The Use of a Transition Rod May Prevent Proximal Junctional Kyphosis in the Thoracic Spine After Scoliosis Surgery A Finite Element Analysis

被引:130
作者
Cahill, Patrick J. [1 ]
Wang, Wenhai [2 ]
Asghar, Jahangir [1 ]
Booker, Rashad [3 ]
Betz, Randal R. [1 ]
Ramsey, Christopher [2 ]
Baran, George [2 ]
机构
[1] Shriners Hosp Children, Philadelphia, PA 19140 USA
[2] Temple Univ, Coll Engn, Philadelphia, PA 19122 USA
[3] Albert Einstein Med Ctr, Dept Orthopaed Surg, Philadelphia, PA 19141 USA
关键词
interspinous ligament; proximal junctional kyphosis; supraspinous ligament; tapered rod; ADOLESCENT IDIOPATHIC SCOLIOSIS; ADJACENT SEGMENT INSTABILITY; HUMAN LUMBAR; BIOMECHANICAL PROPERTIES; MECHANICAL-PROPERTIES; INTERSPINAL LIGAMENT; INSTRUMENTATION; FUSION; BEHAVIOR; LAMINECTOMY;
D O I
10.1097/BRS.0b013e318246d4f2
中图分类号
R74 [神经病学与精神病学];
学科分类号
100204 [神经病学];
摘要
Study Design. Finite element analysis. Objective. Via finite element analysis: (1) to demonstrate the abnormal forces present at the top of a scoliosis construct, (2) to demonstrate the importance of an intact interspinous and supraspinous ligament (ISL/SSL) complex, and (3) to evaluate a transition rod (a rod that has a short taper to a smaller diameter at one end) as an implant solution to diminish these pathomechanics, regardless of the integrity of the ISL/SSL complex. Summary of Background Data. The pathophysiology of increased nucleus pressure and increased angular displacement may contribute to proximal junctional kyphosis. Furthermore, high implant stress can be demonstrated at the upper end of the construct, possibly leading to the risk of implant failure. Methods. A finite element model was constructed to simulate a thoracic spinal fusion. The model was altered to remove the ISL/SSL complex at the level above the construct. Then, the model was altered again by extending the construct one level superior with a transition rod. The angular displacement, the maximum pressure in the nucleus, and stress within the implant were extracted from computational results under 2 conditions: load control and displacement control. The testing was performed with both titanium and stainless steel implants. Results. Pressure in the nucleus and angular displacement are all increased when the ISL/SSL complex is removed immediately above the instrumented levels, whereas the screw pullout force and maximum stress within the screw are decreased. The nucleus pressure increases by more than 50%. The angular displacement increases by 19% to 26%. This absence of the ISL/SSL complex simulates the clinical scenario that occurs when these structures are iatrogenically detached. Abnormal mechanics can be restored to normal level by extending the construct rostral one level with a transition rod. Furthermore, the elevated nucleus pressure and angular displacement noted even when the ISL/SSL complex is intact can be avoided with the use of a transition rod. Under the same bending moment (3 Nm), the nucleus pressure at the level immediately cephalad is up to 23% lower than the pressure in a standard construct. The angular displacement is 18% to 19% less than the standard construct. The maximum implant stress is also decreased by as much as 60%. Conclusion. Finite element modeling suggests that the pathomechanics at the proximal end of a scoliosis construct may be diminished by preserving the ISL/SSL complex and possibly completely eliminated with the use of rods with a diameter transition at the most proximal level.
引用
收藏
页码:E687 / E695
页数:9
相关论文
共 46 条
[1]
The Effect of Posterior Thoracic Spine Anatomical Structures on Motion Segment Flexion Stiffness [J].
Anderson, Andy L. ;
McIff, Terence E. ;
Asher, Marc A. ;
Burton, Douglas C. ;
Glattes, R. Christopher .
SPINE, 2009, 34 (05) :441-446
[2]
Atici Teoman, 2004, Acta Orthop Belg, V70, P344
[3]
Anisotropy of the elastic modulus of trabecular bone specimens from different anatomical locations [J].
Augat, P ;
Link, T ;
Lang, TF ;
Lin, JC ;
Majumdar, S ;
Genant, HK .
MEDICAL ENGINEERING & PHYSICS, 1998, 20 (02) :124-131
[4]
The Dynesys lumbar spinal stabilization system - A preliminary report on positional magnetic resonance imaging findings [J].
Beastall, James ;
Karadimas, Efthimios ;
Siddiqui, Manal ;
Nicol, Malcolm ;
Hughes, Justin ;
Smith, Francis ;
Wardlaw, Douglas .
SPINE, 2007, 32 (06) :685-690
[5]
BIOMECHANICAL PROPERTIES OF SPINAL LIGAMENTS AND A HISTOLOGICAL STUDY OF THE SUPRASPINAL LIGAMENT IN TRACTION [J].
CHAZAL, J ;
TANGUY, A ;
BOURGES, M ;
GAUREL, G ;
ESCANDE, G ;
GUILLOT, M ;
VANNEUVILLE, G .
JOURNAL OF BIOMECHANICS, 1985, 18 (03) :167-176
[6]
Chen LH, 2006, BIO-MED MATER ENG, V16, P261
[7]
Incidence and Risk Factors for Proximal and Distal Junctional Kyphosis Following Surgical Treatment for Scheuermann Kyphosis Minimum Five-Year Follow-up [J].
Denis, Francis ;
Sun, Edward C. ;
Winter, Robert B. .
SPINE, 2009, 34 (20) :E729-E734
[8]
Biomechanical comparison between fusion of two vertebrae and implantation of an artificial intervertebral disc [J].
Denozière, G ;
Ku, DN .
JOURNAL OF BIOMECHANICS, 2006, 39 (04) :766-775
[9]
INSITU MECHANICAL-BEHAVIOR OF POSTERIOR SPINAL LIGAMENTS IN THE LUMBAR REGION - AN INVITRO STUDY [J].
DUMAS, GA ;
BEAUDOIN, L ;
DROUIN, G .
JOURNAL OF BIOMECHANICS, 1987, 20 (03) :301-&
[10]
Edidin A.A., 2006, SPINE TECHNOLOGY HDB, P63, DOI [10.1016/B978-012369390-7/50005-8, DOI 10.1016/B978-012369390-7/50005-8]