Comparison of Complete Versus Incomplete Stent Frame Expansion After Transcatheter Aortic Valve Implantation With Medtronic Coke Valve Bioprosthesis

被引:18
作者
Jilaihawi, Hasan [2 ]
Chin, Derek [1 ]
Spyt, Tomasz [1 ]
Jeilan, Mohamed [1 ]
Vasa-Nicotera, Mariuca [1 ]
Mohamed, Noor [1 ]
Bence, Johan [1 ]
Logtens, Elaine [1 ]
Kovac, Jan [1 ]
机构
[1] Glenfield Gen Hosp, Leicester LE3 9QP, Leics, England
[2] Cedars Sinai Heart Inst, Los Angeles, CA USA
关键词
HIGH-RISK PATIENTS; PROSTHESIS; REPLACEMENT; STENOSIS;
D O I
10.1016/j.amjcard.2011.02.317
中图分类号
R5 [内科学];
学科分类号
100201 [内科学];
摘要
We sought to determine the significance of incomplete stent frame expansion after transcatheter aortic valve implantation with the Medtronic-CoreValve device. Incomplete coronary stent expansion is a well-described phenomenon. Transcatheter valves are mounted on stents; however, the incidence of incomplete stent expansion after transcatheter aortic valve implantation, its sequelae and predictors are poorly elucidated. The 18Fr Core Valve revalving system was used to treat anatomically and clinically suitable patients with severe calcific aortic stenosis. The postdeployment stent dimensions were measured on fluoroscopic images at multiple levels of the stent frame. Incomplete expansion at each was defined as <100% expected and complete expansion as >= 100%. These parameters were correlated to the hemodynamic and clinical end points. A total of 50 consecutive patients underwent transcatheter aortic valve implantation with the Medtronic-CoreValve at a single center from January 2007 to December 2008. For the inflow portion, incomplete expansion was seen in 54% of patients and was unrelated to the aortic valve area, peak or mean aortic valve gradients, or measures of aortic regurgitation, although it was paradoxically associated with a lower incidence of prosthesis patient mismatch. Incomplete expansion of the constrained portion was seen in 62% of the patients and displayed a trend toward a greater incidence of aortic regurgitation grade 2 or greater, although this was rare. Incomplete expansion of the Medtronic Core Valve stent frame is common. For the most part, valvular hemodynamic function was satisfactory, regardless of the degree of expansion of the stent frame that carries it and a strategy of reluctant postdilation in the context of incomplete stent frame expansion was supported. (C) 2011 Elsevier Inc. All rights reserved. (Am J Cardiol 2011;107:1830-1837)
引用
收藏
页码:1830 / 1837
页数:8
相关论文
共 11 条
[1]
The potential of myocardial perfusion scintigraphy for risk stratification of asymptomatic patients with type 2 diabetes [J].
Bax, Jeroen J. ;
Bonow, Robert O. ;
Tschoepe, Diethelm ;
Inzucchi, Silvio E. ;
Barrett, Eugene .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2006, 48 (04) :754-760
[2]
Early experience with percutaneous transcatheter implantation of heart valve prosthesis for the treatment of end-stage inoperable patients with calcific aortic stenosis [J].
Cribier, A ;
Eltchaninoff, H ;
Tron, C ;
Bauer, F ;
Agatiello, C ;
Sebagh, L ;
Bash, A ;
Nusimovici, D ;
Litzler, PY ;
Bessou, JP ;
Leon, MB .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2004, 43 (04) :698-703
[3]
Percutaneous transcatheter implantation of an aortic valve prosthesis for calcific aortic stenosis - First human case description [J].
Cribier, A ;
Eltchaninoff, H ;
Bash, A ;
Borenstein, N ;
Tron, C ;
Bauer, F ;
Derumeaux, G ;
Anselme, F ;
Laborde, F ;
Leon, MB .
CIRCULATION, 2002, 106 (24) :3006-3008
[4]
Percutaneous aortic valve replacement for severe aortic stenosis in high-rick patients using the second- and current third-generation self-expanding CoreValve prosthesis - Device success and 30-day clinical outcome [J].
Grube, Eberhard ;
Schuler, Gerhard ;
Buellesfeld, Lutz ;
Gerckens, Ulrich ;
Linke, Axel ;
Wenaweser, Peter ;
Sauren, Barthel ;
Mohr, Friedrich-Wilhelm ;
Walther, Thomas ;
Zickmann, Bernfried ;
Iversen, Stein ;
Felderhoff, Thomas ;
Cartier, Raymond ;
Bonan, Raoul .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2007, 50 (01) :69-76
[5]
Percutaneous implantation of the CoreValve self-expanding valve prosthesis in high-risk patients with aortic valve disease - The Siegburg First-in-Man Study [J].
Grube, Eberhard ;
Laborde, Jean C. ;
Gerckens, Ulrich ;
Felderhoff, Thomas ;
Sauren, Barthel ;
Buellesfeld, Lutz ;
Mueller, Ralf ;
Menichelli, Maurizio ;
Schmidt, Thomas ;
Zickmann, Bernfried ;
Iversen, Stein ;
Stone, Gregg W. .
CIRCULATION, 2006, 114 (15) :1616-1624
[6]
Prosthesis-patient mismatch after transcatheter aortic valve implantation with the Medtronic-Corevalve bioprosthesis [J].
Jilaihawi, Hasan ;
Chin, Derek ;
Spyt, Tomasz ;
Jeilan, Mohamed ;
Vasa-Nicotera, Mariuca ;
Bence, Johan ;
Logtens, Elaine ;
Kovac, Jan .
EUROPEAN HEART JOURNAL, 2010, 31 (07) :857-864
[7]
Geometry and Degree of Apposition of the CoreValve ReValving System With Multislice Computed Tomography After Implantation in Patients With Aortic Stenosis [J].
Schultz, Carl J. ;
Weustink, Annick ;
Piazza, Nicolo ;
Otten, Amber ;
Mollet, Nico ;
Krestin, Gabriel ;
van Geuns, Robert J. ;
de Feyter, Pim ;
Serruys, Patrick W. J. ;
de Jaegere, Peter .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2009, 54 (10) :911-918
[8]
Safety of coronary Sirolimus-Eluting Stents in daily clinical practice - One-year follow-up of the e-cypher registry [J].
Urban, P ;
Gershlick, AH ;
Guagliumi, G ;
Guyon, P ;
Lotan, C ;
Schofer, J ;
Seth, A ;
Sousa, JE ;
Wijns, W ;
Berge, C ;
Deme, M ;
Stoll, HP .
CIRCULATION, 2006, 113 (11) :1434-1441
[9]
Transapical minimally invasive aortic valve implantation - Multicenter experience [J].
Walther, Thomas ;
Simon, Paul ;
Dewey, Todd ;
Wimmer-Greinecker, Gerhard ;
Falk, Volkmar ;
Kasimir, Marie T. ;
Doss, Mirko ;
Borger, Michael A. ;
Schuler, Gerhard ;
Glogar, Dietmar ;
Fehske, Wolfgang ;
Wolner, Ernst ;
Mohr, Friedrich W. ;
Mack, Michael .
CIRCULATION, 2007, 116 (11) :I240-I245
[10]
Percutaneous transarterial aortic valve replacement in selected high-risk patients with aortic stenosis [J].
Webb, John G. ;
Pasupati, Sanjeevan ;
Humphries, Karin ;
Thompson, Christopher ;
Altwegg, Lukas ;
Moss, Robert ;
Sinhal, Ajay ;
Carere, Ronald G. ;
Munt, Brad ;
Ricci, Donald ;
Ye, Jian ;
Cheung, Anson ;
Lichtenstein, Sam V. .
CIRCULATION, 2007, 116 (07) :755-763