Recent innovations in bipolar electrosurgery

被引:33
作者
Kennedy, JS [1 ]
Buysse, SP [1 ]
Lawes, KR [1 ]
Ryan, TP [1 ]
机构
[1] Valleylab Inc, Boulder, CO 80301 USA
关键词
bipolar; bipolar electrosurgery; vessel ligation;
D O I
10.3109/13645709909153141
中图分类号
R61 [外科手术学];
学科分类号
摘要
Bipolar electrosurgery has been used routinely in endoscopic surgery since its inception and is associated with increased safety and precise energy delivery. The most common tool is the laparoscopic bipolar forceps. In the last decade variations on this tool, along with other configurations, have been developed that expand the use of bipolar electrosurgery. These variations include bipolar hooks, needles, scissors, forceps with cutting capability and optimised instrument/energy source systems. These new technologies make bipolar electrosurgery a very versatile tool for haemostatic dissection in surgery. This paper reviews laparoscopic bipolar instruments and systems, including the advantages and limitations of each configuration. A new optimised forceps/generator system for surgical vessel-sealing, called the 'LigaSure' (Valleylab Inc., Boulder, CO, USA) vessel-seating system is also discussed. The LigaSure system offers increased reliability with a single application and utility on a broader range of vessel sizes than conventional bipolar electrosurgery. A study comparing seal integrity of standard bipolar coagulation, ultrasonic coagulation and vessel-sealing technology is presented.
引用
收藏
页码:95 / 99
页数:5
相关论文
共 15 条
[1]  
BAGGISH MS, 1995, FERTIL STERIL, V63, P422
[2]   STUDIES ON COAGULATION AND THE DEVELOPMENT OF AN AUTOMATIC COMPUTERIZED BIPOLAR COAGULATOR - TECHNICAL NOTE [J].
BERGDAHL, B ;
VALLFORS, B .
JOURNAL OF NEUROSURGERY, 1991, 75 (01) :148-151
[3]  
EDELMAN DS, 1995, SURG LAPAROSC ENDOSC, V5, P459
[4]   ARTERY WELD STRENGTH AFTER BIPOLAR ELECTROCOAGULATION IS AFFECTED BY ELECTROLYTE-SOLUTIONS AROUND THE ARTERY [J].
HARRISON, JD ;
MORRIS, DL .
GASTROINTESTINAL ENDOSCOPY, 1993, 39 (04) :546-548
[5]   A randomised controlled comparison of injection, thermal, and mechanical endoscopic methods of haemostasis on mesenteric vessels [J].
Hepworth, CC ;
Kadirkamanathan, SS ;
Gong, F ;
Swain, CP .
GUT, 1998, 42 (04) :462-469
[6]   Controlled radio frequency vessel sealing system for surgical applications [J].
Kennedy, J ;
Buysse, S ;
Chandler, J ;
Eggleston, J ;
Taylor, K ;
Thomsen, S .
SURGICAL APPLICATIONS OF ENERGY, PROCEEDINGS OF, 1998, 3249 :125-129
[7]   High-burst-strength, feedback-controlled bipolar vessel sealing [J].
Kennedy, JS ;
Stranahan, PL ;
Taylor, KD ;
Chandler, JG .
SURGICAL ENDOSCOPY-ULTRASOUND AND INTERVENTIONAL TECHNIQUES, 1998, 12 (06) :876-878
[8]   New application of bipolar coagulation in laparoscopic surgery [J].
McKernan, JB ;
Stuto, A ;
Champion, JK .
SURGICAL LAPAROSCOPY & ENDOSCOPY, 1996, 6 (05) :335-340
[9]  
Mueller W, 1993, Endosc Surg Allied Technol, V1, P91
[10]  
PHIPPS JH, 1994, GYNAECOL ENDOSC, P35