Fully automated open access platform for rapid, combined serial evaporation and sample reformatting

被引:3
作者
Benali, Otman [1 ]
Davies, Gary [1 ]
Deal, Martyn [1 ]
Farrant, Elizabeth [1 ]
Guthrie, Duncan [1 ,2 ]
Holden, John [1 ,3 ]
Wheeler, Rob [1 ]
机构
[1] GlaxoSmithKline Pharmaceut Technol Dev, Harlow CM19 5AW, Essex, England
[2] Vaportec Ltd, Ingham IP31 1NQ, Suffolk, England
[3] Aitken Sci, Thame OX9 2AH, Oxon, England
来源
JOURNAL OF COMBINATORIAL CHEMISTRY | 2008年 / 10卷 / 01期
关键词
D O I
10.1021/cc700149w
中图分类号
O69 [应用化学];
学科分类号
081704 ;
摘要
This paper reports a novel evaporator and its integration with art automated sample handling system to create a high throughput evaporation platform. The Vaportec V-10 evaporator uses a high speed rotation motor (similar to 6000 rpm) to spin the vial containing a sample, creating a thin film of solvent which can be readily evaporated by the application of heat to the vial, while the consequent centrifugal force prevents "bumping". An intelligent algorithm controls pressure and temperature for optimum solvent removal conditions and end of run detection, critical for automation. The system allows the option of evaporation directly from a sample source vial, or alternatively, integrated liquid handling facilities provide the capability of transferring samples portionwise from a (large) source vial or bottle to a (small) daughter container, enabling efficient sample reformatting, with minimum user intervention. The open access system makes significant advances over current vacuum centrifugal evaporators in terms of evaporation rate and ease of automation. The evaporator's main features, the integration of robotics to provide automation, and examples of evaporation rates of a wide range of solvents from a variety of containers are described.
引用
收藏
页码:52 / 55
页数:4
相关论文
共 9 条
  • [1] A flow process for the multi-step synthesis of the alkaloid natural product oxomaritidine: a new paradigm for molecular assembly
    Baxendale, Ian R.
    Deeley, Jon
    Griffiths-Jones, Charlotte M.
    Ley, Steven V.
    Saaby, Steen
    Tranmer, Geoffrey K.
    [J]. CHEMICAL COMMUNICATIONS, 2006, (24) : 2566 - 2568
  • [2] A simple and inexpensive device for removal of solvent from a large collection of sample tubes
    Gelb, MH
    Holm, BP
    [J]. JOURNAL OF COMBINATORIAL CHEMISTRY, 2006, 8 (01): : 15 - 17
  • [3] GUTHRIE D, 2005, Patent No. 065799
  • [4] GUTHRIE D, 2002, Patent No. 2384724
  • [5] Centrifugal concentrator for the substitution of nitrogen blow-down micro-concentration in dioxin/polychlorinated biphenyl sample preparation
    Kemmochi, Y
    Tsutsumi, K
    Arikawa, A
    Nakazawa, H
    [J]. JOURNAL OF CHROMATOGRAPHY A, 2002, 943 (02) : 295 - 297
  • [6] Further development of a robust workup process for solution-phase high-throughput library synthesis to address environmental and sample tracking issues
    Kuroda, Noritaka
    Hird, Nick
    Cork, David G.
    [J]. JOURNAL OF COMBINATORIAL CHEMISTRY, 2006, 8 (04): : 505 - 512
  • [7] Mullin R, 2004, CHEM ENG NEWS, V82, P23
  • [8] SPADONI S, 2004, CHEM TODAY, V23, P36
  • [9] Plate-to-plate fluorous solid-phase extraction for solution-phase parallel synthesis
    Zhang, W
    Lu, YM
    Nagashima, T
    [J]. JOURNAL OF COMBINATORIAL CHEMISTRY, 2005, 7 (06): : 893 - 897