Characterization of a new qQq-FTICR mass spectrometer for post-translational modification analysis and top-down tandem mass Spectrometry of whole proteins

被引:48
作者
Jebanathirajah, JA
Pittman, JL
Thomson, BA
Budnik, BA
Kaur, P
Rape, M
Kirschner, M
Costello, CE
O'Connor, PB
机构
[1] Boston Univ, Dept Biochem, Sch Med, Boston, MA 02118 USA
[2] MDS SCIEX, Concord, ON, Canada
[3] Boston Univ, Sch Med, Cardiovasc Proteom Ctr, Boston, MA 02215 USA
[4] Harvard Univ, Sch Med, Dept Syst Biol, Boston, MA 02115 USA
关键词
D O I
10.1016/j.jasms.2005.08.008
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The use of a new electrospray qQq Fourier transform ion cyclotron mass spectrometer (qQq-FTICR MS) instrument for biologic applications is described. This qQq-FTICR mass spectrometer was designed for the study of post-translationally modified proteins and for top-down analysis of biologically relevant protein samples. The utility of the instrument for the analysis of phosphorylation, a common and important post-translational modification, was investigated. Phosphorylation was chosen as an example because it is ubiquitous and challenging to analyze. In addition, the use of the instrument for top-down sequencing of proteins was explored since this instrument offers particular advantages to this approach. Top-down sequencing was performed on different proteins, including commercially available proteins and biologically derived samples such as the human E2 ubiquitin conjugating enzyme, UbCH10. A good sequence tag was obtained for the human UbCH10, allowing the unambiguous identification of the protein. The instrument was built with a commercially produced front end: a focusing rf-only quadrupole (Q0), followed by a resolving quadrupole (Q1), and a LINAC quadrupole collision cell (Q2), in combination with an FTICR mass analyzer. It has utility in the analysis of samples found in substoichiometric concentrations, as ions can be isolated in the mass resolving Q1 and accumulated in Q2 before analysis in the ICR cell. The speed and efficacy of the Q2 cooling and fragmentation was demonstrated on an LCMS-compatible time scale, and detection limits for phosphopeptides in the 10 amol/mu L range (pM) were demonstrated. The instrument was designed to make several fragmentation methods available, including nozzle-skimmer fragmentation, Q2 collisionally activated dissociation (Q2 CAD), multipole storage assisted dissociation (MSAD), electron capture dissociation (ECD), infrared multiphoton induced dissociation (IRMPD), and sustained off resonance irradiation (SORI) CAD, thus allowing a variety of MSn experiments. A particularly useful aspect of the system was the use of Q1 to isolate ions from complex mixtures with narrow windows of isolation less than I m/z. These features enable top-down protein analysis experiments as well structural characterization of minor components of complex mixtures.
引用
收藏
页码:1985 / 1999
页数:15
相关论文
共 75 条
[31]   Novel Linac II electrode geometry for creating an axial field in a multipole ion guide [J].
Loboda, A ;
Krutchinsky, A ;
Loboda, O ;
McNabb, J ;
Spicer, V ;
Ens, W ;
Standing, K .
EUROPEAN JOURNAL OF MASS SPECTROMETRY, 2000, 6 (06) :531-536
[32]   ERROR TOLERANT IDENTIFICATION OF PEPTIDES IN SEQUENCE DATABASES BY PEPTIDE SEQUENCE TAGS [J].
MANN, M ;
WILM, M .
ANALYTICAL CHEMISTRY, 1994, 66 (24) :4390-4399
[33]  
Marshall AG, 2002, ANAL CHEM, V74, p253A
[34]   TAILORED EXCITATION FOR FOURIER-TRANSFORM ION-CYCLOTRON RESONANCE MASS-SPECTROMETRY [J].
MARSHALL, AG ;
WANG, TCL ;
RICCA, TL .
JOURNAL OF THE AMERICAN CHEMICAL SOCIETY, 1985, 107 (26) :7893-7897
[35]   Ion "threshing": Collisionally activated dissociation in an external octopole ion trap by oscillation of an axial electric potential gradient [J].
McFarland, MA ;
Hendrickson, CL ;
Marshall, AG .
ANALYTICAL CHEMISTRY, 2004, 76 (06) :1545-1549
[36]   HIGH-RESOLUTION TANDEM FT MASS-SPECTROMETRY ABOVE 10-KDA [J].
MCLAFFERTY, FW .
ACCOUNTS OF CHEMICAL RESEARCH, 1994, 27 (11) :379-386
[37]   Molecular-level description of proteins from Saccharomyces cerevisiae using quadrupole FT hybrid mass spectrometry for top down proteomics [J].
Meng, FY ;
Du, Y ;
Miller, LM ;
Patrie, SM ;
Robinson, DE ;
Kelleher, NL .
ANALYTICAL CHEMISTRY, 2004, 76 (10) :2852-2858
[38]   Localization of O-glycosylation sites in peptides by electron capture dissociation in a fourier transform mass spectrometer [J].
Mirgorodskaya, E ;
Roepstorff, P ;
Zubarev, RA .
ANALYTICAL CHEMISTRY, 1999, 71 (20) :4431-4436
[39]   A general method for precalculation of parameters for sustained off resonance irradiation/collision-induced dissociation [J].
Mirgorodskaya, E ;
O'Connor, PB ;
Costello, CE .
JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 2002, 13 (04) :318-324
[40]   Mass spectrometric determination of O-glycosylation sites using β-elimination and partial acid hydrolysis [J].
Mirgorodskaya, E ;
Hassan, H ;
Clausen, H ;
Roepstorff, P .
ANALYTICAL CHEMISTRY, 2001, 73 (06) :1263-1269