALCANTON, JJ ;
HIGUCHI, N ;
FEINBERG, DA ;
ROSEN, BR ;
MCKINSTRY, RC ;
HUSHEK, SG .
MAGNETIC RESONANCE IN MEDICINE, 1991, 21 (01) :132-137
[2]  
BOTNAR R, 1997, P 5 ANN M ISMRM VANC
[3]  
Callaghan PT., 1993, Principles of Nuclear Magnetic Resonance Microscopy
[4]   Optimization of spoiled gradient-echo phase imaging for in vivo localization of a focused ultrasound beam [J].
Chung, AH ;
Hynynen, K ;
Colucci, V ;
Oshio, K ;
Cline, HE ;
Jolesz, FA .
MAGNETIC RESONANCE IN MEDICINE, 1996, 36 (05) :745-752
[5]   MR TEMPERATURE MAPPING OF FOCUSED ULTRASOUND SURGERY [J].
CLINE, HE ;
HYNYNEN, K ;
HARDY, CJ ;
WATKINS, RD ;
SCHENCK, JF ;
JOLESZ, FA .
MAGNETIC RESONANCE IN MEDICINE, 1994, 31 (06) :628-636
[6]   Simultaneous magnetic resonance phase and magnitude temperature maps in muscle [J].
Cline, HE ;
Hynynen, K ;
Schneider, E ;
Hardy, CJ ;
Maier, SE ;
Watkins, RD ;
Jolesz, FA .
MAGNETIC RESONANCE IN MEDICINE, 1996, 35 (03) :309-315
[7]   ONLINE MONITORING OF ULTRASONIC SURGERY WITH MR-IMAGING [J].
DARKAZANLI, A ;
HYNYNEN, K ;
UNGER, EC ;
SCHENCK, JF .
JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1993, 3 (03) :509-514
[8]  
DEPOORTER J, 1995, MAGN RESON MED, V34, P359, DOI 10.1002/mrm.1910340313
[9]  
DEPOORTER J, 1995, MAGNET RESON MED, V33, P74, DOI 10.1002/mrm.1910330111
[10]   ADAPTIVE TECHNIQUE FOR HIGH-DEFINITION MR IMAGING OF MOVING STRUCTURES [J].
EHMAN, RL ;
FELMLEE, JP .
RADIOLOGY, 1989, 173 (01) :255-263