Migration and right atrial perforation of an accufix atrial lead retention wire following partial lead removal during myomectomy

被引:1
作者
Gerstenfeld, EP [1 ]
Balarajan, Y [1 ]
Cooke, R [1 ]
Mittleman, RS [1 ]
机构
[1] Univ Massachusetts, Med Ctr, Dept Med, Div Cardiol,Sect Cardiac Electrophysiol & Pacing, Worcester, MA 01655 USA
关键词
atrial lead; hypertrophic obstructive cardiomyopathy; lead extraction; pacemaker;
D O I
10.1378/chest.114.2.637
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
A 36-year-old man with a history of hypertrophic obstructive cardiomyopathy presented to the emergency room with "stabbing" chest pain. He had undergone dual-chamber pacemaker implantation in 1993 using an atrial lead (Accufix; Telectronics; Englewood, Cole) and a myomectomy in 1996 during which the distal portion of the atrial lead was removed. Digital fluoroscopy revealed that the retention wire had migrated out of the remaining atrial lead and perforated the right atrium. The retention wire was successfully removed pel cutaneously. The need for complete removal of the retention wire in the Accufix lead at the time of open-heart surgery is emphasized.
引用
收藏
页码:637 / 639
页数:5
相关论文
共 3 条
[1]  
DAUD EG, 1996, AM HEART J, V131, P266
[2]   Extraction of the Telectronics Accufix 330-801 atrial lead: The Mayo Clinic experience [J].
Lloyd, MA ;
Hayes, DL ;
Holmes, DR ;
Stanson, AW ;
Espinosa, RE ;
Osborn, MJ ;
McGoon, MD .
MAYO CLINIC PROCEEDINGS, 1996, 71 (03) :230-234
[3]   ATRIAL J PACING LEAD RETENTION WIRE FRACTURE - RADIOGRAPHIC ASSESSMENT, INCIDENCE OF FRACTURE, AND CLINICAL MANAGEMENT [J].
LLOYD, MA ;
HAYES, DL ;
HOLMES, DR .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1995, 18 (05) :958-964