Extraction of the Telectronics Accufix 330-801 atrial lead: The Mayo Clinic experience

被引:15
作者
Lloyd, MA
Hayes, DL
Holmes, DR
Stanson, AW
Espinosa, RE
Osborn, MJ
McGoon, MD
机构
[1] MAYO CLIN,DEPT DIAGNOST RADIOL,ROCHESTER,MN
[2] MAYO CLIN,DIV CARDIOVASC DIS & INTERNAL MED,ROCHESTER,MN
关键词
D O I
10.4065/71.3.230
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objective: To document a cumulative experience with removal of the Telectronics 330-801 Accufix atrial lead after it was withdrawn from the market. Design: We reviewed our results with 96 attempted lead extraction procedures through Dec. 1, 1995. Material and Methods: All patients underwent preoperative assessment, including fluoroscopic screening of the lead. Patients had blood typing and screening done before the extraction procedure, and extractions were performed with cardiac surgical intervention available. Pacing dependence was assessed to determine the need for temporary pacing. Results: The mean patient age was 65.7 +/- 1.6 years, and the mean duration of lead implantation was 31.0 +/- 1.4 months. Of the 96 lead extractions attempted, 94 were successful. Of the 96 cases, the retention wires were normal in 29, fractured without protrusion in 34, and fractured with protrusion in 33. Fifty-three leads were removed with use of simple traction only, and seven leads were removed by using a locking stylet and telescoping sheaths. The Cook workstation was used in the removal of 32 leads through the femoral vein; in 1 of these 32 patients, the retention wire was removed before lead extraction. The retention wire was removed but the lead was left intact in one patient. One lead was removed during an operative maze procedure. Associated complications were minimal. Conclusion: Although lead extraction is a potentially fatal procedure, the Telectronics 330-801 Accufix atrial lead can be extracted safely with minimal morbidity when performed by experienced personnel.
引用
收藏
页码:230 / 234
页数:5
相关论文
共 8 条
[1]  
BYRD CL, 1990, PACE PACING CLIN ELE, V13, P1870
[2]  
DENNIS JW, 1994, IMPORTANT PRODUCT NO
[3]   INTRAVASCULAR LEAD EXTRACTION USING LOCKING STYLETS, SHEATHS, AND OTHER TECHNIQUES [J].
FEARNOT, NE ;
SMITH, HJ ;
GOODE, LB ;
BYRD, CL ;
WILKOFF, BL ;
SELLERS, TD .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1990, 13 (12) :1864-1870
[4]   ATRIAL J PACING LEAD RETENTION WIRE FRACTURE - RADIOGRAPHIC ASSESSMENT, INCIDENCE OF FRACTURE, AND CLINICAL MANAGEMENT [J].
LLOYD, MA ;
HAYES, DL ;
HOLMES, DR .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1995, 18 (05) :958-964
[5]   SNARE REMOVAL OF A TELECTRONICS ACCUFIX ATRIAL J RETENTION WIRE [J].
LLOYD, MA ;
HAYES, DL ;
STANSON, AW ;
HOLMES, DR .
MAYO CLINIC PROCEEDINGS, 1995, 70 (04) :376-379
[6]   5-YEARS EXPERIENCE WITH INTRAVASCULAR LEAD EXTRACTION [J].
SMITH, HJ ;
FEARNOT, NE ;
BYRD, CL ;
WILKOFF, BL ;
LOVE, CJ ;
SELLERS, TD .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1994, 17 (11) :2016-2020
[7]   VENOUS OBSTRUCTION DUE TO PERMANENT TRANSVENOUS PACEMAKER ELECTRODES - TREATMENT WITH PERCUTANEOUS TRANSLUMINAL BALLOON VENOPLASTY [J].
SPITTELL, PC ;
VLIETSTRA, RE ;
HAYES, DL ;
HIGANO, ST .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1990, 13 (03) :271-274
[8]  
1995, MDDI REPORTS GRAY SH, V2, P12