Lessons learned from the first clinical implants of the DeBakey ventricular assist device axial pump: A single center report

被引:30
作者
Wieselthaler, GM [1 ]
Schima, H [1 ]
Lassnigg, AM [1 ]
Dworschak, M [1 ]
Pacher, R [1 ]
Grimm, M [1 ]
Wolner, E [1 ]
机构
[1] Univ Vienna, Dept Cardiothorac Surg, A-1090 Vienna, Austria
关键词
D O I
10.1016/S0003-4975(00)02636-9
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background. The bridge to transplantation with pulsatile mechanical assist devices became a standard procedure for patients deteriorating on the waiting list. Recently, continuous flow axial impeller pumps were introduced to clinical application offering new advantages. Methods. From November 1998 till September 2000, 6 male patients (mean age 53 plus or minus 11 years) with end-stage left heart failure were implanted with a DeBakey ventricular assist device (VAD) axial-flow pump for bridge to transplantation. Results. Three patients were successfully transplanted after 74, 115, and 117 days, respectively. Two other patients died after 25 and 133 days. One patient is still on the device after 108 days. Because of modification of the implantation technique after the first 2 patients, mean pump-flow within the first 3 weeks was increased from 4.3 +/- 0.6 L/min to 6.7 +/- 0.3 L/min. Patients were put on regular bicycle-ergometer training and improved their exercise capacities up to a mean maximum oxygen consumption of 20.2 ml/kg/min. Conclusions. Initial implants of the DeBakey VAD demonstrated support properties comparable to pulsatile pumps but without significant restrictions for extended use. (C) 2001 by The Society of Thoracic Surgeons.
引用
收藏
页码:S139 / S143
页数:5
相关论文
共 10 条
[1]   Development of the Nimbus/University of Pittsburgh innovative ventricular assist system [J].
Butler, KC ;
Dow, JJ ;
Litwak, P ;
Kormos, RL ;
Borovetz, HS .
ANNALS OF THORACIC SURGERY, 1999, 68 (02) :790-794
[2]  
DEBAKEY ME, 1998, HEART REPLACEMENT AR, P407
[3]   IMPROVED MORTALITY AND REHABILITATION OF TRANSPLANT CANDIDATES TREATED WITH A LONG-TERM IMPLANTABLE LEFT-VENTRICULAR ASSIST SYSTEM [J].
FRAZIER, OH ;
ROSE, EA ;
MCCARTHY, P ;
BURTON, NA ;
TECTOR, A ;
LEVIN, H ;
KAYNE, HL ;
POIRIER, VL ;
DASSE, KA .
ANNALS OF SURGERY, 1995, 222 (03) :327-338
[4]   Infection during circulatory support with ventricular assist devices [J].
Holman, WL ;
Skinner, JL ;
Waites, KB ;
Benza, RL ;
McGiffin, DC ;
Kirklin, JK .
ANNALS OF THORACIC SURGERY, 1999, 68 (02) :711-716
[5]  
JARVIK R, 2000, ASAIO J, V46, P156
[6]   Comparison of functional capacity in patients with end-stage heart failure following implantation of a left ventricular assist device versus heart transplantation:: Results of the experience with left ventricular assist device with exercise trial [J].
Jaski, BE ;
Lingle, RJ ;
Kim, J ;
Branch, KR ;
Goldsmith, R ;
Johnson, MR ;
Lahpor, JR ;
Icenogle, TB ;
Piña, I ;
Adamson, R ;
Favrot, LK ;
Dembitsky, WP .
JOURNAL OF HEART AND LUNG TRANSPLANTATION, 1999, 18 (11) :1031-1040
[7]   One hundred patients with the HeartMate left ventricular assist device: Evolving concepts and technology [J].
McCarthy, PM ;
Smedira, NO ;
Vargo, RL ;
Goormastic, M ;
Hobbs, RE ;
Starling, RC ;
Young, JB .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 1998, 115 (04) :904-912
[8]   Exercise capacity recovers slowly but fully in patients with a left ventricular assist device [J].
Nishimura, M ;
Radovancevic, B ;
Odegaard, P ;
Myers, T ;
Springer, W ;
Frazier, OH .
ASAIO JOURNAL, 1996, 42 (05) :M568-M570
[9]   100 long-term implantable left ventricular assist devices: The Columbia Presbyterian Interim Experience [J].
Sun, BC ;
Catanese, KA ;
Spanier, TB ;
Flannery, MR ;
Gardocki, MT ;
Marcus, LS ;
Levin, HR ;
Rose, EA ;
Oz, MC .
ANNALS OF THORACIC SURGERY, 1999, 68 (02) :688-694
[10]   First clinical experience with the DeBakey VAD continuous-axial-flow pump for bridge to transplantation [J].
Wieselthaler, GM ;
Schima, H ;
Hiesmayr, M ;
Pacher, R ;
Laufer, G ;
Noon, GP ;
DeBakey, M ;
Wolner, E .
CIRCULATION, 2000, 101 (04) :356-359