Uncertainty contributions to species specific isotope dilution analysis. Part 2. Determination of methylmercury by HPLC coupled with quadrupole and multicollector ICP-MS

被引:19
作者
Clough, R
Belt, ST
Evans, EH
Fairman, B
Catterick, T
机构
[1] Univ Plymouth, Sch Environm Sci, Plymouth Environm Res Ctr, Plymouth PL4 8AA, Devon, England
[2] LGC, Teddington TW11 0LY, Middx, England
关键词
D O I
10.1039/b305454n
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Species-specific isotope dilution analysis (IDA) has been used to determine the mass fraction of methylmercury in a fish muscle certified reference material, DORM-2. Two spike solutions were prepared and their mass fractions determined by coupling high-performance liquid chromatography (HPLC) with both quadrupole (Q) and multicollector (MC) inductively coupled plasma mass spectrometry (ICP-MS) and uncertainty budgets calculated. For the Q-ICP-MS determination the relative standard uncertainty of the spike was 4.1%. The major uncertainty contribution (62%) arose from the uncertainties associated with the relative isotopic abundances of Hg isotopes as given by IUPAC. The measured isotope amount ratios (Hg-200:Hg-199 and Tl-205:Tl-203) contributed 37% of the combined uncertainty. For the multicollector instrument the relative standard uncertainty of the spike solution mass fraction was 0.7%. In this case the uncertainty budget was dominated by the Hg isotopic uncertainties, with a contribution of 99%. The uncertainty contributions from the measured isotope amount ratios were reduced to 0.7% of the total when using MC-ICP-MS. The two spike solutions were separately used to determine the mass fraction of methylmercury in DORM-2 CRM by species-specific IDA, using HPLC coupled with quadrupole ICP-MS. The values found were 4.45 +/- 0.90 mug g(-1) and 4.25 +/- 0.47 mug g(-1) using spikes characterised by Q- and MC-ICP-MS, respectively (the uncertainties quoted are the expanded uncertainty, k = 2), compared with a certified value of 4.47 +/- 0.32. The major uncertainty contribution for each determination was from the measured isotope amount ratios (70%) with lesser contributions from the uncertainties associated with the natural isotopic abundance of Hg and the spike mass fraction.
引用
收藏
页码:1039 / 1046
页数:8
相关论文
共 23 条
  • [1] Estimation of uncertainties in ICP-MS analysis: a practical methodology
    Barwick, VJ
    Ellison, SLR
    Fairman, B
    [J]. ANALYTICA CHIMICA ACTA, 1999, 394 (2-3) : 281 - 291
  • [2] Precise and accurate isotope ratio measurements by ICP-MS
    Becker, JS
    Dietze, HJ
    [J]. FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 2000, 368 (01): : 23 - 30
  • [3] Artifact formation of methyl mercury during aqueous distillation and alternative techniques for the extraction of methyl mercury from environmental samples
    Bloom, NS
    Colman, JA
    Barber, L
    [J]. FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1997, 358 (03): : 371 - 377
  • [4] BRICHE C, 1998, THESIS U ANTWERPEN
  • [5] BRICHE C, 1998, THESIS U INSTELLING
  • [6] Briche CSJW, 2001, ANAL CHIM ACTA, V437, P1
  • [7] Structured approach to achieving high accuracy measurements with isotope dilution inductively coupled plasma mass spectrometry
    Catterick, T
    Fairman, B
    Harrington, C
    [J]. JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1998, 13 (09) : 1009 - 1013
  • [8] Uncertainty contributions to species specific isotope dilution analysis.: Part 1.: Characterisation and stability of 199Hg and 13C isotopically enriched methylmercury by 1H NMR
    Clough, R
    Belt, ST
    Evans, EH
    Sutton, P
    Fairman, B
    Catterick, T
    [J]. JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2003, 18 (09) : 1033 - 1038
  • [9] Isotope dilution ICP-MS for speciation studies
    Clough, R
    Truscatt, J
    Belt, ST
    Evans, EH
    Fairman, B
    Catterick, T
    [J]. APPLIED SPECTROSCOPY REVIEWS, 2003, 38 (01) : 101 - 132
  • [10] CLOUGH R, IN PRESS ANAL CHIM A